Monte Carlo simulation of image properties of an X-ray intensifying screen

被引:1
作者
Wang, Y [1 ]
Wang, JJ
Wang, KL
Liu, GZ
Liu, YQ
机构
[1] Tsing Hua Univ, Dept Engn Phys, Beijing 100084, Peoples R China
[2] NW Inst Nucl Technol, Xian 710024, Peoples R China
关键词
Monte Carlo simulation; intensifying screen; point spread function; detective quantum efficiency;
D O I
10.1016/S0168-9002(00)00289-8
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A Monte Carlo simulation program named MCPEP has been developed. Based on the existing simulation program that simulates the transfer of X-ray photons and the secondary electrons, MCPEP also simulates the light photons in the screen. The performances of an intensifying screen (Gd-2 O2S:Tb) with different thickness and different X-ray energies have been analyzed by MCPEP. The calculated light photon probability distribution, average light photon number per absorbed X-ray photon, statistical factor for light emission, X-ray detection efficiency, detective quantum efficiency (DQE) and point spread function (PSF) of the screen are presented. (C) 2000 Published by Elsevier Science B.V. All rights reserved.
引用
收藏
页码:567 / 570
页数:4
相关论文
共 8 条
[1]  
AERTS W, 1982, MATER EVAL, V40, P1071
[2]   STUDIES OF X-RAY-ENERGY ABSORPTION AND QUANTUM NOISE PROPERTIES OF X-RAY SCREENS BY USE OF MONTE-CARLO SIMULATION [J].
CHAN, HP ;
DOI, K .
MEDICAL PHYSICS, 1984, 11 (01) :37-46
[3]  
Dainty JC., 1974, IMAGE SCI
[4]   IMAGE-INFORMATION TRANSFER PROPERTIES OF X-RAY FLUORESCENT SCREENS [J].
DICK, CE ;
MOTZ, JW .
MEDICAL PHYSICS, 1981, 8 (03) :337-346
[5]   UTILIZATION OF MONOENERGETIC X-RAY-BEAMS TO EXAMINE THE PROPERTIES OF RADIOGRAPHIC INTENSIFYING SCREENS [J].
DICK, CE ;
MOTZ, JW .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1981, 28 (02) :1554-1558
[6]   Y2O2S-EU PHOSPHOR SCREENS EVALUATION [J].
GIAKOUMAKIS, GE ;
NOMICOS, CD ;
SKOUNTZOS, P ;
KOUTROUBAS, S ;
ZISOS, A ;
YIAKOUMAKIS, EN ;
KATSARIOTI, MC ;
KALIAKATSOS, JA ;
ROVITHI, M ;
PANAYIOTAKIS, GS ;
EVANGELOU, EK .
MEDICAL PHYSICS, 1993, 20 (01) :79-83
[7]   IMAGE-INFORMATION TRANSFER PROPERTIES OF X-RAY INTENSIFYING SCREENS IN THE ENERGY-RANGE FROM 17 TO 320 KEV [J].
GINZBURG, A ;
DICK, CE .
MEDICAL PHYSICS, 1993, 20 (04) :1013-1021
[8]   CALCULATION OF MODULATION TRANSFER-FUNCTIONS OF X-RAY FLUORESCENT SCREENS [J].
SWANK, RK .
APPLIED OPTICS, 1973, 12 (08) :1865-1870