Renal Arterial Blood Flow Measurement by Breath-Held MRI: Accuracy in Phantom Scans and Reproducibility in Healthy Subjects

被引:18
作者
Dambreville, Samuel [1 ]
Chapman, Arlene B.
Torres, Vicente E. [2 ]
King, Bernard F. [2 ]
Wallin, Ashley K. [1 ]
Frakes, David H. [3 ]
Yoganathan, Ajit P. [1 ]
Wijayawardana, Sameera R.
Easley, Kirk
Bae, Kyongtae T. [4 ]
Brummer, Marijn E.
机构
[1] Georgia Inst Technol, Atlanta, GA 30332 USA
[2] Mayo Clin, Rochester, MN USA
[3] Arizona State Univ, Tempe, AZ USA
[4] Univ Pittsburgh, Pittsburgh, PA USA
关键词
quantitative flow; renal blood flow; flow validation; kidney imaging; phase velocity imaging; QUANTITATION; CLEARANCE;
D O I
10.1002/mrm.22278
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
This study evaluates reliability of current technology for measurement of renal arterial blood flow by breath-held velocity-encoded MRI. Overall accuracy was determined by comparing MRI measurements with known flow in controlled-flow-loop phantom studies. Measurements using prospective and retrospective gating methods were compared in phantom studies with pulsatile flow, not revealing significant differences. Phantom study results showed good accuracy, with deviations from true flow consistently below 13% for vessel diameters 3mm and above. Reproducibility in human subjects was evaluated by repeated studies in six healthy control subjects, comparing immediate repetition of the scan, repetition of the scan plane scouting, and week-to-week variation in repeated studies. The standard deviation in the 4-week protocol of repeated in vivo measurements of single-kidney renal flow in normal subjects was 59.7 mL/min, corresponding with an average coefficient of variation of 10.55%. Comparison of renal arterial blood flow reproducibility with and without gadolinium contrast showed no significant differences in mean or standard deviation. A breakdown among error components showed corresponding marginal standard deviations (coefficients of variation) 23.8 mL/min (4.21%) for immediate repetition of the breath-held flow scan, 39.13 mL/min (6.90%) for repeated plane scouting, and 40.76 mL/min (7.20%) for weekly fluctuations in renal blood flow. Magn Reson Med 63:940-950, 2010. (C) 2010 Wiley-Liss, Inc.
引用
收藏
页码:940 / 950
页数:11
相关论文
共 32 条
[1]   BLOOD-FLOW EFFECTS IN MAGNETIC-RESONANCE IMAGING [J].
AXEL, L .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1984, 143 (06) :1157-1166
[2]   Renal blood flow measurements with use of phase-contrast magnetic resonance imaging: Normal values and reproducibility [J].
Bax, L ;
Bakker, CJG ;
Klein, WM ;
Blanken, N ;
Beutler, JJ ;
Mali, WPTRM .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2005, 16 (06) :807-814
[3]   Association of gadolinium based magnetic resonance imaging contrast agents and nephrogenic systemic fibrosis [J].
Bhave, Gautam ;
Lewis, Julia B. ;
Chang, Sam S. .
JOURNAL OF UROLOGY, 2008, 180 (03) :830-835
[4]  
Bland JM, 1999, STAT METHODS MED RES, V8, P135, DOI 10.1177/096228029900800204
[5]   Interleaved gradient echo planar (IGEPI] and phase contrast CINE-PC flow measurements in the renal artery [J].
Bock, M ;
Schoenberg, SO ;
Schad, LR ;
Knopp, MV ;
Essig, M ;
van Kaick, G .
JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1998, 8 (04) :889-895
[6]   COMPARING PAIRED DATA - A SIMULTANEOUS TEST FOR MEANS AND VARIANCES [J].
BRADLEY, EL ;
BLACKWOOD, LG .
AMERICAN STATISTICIAN, 1989, 43 (04) :234-235
[7]   MEASUREMENT OF FLOW WITH NMR IMAGING USING A GRADIENT PULSE AND PHASE DIFFERENCE TECHNIQUE [J].
BRYANT, DJ ;
PAYNE, JA ;
FIRMIN, DN ;
LONGMORE, DB .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1984, 8 (04) :588-593
[8]   Polyvinyl alcohol cryogel: An ideal phantom material for MR studies of arterial flow and elasticity [J].
Chu, KC ;
Rutt, BK .
MAGNETIC RESONANCE IN MEDICINE, 1997, 37 (02) :314-319
[9]  
Cortsen M, 1996, ACTA RADIOL, V37, P79
[10]   Scleromyxoedema-like cutaneous diseases in renal-dialysis patients [J].
Cowper, SE ;
Robin, HS ;
Steinberg, SM ;
Su, LD ;
Gupta, S ;
LeBoit, PE .
LANCET, 2000, 356 (9234) :1000-1001