Comparison between electrocardiogram- and photoplethysmogram-derived features for atrial fibrillation detection in free-living conditions

被引:50
作者
Eerikainen, Linda M. [1 ,2 ]
Bonomi, Alberto G. [2 ]
Schipper, Fons [2 ]
Dekker, Lukas R. C. [1 ,3 ]
Vullings, Rik [1 ]
de Morree, Helma M. [2 ]
Aarts, Ronald M. [1 ,2 ]
机构
[1] Eindhoven Univ Technol, Dept Elect Engn, NL-5612 AZ Eindhoven, Netherlands
[2] Philips Res, Eindhoven, Netherlands
[3] Catharina Hosp, Dept Cardiol, Eindhoven, Netherlands
关键词
atrial fibrillation; electrocardiography; photoplet hysinogra phy; movement artefacts; PHYSIOLOGICAL TIME-SERIES; ENTROPY;
D O I
10.1088/1361-6579/aad2c0
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Objective: Atrial fibrillation (AF) is the most commonly experienced arrhythmia and it increases the risk of stroke and heart failure. The challenge in detecting the presence of AF is the occasional and asymptomatic manifestation of the condition. Long-term monitoring can increase the sensitivity of detecting intermittent AF episodes, however it is either cumbersome or invasive and costly with electrocardiography (ECG). Photoplethysmography (PPG) is an unobtrusive measuring modality enabling heart rate monitoring, and promising results have been presented in detecting AF. However, there is still limited knowledge about the applicability of the PPG solutions in free-living conditions. The aim of this study was to compare the inter-beat interval derived features for AF detection between ECG and wrist-worn PPG in daily life. Approach: The data consisted of 24 h ECG, PPG, and accelerometer measurements from 27 patients (eight AF, 19 non-AF). In total, seven features (Shannon entropy, root mean square of successive differences (RMSSD), normalized RMSSD, pNN40, pNN70, sample entropy, and coefficient of sample entropy (CosEn)) were compared. Body movement was measured with the accelerometer and used with three different thresholds to exclude PPG segments affected by movement. Main results: CosEn resulted as the best performing feature from ECG with Cohens kappa 0.95. When the strictest movement threshold was applied, the same performance was obtained with PPG (kappa = 0.96). In addition, pNN40 and pNN70 reached similar results with the same threshold (kappa = 0.95 and 0.94), but were more robust with respect to movement artefacts. The coverage of PPG was 24.0%-57.6% depending on the movement threshold compared to 92.1% of ECG. Significance: The inter-beat interval features derived from PPG are equivalent to the ones from ECG for AF detection. Movement artefacts substantially worsen PPG-based AF monitoring in free-living conditions, therefore monitoring coverage needs to be carefully selected. Wrist-worn PPG still provides a promising technology for long-term AF monitoring.
引用
收藏
页数:11
相关论文
共 30 条
[1]   Photoplethysmography and its application in clinical physiological measurement [J].
Allen, John .
PHYSIOLOGICAL MEASUREMENT, 2007, 28 (03) :R1-R39
[2]  
Bonomi AG, 2016, COMPUT CARDIOL CONF, V43, P277
[3]   Uncovering Atrial Fibrillation Beyond Short-Term Monitoring in Cryptogenic Stroke Patients Three-Year Results From the Cryptogenic Stroke and Underlying Atrial Fibrillation Trial [J].
Brachmann, Johannes ;
Morillo, Carlos A. ;
Sanna, Tommaso ;
Di Lazzaro, Vincenzo ;
Diener, Hans-Christoph ;
Bernstein, Richard A. ;
Rymer, Marylin ;
Ziegler, Paul D. ;
Liu, Shufeng ;
Passman, Rod S. .
CIRCULATION-ARRHYTHMIA AND ELECTROPHYSIOLOGY, 2016, 9 (01)
[4]  
Camm AJ, 2010, EUR HEART J, V31, P2369, DOI [10.1093/eurheartj/ehq278, 10.1093/europace/euq350]
[5]  
Chan P. H., 2016, J AM HEART ASSOC, V5, P1
[6]  
Chong J.W., 2015, IEEE T BIOMED ENG, V19, P1
[7]   A COEFFICIENT OF AGREEMENT FOR NOMINAL SCALES [J].
COHEN, J .
EDUCATIONAL AND PSYCHOLOGICAL MEASUREMENT, 1960, 20 (01) :37-46
[8]   Detection of atrial fibrillation episodes using a wristband device [J].
Corino, Valentina D. A. ;
Laureanti, Rita ;
Ferranti, Lorenzo ;
Scarpini, Giorgio ;
Lombardi, Federico ;
Mainardi, Luca T. .
PHYSIOLOGICAL MEASUREMENT, 2017, 38 (05) :787-799
[9]   Automatic Real Time Detection of Atrial Fibrillation [J].
Dash, S. ;
Chon, K. H. ;
Lu, S. ;
Raeder, E. A. .
ANNALS OF BIOMEDICAL ENGINEERING, 2009, 37 (09) :1701-1709
[10]  
Eerikainen LM, 2017, 44 COMP CARD C CINC, P3