A digital method for computing target margins in radiotherapy

被引:28
作者
Bedford, JL [1 ]
Shentall, GS
机构
[1] Inst Canc Res, Joint Dept Phys, Surrey SM2 5PT, England
[2] Royal Marsden NHS Trust, Surrey SM2 5PT, England
[3] Royal Preston Hosp, Lancashire & Lakeland Radiotherpy Unit, Preston PR2 9HT, Lancs, England
关键词
radiotherapy; treatment planning; outline; margin; planning target volume;
D O I
10.1118/1.598184
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Computer methods for determining the planning target volume from the gross tumor volume for both conventional and conformal radiotherapy are presented. Production of a two-dimensional (2D) treatment plan is assisted by projecting outlines of the gross tumor onto a single transverse plane, so that the total extent of the tumor can be easily visualized. A 2D margin can then be added to the resulting outline, so as to account for microscopic tumor spread, organ motion, and setup uncertainty. The margin may be anisotropic to account for the known differences in setup accuracy in the anterior, posterior, left and right directions. For three-dimensional (3D) treatment planning, it is necessary to add a 3D margin to the gross tumor volume to define the planning target volume, the anisotropy of the margin now being allowed to extend to the superior and inferior directions also. Robust methods for automatically calculating these regions are described, and illustrated for the case of a prostate tumor. It is demonstrated that while a slicewise 2D margin is adequate for 2D planning, a fully 3D margin must be used for 3D conformal planning to avoid underdosing the superior and inferior extremities of the clinical target volume. (C) 1998 American Association of Physicists in Medicine.
引用
收藏
页码:224 / 231
页数:8
相关论文
共 17 条
  • [1] Three dimensional planning target volumes: A model and a software tool
    AustinSeymour, M
    Kalet, I
    McDonald, J
    KromhoutSchiro, S
    Jacky, J
    Hummel, S
    Unger, J
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1995, 33 (05): : 1073 - 1080
  • [2] MEASUREMENT OF PROSTATE MOVEMENT OVER THE COURSE OF ROUTINE RADIOTHERAPY USING IMPLANTED MARKERS
    BALTER, JM
    SANDLER, HM
    LAM, K
    BREE, RL
    LICHTER, AS
    TENHAKEN, RK
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1995, 31 (01): : 113 - 118
  • [3] Target margins for random geometrical treatment uncertainties in conformal radiotherapy
    Bel, A
    vanHerk, M
    Lebesque, JV
    [J]. MEDICAL PHYSICS, 1996, 23 (09) : 1537 - 1545
  • [4] Implications of 3-dimensional target shape and motion in aperture design
    Fontenla, E
    Pelizzari, CA
    Chen, GTY
    [J]. MEDICAL PHYSICS, 1996, 23 (08) : 1431 - 1441
  • [5] A RANDOMIZED TRIAL OF PATIENT REPOSITIONING DURING RADIOTHERAPY USING A MEGAVOLTAGE IMAGING-SYSTEM
    GILDERSLEVE, J
    DEARNALEY, DP
    EVANS, PM
    LAW, M
    RAWLINGS, C
    SWINDELL, W
    [J]. RADIOTHERAPY AND ONCOLOGY, 1994, 31 (02) : 161 - 168
  • [6] ACCURACY OF FIELD ALIGNMENT IN RADIOTHERAPY OF HEAD AND NECK-CANCER UTILIZING INDIVIDUALIZED FACE MASK IMMOBILIZATION - A RETROSPECTIVE ANALYSIS OF CLINICAL-PRACTICE
    HESS, CF
    KORTMANN, RD
    JANY, R
    HAMBERGER, A
    BAMBERG, M
    [J]. RADIOTHERAPY AND ONCOLOGY, 1995, 34 (01) : 69 - 72
  • [7] Accuracy of pelvic radiotherapy: Prospective analysis of 90 patients in a randomised trial of blocked versus standard radiotherapy
    Huddart, RA
    Nahum, A
    Neal, A
    McLean, M
    Dearnaley, DP
    Law, M
    Dyer, J
    Tait, D
    [J]. RADIOTHERAPY AND ONCOLOGY, 1996, 39 (01) : 19 - 29
  • [8] AN ESTIMATE OF THE MARGIN REQUIRED WHEN DEFINING BLOCKS AROUND THE PROSTATE
    JONES, D
    HAFERMANN, MD
    RIEKE, JW
    VERMEULEN, SS
    [J]. BRITISH JOURNAL OF RADIOLOGY, 1995, 68 (811) : 740 - 746
  • [9] THE ROYAL-MARSDEN-HOSPITAL PELVIC RADIOTHERAPY TRIAL - TECHNICAL ASPECTS AND QUALITY ASSURANCE
    MAYLES, WPM
    CHOW, M
    DYER, J
    FERNANDEZ, EM
    HEISIG, S
    KNIGHT, RT
    MOORE, I
    NAHUM, AE
    SHENTALL, GS
    TAIT, DM
    [J]. RADIOTHERAPY AND ONCOLOGY, 1993, 29 (02) : 184 - 191
  • [10] THE VALUE OF NONUNIFORM MARGINS FOR 6-FIELD CONFORMAL IRRADIATION OF LOCALIZED PROSTATE-CANCER
    PICKETT, B
    ROACH, M
    VERHEY, L
    HORINE, P
    MALFATTI, C
    AKAZAWA, C
    DEA, D
    VARAD, B
    RATHBUN, C
    PHILLIPS, TL
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1995, 32 (01): : 211 - 218