Tribological behaviour of 3D printed materials for small joint implants: A pilot study

被引:9
作者
Ranusa, Matus [1 ]
Cipek, Pavel [1 ]
Vrbka, Martin [1 ]
Palousek, David [1 ]
Krupka, Ivan [1 ]
Hartl, Martin [1 ]
机构
[1] Brno Univ Technol, Fac Mech Engn, Tech 2896-2, Brno 61669, Czech Republic
关键词
Additive manufacturing; Small joints; Ti6Al4V alloy; CoCrMo alloy; Synovial fluid; Fluorescent microscopy; HALLUX-RIGIDUS; FILM FORMATION; METALLIC BIOMATERIALS; SURFACE MODIFICATION; CLINICAL-OUTCOMES; LUBRICATION; ALBUMIN; WEAR; ARTHROPLASTY; ADSORPTION;
D O I
10.1016/j.jmbbm.2022.105274
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Additive manufacturing is a progressive method in endoprosthetics enabling customisation of implants. However, the challenge is to design articulating surfaces that are wear resistant in a long term. To tackle this challenge, it is necessary to understand the interaction between the surfaces and the lubricant synovial fluid as well as the mechanism of lubrication film formation. In this study we observed three synovial fluid constituents (albumin, gamma-globulin a hyaluronic acid) in the contact area simultaneously with the coefficient of friction (CoF). Two metal alloys, CoCrMo and Ti6Al4V covered by DLC, were selected for the experiments as they are both suitable for the additive technology and commonly used in implants manufacturing. The tests were running on a custom-made pin-on-plate tribometer equipped for optical fluorescence measurements. The test apparatus allowed reciprocating motion and observation of the contact area. Our results showed differences in the pace of the CoF increase between the alloys and differences between the samples manufactured by the conventional and the additive manufacturing method. Both the conventionally and additively manufactured CrCrMo samples showed a stable CoF values from the beginning of the experiments: 0.66 (SD 0.02) for the conventional manufacturing CrCrMo samples and 0.53 (SD 0.01) for the additive manufacturing CrCrMo samples. The Ti6Al4V/DLC samples showed a stable CoF values similar to those of the CoCrMo samples not until the 240 s of experiment. These results are related to the protein formation in the contact areas as suggested by a similar increasing trend of the individual synovial fluid constituents in the contact. Increasing protein amounts in the contact led to CoF increase. There were also differences in the ratios of the individual constituents, where both the CoCrMo and the Ti6Al4V/ DLC samples manufactured additionally showed lower concentrations of gamma-globulin and hyalumnic acid. These pilot results, on the one hand, support the potential of the additive manufacturing in the implantology and, on the other hand, demonstrate the application of a method suitable for the analysis of the lubricant behaviour in the contact. The method is limited in using the intensity of the emitted light to observe the behaviour of the lubricant film. Future development of the method will require a direct quantification of film thickness.
引用
收藏
页数:14
相关论文
共 70 条
[1]   Influence of albumin on the tribological behavior of Ag-Ti (C, N) thin films for orthopedic implants [J].
Almeida Alves, C. F. ;
Oliveira, F. ;
Carvalho, I. ;
Piedade, A. P. ;
Carvalho, S. .
MATERIALS SCIENCE & ENGINEERING C-MATERIALS FOR BIOLOGICAL APPLICATIONS, 2014, 34 :22-28
[2]   Arthroplasty of the Small Joints [J].
Arndt, Moritz ;
Henniger, Martina ;
Hoffmann, Tobias ;
Rehart, Stefan .
AKTUELLE RHEUMATOLOGIE, 2020, 45 (01) :77-85
[3]   Implant surface design for improved implant stability - A study on Ti6Al4V dense and cellular structures produced by Selective Laser Melting [J].
Bartolomeu, F. ;
Costa, M. M. ;
Gomes, J. R. ;
Alves, N. ;
Abreu, C. S. ;
Silva, F. S. ;
Miranda, G. .
TRIBOLOGY INTERNATIONAL, 2019, 129 :272-282
[4]   Surface conditioning of additively manufactured titanium implants and its influence on materials properties and in vitro biocompatibility [J].
Bernhardt, Anne ;
Schneider, Jakob ;
Schroeder, Andreas ;
Papadopoulous, Kiriaki ;
Lopez, Elena ;
Brueckner, Frank ;
Botzenhart, Ute .
MATERIALS SCIENCE & ENGINEERING C-MATERIALS FOR BIOLOGICAL APPLICATIONS, 2021, 119
[5]  
Brand Richard A, 2005, Iowa Orthop J, V25, P82
[6]  
Brown S.A., 1996, Medical Applications o f Titanium and Its Alloys, P231
[7]   Effects of surface coating on reducing friction and wear of orthopaedic implants [J].
Ching, Hee Ay ;
Choudhury, Dipankar ;
Nine, Md Julker ;
Abu Osman, Noor Azuan .
SCIENCE AND TECHNOLOGY OF ADVANCED MATERIALS, 2014, 15 (01)
[8]  
Choudhury D., 2015, TRIBOL T, P1
[9]   A novel functional layered diamond like carbon coating for orthopedics applications [J].
Choudhury, Dipankar ;
Morita, Takehiro ;
Sawae, Yoshinori ;
Lackner, Juergen M. ;
Towler, Mark ;
Krupka, Ivan .
DIAMOND AND RELATED MATERIALS, 2016, 61 :56-69
[10]  
Cipek P, 2019, Tribology Ind, V41, P387, DOI [10.24874/ti.2019.41.03.08, DOI 10.24874/TI.2019.41.03.08]