Investigation of the influence of blade configuration on the hemodynamic performance and blood damage of the centrifugal blood pump

被引:26
作者
Li, Yuan [1 ]
Yu, Jiachen [2 ]
Wang, Hongyu [1 ]
Xi, Yifeng [1 ]
Deng, Xiaoyan [1 ]
Chen, Zengsheng [1 ]
Fan, Yubo [1 ]
机构
[1] Beihang Univ, Key Lab Biomech & Mechanobiol, Beijing Adv Innovat Ctr Biomed Engn, Sch Biol Sci & Med Engn,Minist Educ, Beijing 100083, Peoples R China
[2] Beihang Univ, Sch SinoFrench Engineer, Beijing, Peoples R China
基金
中国国家自然科学基金; 国家重点研发计划;
关键词
blade configuration; centrifugal blood pump; hemolysis; platelet activation state; pressure head; VENTRICULAR ASSIST DEVICE; FLUID-DYNAMICS ANALYSIS; PLATELET ACTIVATION; SHEAR-STRESS; HEMOLYSIS; GEOMETRY; DESIGN; MODELS;
D O I
10.1111/aor.14265
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Purpose The design and optimization of centrifugal blood pumps are crucial for improved extracorporeal membrane oxygenation system performance. Secondary flow passages are common in centrifugal blood pumps, allowing for a high volume of unstable flow. Traditional design theory offers minimal guidance on the design and optimization of centrifugal blood pumps, so it's critical to understand how design parameter variables affect hydraulic performance and hemocompatibility. Methods Computational fluid dynamics (CFD) was employed to investigate the effects of blade number, blade wrap angle, blade thickness, and splitters on pressure head, hemolysis, and platelet activation state. Eulerian and Lagrangian features were used to analyze the flow fields and hemocompatibility metrics such as scalar shear stress, velocity distribution, and their correlation. Results The equalization of frictional and flow losses allow impellers with more blades and smaller wrap angles to have higher pressure heads, whereas the trade-off between the volume of high scalar shear stress and exposure time allows impellers with fewer blades and larger blade wrap angles to have a lower HI; there are configurations that increase the possibility of platelet activation for both number of blades and wrap angles. The hydraulic performance and hemocompatibility of centrifugal blood pumps are not affected by blade thickness. Compared to the main blades, splitters can improve the blood compatibility of a centrifugal blood pump with a small reduction in pressure head, but there is a trade-off between the length and location of the splitter that suppresses flow losses while reducing the velocity gradient. According to correlation analysis, pressure head, HI, and the volume of high shear stress were all substantially connected, and exposure time had a significant impact on HI. The platelet activation state was influenced by the average scalar shear stress and the volume of low velocity. Conclusion The findings reveal the impact of design variables on the performance of centrifugal blood pumps with secondary flow passages, as well as the relationship between hemocompatibility, hydraulic performance, and flow characteristics, and are useful for the design and optimization of this type of blood pump, as well as the prediction of clinical complications.
引用
收藏
页码:1817 / 1832
页数:16
相关论文
共 46 条
  • [1] [Anonymous], 2016, Crit Care, V20, P94
  • [2] Towards Non-thrombogenic Performance of Blood Recirculating Devices
    Bluestein, D.
    Chandran, K. B.
    Manning, K. B.
    [J]. ANNALS OF BIOMEDICAL ENGINEERING, 2010, 38 (03) : 1236 - 1256
  • [3] Thrombotic Risk of Rotor Speed Modulation Regimes of Contemporary Centrifugal Continuous-flow Left Ventricular Assist Devices
    Boraschi, Andrea
    Bozzi, Silvia
    Thamsen, Bente
    Granegger, Marcus
    Wiegmann, Lena
    Pappalardo, Federico
    Slepian, Marvin J.
    Kurtcuoglu, Vartan
    Redaelli, Alberto
    De Zelicourt, Diane
    Consolo, Filippo
    [J]. ASAIO JOURNAL, 2021, 67 (07) : 737 - 745
  • [4] High shear induces platelet dysfunction leading to enhanced thrombotic propensity and diminished hemostatic capacity
    Chen, Zengsheng
    Mondal, Nandan K.
    Zheng, Shirong
    Koenig, Steven C.
    Slaughter, Mark S.
    Griffith, Bartley P.
    Wu, Zhongjun J.
    [J]. PLATELETS, 2019, 30 (01) : 112 - 119
  • [5] Flow features and device-induced blood trauma in CF-VADs under a pulsatile blood flow condition: A CFD comparative study
    Chen, Zengsheng
    Jena, Sofen K.
    Giridharan, Guruprasad A.
    Koenig, Steven C.
    Slaughter, Mark S.
    Griffith, Bartley P.
    Wu, Zhongjun J.
    [J]. INTERNATIONAL JOURNAL FOR NUMERICAL METHODS IN BIOMEDICAL ENGINEERING, 2018, 34 (02)
  • [6] Paradoxical Effect of Nonphysiological Shear Stress on Platelets and von Willebrand Factor
    Chen, Zengsheng
    Mondal, Nandan K.
    Ding, Jun
    Koenig, Steven C.
    Slaughter, Mark S.
    Wu, Zhongjun J.
    [J]. ARTIFICIAL ORGANS, 2016, 40 (07) : 659 - 668
  • [7] Left Ventricular Assist Device Inflow Cannula Insertion Depth Influences Thrombosis Risk
    Chivukula, Venkat Keshav
    Beckman, Jennifer A.
    Li, Song
    Masri, S. Carolina
    Levy, Wayne C.
    Lin, Shin
    Cheng, Richard K.
    Farris, Stephen D.
    Wood, Gregory
    Dardas, Todd F.
    Kirkpatrick, James N.
    Koomalsingh, Kevin
    Zimpfer, Daniel
    Mackensen, G. Burkhard
    Chassagne, Fanette
    Mahr, Claudius
    Aliseda, Alberto
    [J]. ASAIO JOURNAL, 2020, 66 (07) : 766 - 773
  • [8] Left Ventricular Assist Device Inflow Cannula Angle and Thrombosis Risk
    Chivukula, Venkat Keshav
    Beckman, Jennifer A.
    Prisco, Anthony R.
    Dardas, Todd
    Lin, Shin
    Smith, Jason W.
    Mokadam, Nahush A.
    Aliseda, Alberto
    Mahr, Claudius
    [J]. CIRCULATION-HEART FAILURE, 2018, 11 (04)
  • [9] High Frequency Components of Hemodynamic Shear Stress Profiles are a Major Determinant of ShearMediated Platelet Activation in Therapeutic Blood Recirculating Devices
    Consolo, Filippo
    Sheriff, Jawaad
    Gorla, Silvia
    Magri, Nicolo
    Bluestein, Danny
    Pappalardo, Federico
    Slepian, Marvin J.
    Fiore, Gianfranco B.
    Redaelli, Alberto
    [J]. SCIENTIFIC REPORTS, 2017, 7
  • [10] A two-phase flow approach for modeling blood stasis and estimating the thrombosis potential of a ventricular assist device
    Dai, Wei-Feng
    Wu, Peng
    Liu, Guang-Mao
    [J]. INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2021, 44 (07) : 471 - 480