Instrument modifications that produced reduced plate heights <2 with sub-2 μm particles and 95% of theoretical efficiency at k=2 in supercritical fluid chromatography

被引:17
作者
Berger, Terry A. [1 ]
机构
[1] SFC Solut Inc, Washington, DC USA
关键词
Supercritical fluid chromatography (SFC); 1.8 mu m particles; Theoretical efficiency; h(r)< 2 @ k=2; Peak fidelity; Temperature gradients; BAND-BROADENING CONTRIBUTIONS; LIQUID-CHROMATOGRAPHY; TEMPERATURE-GRADIENTS; ACCURATE MEASUREMENTS; POROUS PARTICLES; PRESSURE DROPS; COLUMN; GAS; RESOLUTION; EQUATION;
D O I
10.1016/j.chroma.2016.03.021
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The concept of peak fidelity was shown to be helpful in modeling tubing and detector cell dimensions. Connection tubing and flow cell variances were modeled to determine appropriate internal ID's, lengths, and volumes. A low dispersion plumbing configuration, based on these calculations, was assembled to replace the standard plumbing and produced the reported results. The modifications made were straightforward using commercially available parts. The full theoretical efficiency of a 3 x 100 mm column packed with 1.8 mu m totally porous particles was achieved for the first time in supercritical fluid chromatography (SFC). Peak fidelity of >0.95 was maintained to below k= 2. A reduced plate height as low as 1.87 was measured. Thus, true "ultra high performance" SFC was achieved, with the results a major improvement from all previous SFC reports. Since there were no efficiency losses, none could be attributed to thermal gradients caused by the expansion of the fluid over large pressure drops, under the conditions used. Similarly, changes in diffusion coefficients caused by significant decreases in density during expansion are apparently balanced by the increase in linear velocity, keeping the ratio between the diffusion coefficient and the linear velocity a constant. Changing modifier concentration to change retention was shown to not be a significant problem. All these issues have been a concern in the past. Diffusion coefficients, and viscosity data needs to be collected at high pressures before the actual limits of SFC can be discovered. (C) 2016 Elsevier B.V. All rights reserved.
引用
收藏
页码:129 / 144
页数:16
相关论文
共 42 条
  • [1] BERGER TA, 1995, CHROMATOGRAPHIA, V41, P133, DOI [10.1007/BF02688014, 10.1007/BF02267944]
  • [2] BERGER TA, 1995, CHROMATOGRAPHIA, V41, P471, DOI 10.1007/BF02688070
  • [3] PACKED-COLUMN SUPERCRITICAL-FLUID CHROMATOGRAPHY WITH 220000 PLATES
    BERGER, TA
    WILSON, WH
    [J]. ANALYTICAL CHEMISTRY, 1993, 65 (10) : 1451 - 1455
  • [4] Berger TA., 1995, PACKED COLUMN SFC, DOI [10.1039/9781847551603, DOI 10.1039/9781847551603]
  • [5] Demonstration of High Speeds with Low Pressure Drops Using 1.8 μm Particles in SFC
    Berger, Terry A.
    [J]. CHROMATOGRAPHIA, 2010, 72 (7-8) : 597 - 602
  • [6] VARIANCE OF A ZONE MIGRATING IN A NONUNIFORM TIME-INVARIANT LINEAR MEDIUM
    BLUMBERG, LM
    BERGER, TA
    [J]. JOURNAL OF CHROMATOGRAPHY, 1992, 596 (01): : 1 - 13
  • [7] On axial temperature gradients due to large pressure drops in dense fluid chromatography
    Colgate, Sam O.
    Berger, Terry A.
    [J]. JOURNAL OF CHROMATOGRAPHY A, 2015, 1385 : 94 - 102
  • [8] Understanding and diminishing the extra-column band broadening effects in supercritical fluid chromatography
    De Pauw, Ruben
    Shoykhet , Konstantin
    Desmet, Gert
    Broeckhoven, Ken
    [J]. JOURNAL OF CHROMATOGRAPHY A, 2015, 1403 : 132 - 137
  • [9] Exploring the speed-resolution limits of supercritical fluid chromatography at ultra-high pressures
    De Pauw, Ruben
    Shoykhet , Konstantin
    Desmet, Gert
    Broeckhoven, Ken
    [J]. JOURNAL OF CHROMATOGRAPHY A, 2014, 1374 : 247 - 253
  • [10] de Reuck K. M., 1993, METHANOL INT THERMOD