The design features are described of a prompt gamma neutron activation analysis system at HANARO, a 30 MW research reactor in the Korea Atomic Energy Research Institute. The beam consists of polychromatic thermal neutrons diffracted by a set of pyrolytic graphite crystals at orders n in the range 1 less than or equal to n less than or equal to 6 at a Bragg angle of 45 degrees on a horizontal beam line. A neutron flux of 1.0 10(8) n.cm(-2) S-1 is calculated at the sample position from the reflectivity of the crystal which has been confirmed in a measurement of a diffracted neutron spectrum using a time-of-flight spectrometer and gold-wire activation. The fast neutron and gamma backgrounds will be low due to the use of a diffracted beam and a tapered collimator. The detection system comprises a 30% n-type HPGe detector, signal electronics and a fast ADC. The first application of this system Mlill be the analysis of boron concentration in biological samples for neutron capture therapy. Construction of the beam line and the arrangement of the detection system is proceeding.
引用
收藏
页码:413 / 416
页数:4
相关论文
共 14 条
[1]   DETERMINATION OF BORON IN FOOD AND BIOLOGICAL REFERENCE MATERIALS BY NEUTRON-CAPTURE PROMPT-GAMMA ACTIVATION [J].
ANDERSON, DL ;
CUNNINGHAM, WC ;
MACKEY, EA .
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 1990, 338 (04) :554-558
[2]   SECONDARY EXTINCTION AND NEUTRON CRYSTALLOGRAPHY [J].
BACON, GE ;
LOWDE, RD .
ACTA CRYSTALLOGRAPHICA, 1948, 1 (1-6) :303-314
[3]   COMPARISONS OF ATOMIC THERMAL MOTIONS FOR GRAPHITE AT 300-K BASED ON X-RAY, NEUTRON, AND PHONON-SPECTRUM DATA [J].
CHEN, R ;
TRUCANO, P .
ACTA CRYSTALLOGRAPHICA SECTION A, 1978, 34 (NOV) :979-982
[4]  
CHOI HD, 1998, P 6 M INT GROUP RES
[5]   A PROMPT GAMMA NEUTRON-ACTIVATION ANALYSIS FACILITY USING A DIFFRACTED BEAM [J].
HARLING, OK ;
CHABEUF, JM ;
LAMBERT, F ;
YASUDA, G .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 1993, 83 (04) :557-562
[6]  
JUN BJ, 1999, P 6 AS S RES REACT J
[7]   EFFECTS OF TARGET SHAPE AND NEUTRON-SCATTERING ON ELEMENT SENSITIVITIES FOR NEUTRON-CAPTURE PROMPT GAMMA-RAY ACTIVATION-ANALYSIS [J].
MACKEY, EA ;
GORDON, GE ;
LINDSTROM, RM ;
ANDERSON, DL .
ANALYTICAL CHEMISTRY, 1991, 63 (03) :288-292
[8]   USE OF SPHERICAL TARGETS TO MINIMIZE EFFECTS OF NEUTRON-SCATTERING BY HYDROGEN IN NEUTRON-CAPTURE PROMPT GAMMA-RAY ACTIVATION-ANALYSIS [J].
MACKEY, EA ;
GORDON, GE ;
LINDSTROM, RM ;
ANDERSON, DL .
ANALYTICAL CHEMISTRY, 1992, 64 (20) :2366-2371
[9]   THE SMALL-ANGLE NEUTRON-SCATTERING SPECTROMETER AT THE UNIVERSITY-OF-MISSOURI RESEARCH REACTOR [J].
MILDNER, DFR ;
BERLINER, R ;
PRINGLE, OA ;
KING, JS .
JOURNAL OF APPLIED CRYSTALLOGRAPHY, 1981, 14 (DEC) :370-382
[10]   Prompt-gamma activation analysis using the k0 approach [J].
Molnar, GL ;
Revay, Z ;
Paul, RL ;
Lindstrom, RM .
JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY, 1998, 234 (1-2) :21-26