A cartridge based sensor array platform for multiple coagulation measurements from plasma

被引:41
作者
Cakmak, O. [1 ]
Ermek, E. [1 ]
Kilinc, N. [1 ,2 ]
Bulut, S. [1 ]
Baris, I. [3 ]
Kavakli, I. H. [3 ]
Yaralioglu, G. G. [4 ]
Urey, Hakan [1 ]
机构
[1] Koc Univ, TR-34450 Istanbul, Turkey
[2] Gebze Inst Technol, Dept Phys, TR-41400 Gebze, Turkey
[3] Koc Univ, Dept Mol Biol & Genet, TR-34450 Istanbul, Turkey
[4] Ozyegin Univ, Fac Engn, Istanbul, Turkey
关键词
FREQUENCY-RESPONSE; BLOOD-COAGULATION; VISCOUS FLUIDS; POINT;
D O I
10.1039/c4lc00809j
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
This paper proposes a MEMS-based sensor array enabling multiple clot-time tests for plasma in one disposable microfluidic cartridge. The versatile LoC (Lab-on-Chip) platform technology is demonstrated here for real-time coagulation tests (activated Partial Thromboplastin Time (aPTT) and Prothrombin Time (PT)). The system has a reader unit and a disposable cartridge. The reader has no electrical connections to the cartridge. This enables simple and low-cost cartridge designs and avoids reliability problems associated with electrical connections. The cartridge consists of microfluidic channels and MEMS microcantilevers placed in each channel. The microcantilevers are made of electroplated nickel. They are actuated remotely using an external electro-coil and the read-out is also conducted remotely using a laser. The phase difference between the cantilever oscillation and the coil drive is monitored in real time. During coagulation, the viscosity of the blood plasma increases resulting in a change in the phase read-out. The proposed assay was tested on human and control plasma samples for PT and aPTT measurements. PT and aPTT measurements from control plasma samples are comparable with the manufacturer's datasheet and the commercial reference device. The measurement system has an overall 7.28% and 6.33% CV for PT and aPTT, respectively. For further implementation, the microfluidic channels of the cartridge were functionalized for PT and aPTT tests by drying specific reagents in each channel. Since simultaneous PT and aPTT measurements are needed in order to properly evaluate the coagulation system, one of the most prominent features of the proposed assay is enabling parallel measurement of different coagulation parameters. Additionally, the design of the cartridge and the read-out system as well as the obtained reproducible results with 10 mu l of the plasma samples suggest an opportunity for a possible point-of-care application.
引用
收藏
页码:113 / 120
页数:8
相关论文
共 28 条
[1]  
Cakmak O., 2013, MEMS BAS BLOOD PLASM, P1, DOI [10.1109/ICSENS.2013.6688342, DOI 10.1109/ICSENS.2013.6688342]
[2]  
Cakmak O, 2014, 2014 IEEE 27TH INTERNATIONAL CONFERENCE ON MICRO ELECTRO MECHANICAL SYSTEMS (MEMS), P330, DOI 10.1109/MEMSYS.2014.6765643
[3]   MEMS biosensor for blood plasma viscosity measurements [J].
Cakmak, Onur ;
Elbuken, Caglar ;
Ermek, Erhan ;
Bulut, Selma ;
Kilinc, Yasin ;
Baris, Ibrahim ;
Kavakli, Halil ;
Alaca, Erdem ;
Urey, Hakan .
NEW BIOTECHNOLOGY, 2012, 29 :S162-S163
[4]   Microcantilever based disposable viscosity sensor for serum and blood plasma measurements [J].
Cakmak, Onur ;
Elbuken, Caglar ;
Ermek, Erhan ;
Mostafazadeh, Aref ;
Baris, Ibrahim ;
Alaca, B. Erdem ;
Kavakli, Ibrahim Halil ;
Urey, Hakan .
METHODS, 2013, 63 (03) :225-232
[5]   Experimental validation of theoretical models for the frequency response of atomic force microscope cantilever beams immersed in fluids [J].
Chon, JWM ;
Mulvaney, P ;
Sader, JE .
JOURNAL OF APPLIED PHYSICS, 2000, 87 (08) :3978-3988
[6]  
CoaguChek XS, 2014, 06024 COAGUCHEK XS
[7]  
Djakov V., 2012, Google Pat., Patent No. [US8297110 B2, 8297110]
[8]  
Dufour I. A. M., 2012, J SENSORS, V2012, P1
[9]   Coagulation monitoring: Current techniques and clinical use of viscoelastic point-of-care coagulation devices [J].
Ganter, Michael T. ;
Hofer, Christoph K. .
ANESTHESIA AND ANALGESIA, 2008, 106 (05) :1366-1375
[10]   Biomaterial-associated thrombosis: roles of coagulation factors, complement, platelets and leukocytes [J].
Gorbet, MB ;
Sefton, MV .
BIOMATERIALS, 2004, 25 (26) :5681-5703