Biological behavior of sol-gel coated dental implants

被引:47
作者
Ramires, PA
Wennerberg, A
Johansson, CB
Cosentino, F
Tundo, S
Milella, E
机构
[1] PASTIS, CNRSM, Biomat Unit, I-72100 Brindisi, Italy
[2] Univ Gothenburg, Inst Surg Sci, Dept Biomat Handicap Res, Gothenburg, Sweden
[3] CNR, IME, Lecce, Italy
关键词
D O I
10.1023/A:1023412131314
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The biocompatibility of dental implants coated with titania/hydroxyapatite (HA) and titania/ bioactive glass (BG) composites obtained via sol-gel process was investigated using an in vitro and in vivo model. A device for the in vitrotesting of screw-shaped dental implants was developed, in order to well compare the two experimental models studying the behavior of human MG63 osteoblast-like cells seeded onto a particular geometry. The expression of some biochemical parameters of osteoblastic phenotype (alkaline phosphatase specific activity, collagen and osteocalcin production) and some indications on cells morphology obtained by scanning electron microscopy were evaluated. The in vitro and in vivo models were compared after implants insertion in rabbit tibia and femur. The removal torque and histomorphometric parameters ( percentage of bone in contact with implant surface and the amount of bone inside the threaded area) were examined. A good agreement was found between the in vitro and in vivo models. These experiments showed better performances of HA and BG sol-gel coated dental implants with respect to uncoated titanium; in particular, it was found that in vitro the HA coating stimulates osteoblastic cells in producing higher level of ALP and collagen, whereas in vivo this surface modification resulted in a higher removal torque and a larger bone-implant contact area. This behavior could be ascribed to the morphology and the chemical composition of the implants with rough and bioactive surfaces. (C) 2003 Kluwer Academic Publishers.
引用
收藏
页码:539 / 545
页数:7
相关论文
共 39 条
[1]  
Brunette D M, 1988, Int J Oral Maxillofac Implants, V3, P231
[2]   HYDROXYLAPATITE COATING OF POROUS IMPLANTS IMPROVES BONE INGROWTH AND INTERFACE ATTACHMENT STRENGTH [J].
COOK, SD ;
THOMAS, KA ;
DALTON, JE ;
VOLKMAN, TK ;
WHITECLOUD, TS ;
KAY, JF .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1992, 26 (08) :989-1001
[3]  
COSENTINO F, 1999, 15 EUR C BIOM BORD F
[4]   Topographical control of cells [J].
Curtis, A ;
Wilkinson, C .
BIOMATERIALS, 1997, 18 (24) :1573-1583
[5]   Conditions which promote mineralization at the bone-implant interface: A model in vitro study [J].
Dee, KC ;
Rueger, DC ;
Andersen, TT ;
Bizios, R .
BIOMATERIALS, 1996, 17 (02) :209-215
[6]  
DEGROOT K, 1983, BIOCERAMICS CALCIUM, P99
[7]   An image analysis method for the study of cell adhesion to biomaterials [J].
Dubois, JC ;
Souchier, C ;
Couble, ML ;
Exbrayat, P ;
Lissac, M .
BIOMATERIALS, 1999, 20 (19) :1841-1849
[8]   Bioactive ceramics: the effect of surface reactivity on bone formation and bone cell function [J].
Ducheyne, P ;
Qiu, Q .
BIOMATERIALS, 1999, 20 (23-24) :2287-2303
[9]   Short- and long-term animal studies with a plasma-sprayed calcium phosphate-coated implant [J].
Gottlander, M ;
Johansson, CB ;
Albrektsson, T .
CLINICAL ORAL IMPLANTS RESEARCH, 1997, 8 (05) :345-351
[10]   Characterization of sol-gel surfaces for biomedical applications [J].
Haddow, DB ;
James, PF ;
VanNoort, R .
JOURNAL OF MATERIALS SCIENCE-MATERIALS IN MEDICINE, 1996, 7 (05) :255-260