Impact of stent malapposition on intracoronary flow dynamics: An optical coherence tomography-based patient-specific study

被引:8
作者
Wei, Lingling [1 ]
Wang, Jiaqiu [2 ]
Chen, Qiang [1 ]
Li, Zhiyong [1 ,2 ]
机构
[1] Southeast Univ, Sch Biol Sci Med Engn, Biomech Lab, Nanjing 210096, PR, Peoples R China
[2] Queensland Univ Technol, Sch Mech, Med & Proc Engn, Brisbane, Qld 4001, Australia
基金
中国国家自然科学基金;
关键词
Percutaneous coronary intervention; Stent malapposition; Computational fluid dynamics (CFD); Coronary artery; Hemodynamics; SHEAR-STRESS; FLUID-DYNAMICS; THROMBOSIS; APPOSITION; HEMODYNAMICS; BIFURCATIONS; PREDICTORS; MECHANISMS; GEOMETRY; DESIGN;
D O I
10.1016/j.medengphy.2021.06.002
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Percutaneous coronary intervention with stent implantation has emerged as a popular approach to treat coronary artery stenosis. Stent malapposition (SM), also referred as incomplete stent apposition, could reduce stent tissue coverage and hence increase the risk of late stent thrombosis. The objective of this study was to investigate the impact of SM on intracoronary flow dynamics by combining optical coherence tomography (OCT) image-based model reconstruction and computational analysis. Firstly, a stenosed coronary artery model was reconstructed from OCT and angiography imaging data of a patient. Two structural analyses were carried out to simulate two types of coronary artery stent implantations: a fully-apposed (FA) case and a SM case. Then, based on the two deformed coronary geometries, two computational fluid dynamics (CFD) analyses were performed to evaluate the differences of hemodynamic metrics between the FA and the SM cases, including wall shear stress (WSS), time-averaged WSS (TWSS), oscillatory shear index (OSI), WSS gradient (WSSG), time-averaged WSSG (TWSSG), and relative residence time (RRT). The results indicated that maximum flow velocity was higher in the SM case than that of the FA case, due to the incomplete expansion of the stent and artery. Moreover, the SM case had a lower percentage of areas of adverse WSS (< 0.5 Pa) and RRT (> 10/Pa) but a higher percentage of areas of adverse OSI (> 0.1) and WSSG (> 5000 Pa/m). Specifically, the differences of OSI, WSSG, and RRT between the two cases were relatively small. It was suggested that SM might not be responsible for negative hemodynamic metrics which would further result in stent thrombosis on the basis of the present specific model.
引用
收藏
页码:26 / 32
页数:7
相关论文
共 44 条
[21]   Measurement of the uniaxial mechanical properties of healthy and atherosclerotic human coronary arteries [J].
Karimi, Alireza ;
Navidbakhsh, Mahdi ;
Shojaei, Ahmad ;
Faghihi, Shahab .
MATERIALS SCIENCE & ENGINEERING C-MATERIALS FOR BIOLOGICAL APPLICATIONS, 2013, 33 (05) :2550-2554
[22]   Geometry of the carotid bifurcation predicts its exposure to disturbed flow [J].
Lee, Sang-Wook ;
Antiga, Luca ;
Spence, J. David ;
Steinman, David A. .
STROKE, 2008, 39 (08) :2341-2347
[23]   Correlations Among Indicators of Disturbed Flow at the Normal Carotid Bifurcation [J].
Lee, Sang-Wook ;
Antiga, Luca ;
Steinman, David A. .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 2009, 131 (06)
[24]   Materials, fluid dynamics, and solid mechanics aspects of coronary artery stents: A state-of-the-art review [J].
Lewis, Gladius .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART B-APPLIED BIOMATERIALS, 2008, 86B (02) :569-590
[25]   TOWARD THE QUIESCENT CORONARY PLAQUE [J].
MACISAAC, AI ;
THOMAS, JD ;
TOPOL, EJ .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1993, 22 (04) :1228-1241
[26]   Hemodynamic shear stress and its role in atherosclerosis [J].
Malek, AM ;
Alper, SL ;
Izumo, S .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1999, 282 (21) :2035-2042
[27]   In Vivo Wall Shear Stress Distribution in the Carotid Artery Effect of Bifurcation Geometry, Internal Carotid Artery Stenosis, and Recanalization Therapy [J].
Markl, Michael ;
Wegent, Felix ;
Zech, Timo ;
Bauer, Simon ;
Strecker, Christoph ;
Schumacher, Martin ;
Weiller, Cornelius ;
Hennig, Juergen ;
Harloff, Andreas .
CIRCULATION-CARDIOVASCULAR IMAGING, 2010, 3 (06) :647-655
[28]   Patient-specific simulations of stenting procedures in coronary bifurcations: Two clinical cases [J].
Morlacchi, Stefano ;
Colleoni, Sebastian George ;
Cardenes, Ruben ;
Chiastra, Claudio ;
Luis Diez, Jose ;
Larrabide, Ignacio ;
Migliavacca, Francesco .
MEDICAL ENGINEERING & PHYSICS, 2013, 35 (09) :1272-1281
[29]   The stress-strain behavior of coronary stent struts is size dependent [J].
Murphy, BP ;
Savage, P ;
McHugh, PE ;
Quinn, DF .
ANNALS OF BIOMEDICAL ENGINEERING, 2003, 31 (06) :686-691
[30]  
Otsuka Fumiyuki, 2012, Thrombosis, V2012, P608593, DOI 10.1155/2012/608593