Infection Rates of Electrical Leads Used for Percutaneous Neurostimulation of the Peripheral Nervous System

被引:60
作者
Ilfeld, Brian M. [1 ]
Gabriel, Rodney A. [1 ]
Saulino, Michael F. [2 ,3 ]
Chae, John [4 ,5 ,6 ]
Peckham, P. Hunter [7 ]
Grant, Stuart A. [8 ]
Gilmore, Christopher A. [9 ,10 ,11 ]
Donohue, Michael C. [12 ]
deBock, Matthew G. [13 ]
Wongsarnpigoon, Amorn [13 ]
Boggs, Joseph W. [13 ]
机构
[1] Univ Calif San Diego, Dept Anesthesiol, San Diego, CA 92103 USA
[2] MossRehab, Elkins Pk, PA USA
[3] Sidney Kimmel Med Coll, Dept Rehabil Med, Philadelphia, PA USA
[4] Case Western Reserve Univ, Dept Phys Med & Rehabil, Cleveland, OH 44106 USA
[5] MetroHlth Syst, MetroHlth Rehabil Inst, Cleveland, OH USA
[6] Cleveland Funct Elect Stimulat Ctr, Cleveland, OH USA
[7] Case Western Reserve Univ, Biomed Engn & Orthopaed, Cleveland, OH 44106 USA
[8] Duke Univ, Med Ctr, Dept Anesthesiol, Durham, NC 27710 USA
[9] Wake Forest Univ, Baptist Med Ctr, Dept Anesthesiol, Winston Salem, NC 27109 USA
[10] Ctr Clin Res, Winston Salem, NC USA
[11] Carolinas Pain Inst, Winston Salem, NC USA
[12] Univ Southern Calif, Keck Sch Med, Dept Neurol, Los Angeles, CA USA
[13] SPR Therapeut LLC, Chapel Hill, NC USA
基金
美国国家卫生研究院;
关键词
neuromodulation; percutaneous peripheral nerve stimulation; peripheral nerve stimulator; helical lead; small-diameter open-coiled helical lead; postoperative pain; RANDOMIZED CLINICAL-TRIAL; HEMIPLEGIC SHOULDER PAIN; FECAL INCONTINENCE; INTRAMUSCULAR ELECTRODES; SACRAL NEUROMODULATION; LONG-TERM; STIMULATION ELECTRODES; CASE SERIES; FOLLOW-UP; MULTICENTER;
D O I
10.1111/papr.12523
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
BackgroundPercutaneous neurostimulation of the peripheral nervous system involves the insertion of a wire lead through an introducing needle to target a nerve/plexus or a motor point within a muscle. Electrical current may then be passed from an external generator through the skin via the lead for various therapeutic goals, including providing analgesia. With extended use of percutaneous leads sometimes greater than a month, infection is a concern. It was hypothesized that the infection rate of leads with a coiled design is lower than for leads with a noncoiled cylindrical design. MethodsThe literature was retrospectively reviewed for clinical studies of percutaneous neurostimulation of the peripheral nervous system of greater than 2 days that included explicit information on adverse events. The primary endpoint was the number of infections per 1,000 indwelling days. ResultsForty-three studies were identified that met inclusion criteria involving coiled (n = 21) and noncoiled (n = 25) leads (3 studies involved both). The risk of infection with noncoiled leads was estimated to be 25 times greater than with coiled leads (95% confidence interval [CI] 2 to 407, P = 0.006). The infection rates were estimated to be 0.03 (95% CI 0.01 to 0.13) infections per 1,000 indwelling days for coiled leads and 0.83 (95% CI 0.16 to 4.33) infections per 1,000 indwelling days for noncoiled leads (P = 0.006). ConclusionsPercutaneous leads used for neurostimulation of the peripheral nervous system have a much lower risk of infection with a coiled design compared with noncoiled leads: approximately 1 infection for every 30,000 vs. 1,200 indwelling days, respectively.
引用
收藏
页码:753 / 762
页数:10
相关论文
共 55 条
[1]  
Amend B, 2013, BJU INT, V111, P485, DOI 10.1111/j.1464-410X.2012.11263.x
[2]   Urodynamic changes and response rates in patients treated with permanent electrodes compared to conventional wire electrodes in the peripheral nerve evaluation test [J].
Bannowsky, Andreas ;
Wefer, Bjoern ;
Braun, Peter Martin ;
Juenemann, Klaus-Peter .
WORLD JOURNAL OF UROLOGY, 2008, 26 (06) :623-626
[3]   Infectious Risk of Continuous Peripheral Nerve Blocks [J].
Capdevila, Xavier ;
Bringuier, Sophie ;
Borgeat, Alain .
ANESTHESIOLOGY, 2009, 110 (01) :182-188
[4]   Sacral nerve root stimulation for lower urinary tract dysfunction: overcoming the problem of lead migration [J].
Carey, M ;
Fynes, M ;
Murray, C ;
Maher, C .
BJU INTERNATIONAL, 2001, 87 (01) :15-18
[5]   Single-Lead Percutaneous Peripheral Nerve Stimulation for the Treatment of Hemiplegic Shoulder Pain: A Case Series [J].
Chae, John ;
Wilson, Richard D. ;
Bennett, Maria E. ;
Lechman, Tina E. ;
Stager, Kathryn W. .
PAIN PRACTICE, 2013, 13 (01) :59-67
[6]   Intramuscular Electrical Stimulation for Upper Limb Recovery in Chronic Hemiparesis: An Exploratory Randomized Clinical Trial [J].
Chae, John ;
Harley, Mary Y. ;
Hisel, Terri Z. ;
Corrigan, Catherine M. ;
Demchak, Jeffrey A. ;
Wong, Yu-Tung ;
Fang, Zi-Ping .
NEUROREHABILITATION AND NEURAL REPAIR, 2009, 23 (06) :569-578
[7]  
Corey JM, 1990, TISSUE RESPONSE CORR
[8]  
Daly JJ, 2001, J REHABIL RES DEV, V38, P513
[9]   Bacterial colonization of stimulation electrode wires in patients undergoing temporary sacral nerve stimulation [J].
Dudding, T. ;
Vaizey, C. .
COLORECTAL DISEASE, 2010, 12 (02) :141-143
[10]   Economic evaluation of sacral nerve stimulation for faecal incontinence [J].
Dudding, T. C. ;
Lee, E. Meng ;
Faiz, O. ;
Pares, D. ;
Vaizey, C. J. ;
McGuire, A. ;
Kamm, M. A. .
BRITISH JOURNAL OF SURGERY, 2008, 95 (09) :1155-1163