International study on syncope of uncertain aetiology 3 (ISSUE 3): pacemaker therapy for patients with asystolic neurally-mediated syncope: rationale and study design

被引:43
作者
Brignole, M. [1 ]
Andresen, Dietrich [1 ]
Benditt, David [1 ]
Blanc, Jean Jacques [1 ]
Garcia-Civera, Roberto [1 ]
Khran, Andrew [1 ]
Menozzi, Carlo [1 ]
Moya, Angel [1 ]
Sutton, Richard [1 ]
Vardas, Panos [1 ]
Wieling, Wouter [1 ]
机构
[1] Osped Tigullio, Dept Cardiol, I-16033 Lavagna, Italy
来源
EUROPACE | 2007年 / 9卷 / 01期
关键词
syncope; neurally-mediated syncope; pacemaker therapy; electrocardiographic monitoring; implantable loop recorder;
D O I
10.1093/europace/eul135
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Aim To assess the effectiveness of pacing therapy for preventing syncope recurrence in patients with a high probability of cardio-inhibitory neurally-mediated syncope (NMS). Methods Study design: Multi-centre, prospective, double-blind, randomized placebo-controlled study. Inclusion criteria: Eligible patients are at least 40 years of age and have suffered, in the prior 2 years, >= 3 syncope episodes of suspected NMS (with the exception of carotid sinus syndrome), which is considered by the attending physician to have a severe clinical presentation requiring treatment initiation. Patients with positive and negative tilt testing are included. Exclusion criteria: Patients with one or more of the following are excluded: carotid sinus syndrome; suspected or definite heart disease and high likelihood of cardiac syncope; symptomatic orthostatic hypotension diagnosed by standing blood pressure measurement; loss of consciousness different from syncope (e.g. epilepsy, psychiatric, metabolic, drop-attack, TIA, intoxication, cataplexy); subclavian steal syndrome. Study protocol: Eligible patients receive an Implantable Loop Recorder and are followed tilt the first documented syncopal recurrence or a significant asystolic event. Those patients who have an asystolic pause (sinus arrest or AV block) > 6 s or a syncopal asystolic pause >= 3 s receive a dual-chamber pacemaker implantation and are randomized to active therapy (Pm ON) or to placebo therapy (Pm OFF). End-points: Primary end-point is the first syncope recurrence after pacemaker implant. Sample size and duration: A maximum of 710 patients are to be enrolled during an anticipated period of 2 years to allow randomization of 60 patients in the Pm ON arm and 60 in the Pm OFF arm (total 120).
引用
收藏
页码:25 / 30
页数:6
相关论文
共 13 条
  • [1] Early application of an implantable loop recorder allows effective specific therapy in patients with recurrent suspected neurally mediated syncope
    Brignole, M
    Sutton, R
    Menozzi, C
    Garcia-Civera, R
    Moya, A
    Wieling, W
    Andresen, D
    Benditt, DG
    Vardas, P
    [J]. EUROPEAN HEART JOURNAL, 2006, 27 (09) : 1085 - 1092
  • [2] Guidelines on management (diagnosis and treatment) of syncope - Update 2004 - Executive summary - The Task Force on syncope, European Society of Cardiology
    Brignole, M
    Alboni, P
    Benditt, DG
    Bergfeldt, L
    Blanc, JJ
    Thomsen, PEB
    van Dijk, JG
    Fitzpatrick, A
    Hohnloser, S
    Janousek, J
    Kapoor, W
    Kenny, RA
    Kulakowski, P
    Masotti, G
    Moya, A
    Raviele, A
    Sutton, R
    Theodorakis, G
    Ungar, A
    Wieling, W
    [J]. EUROPEAN HEART JOURNAL, 2004, 25 (22) : 2054 - 2072
  • [3] Guidelines on management (diagnosis and treatment) of syncope - Update 2004
    Brignole, M
    Alboni, P
    Benditt, DG
    Bergfeldt, L
    Blanc, JJ
    Thomsen, PEB
    van Dijk, JG
    Fitzpatrick, A
    Hohnloser, S
    Janousek, J
    Kapoor, W
    Kenny, RA
    Kulakowski, P
    Masotti, G
    Moya, A
    Raviele, A
    Sutton, R
    Theodorakis, G
    Ungar, A
    Wieling, W
    [J]. EUROPACE, 2004, 6 (06): : 467 - 537
  • [4] A new management of syncope: prospective systematic guideline-based evaluation of patients referred urgently to general hospitals
    Brignole, M
    Menozzi, C
    Bartoletti, A
    Giada, F
    Lagi, A
    Ungar, A
    Ponassi, I
    Mussi, C
    Maggi, R
    Re, G
    Furlan, R
    Rovelli, G
    Ponzi, P
    Scivales, A
    [J]. EUROPEAN HEART JOURNAL, 2006, 27 (01) : 76 - 82
  • [5] The usage and diagnostic yield of the implantable loop-recorder in detection of the mechanism of syncope and in guiding effective antiarrhythmic therapy in older people
    Brignole, M
    Menozzi, C
    Magi, R
    Solano, A
    Donateo, P
    Bottoni, N
    Lolli, G
    Quartieri, F
    Croci, F
    Oddone, D
    Puggioni, E
    [J]. EUROPACE, 2005, 7 (03): : 273 - 279
  • [6] Proposed electrocardiographic classification of spontaneous syncope documented by an implantable loop recorder
    Brignole, M
    Moya, A
    Menozzi, C
    Garcia-Civera, R
    Sutton, R
    [J]. EUROPACE, 2005, 7 (01): : 14 - 18
  • [7] New classification of haemodynamics of vasovagal syncope: beyond the VASIS classification - Analysis of the pre-syncopal phase of the tilt test without and with nitroglycerin challenge
    Brignole, M
    Menozzi, C
    Del Rosso, A
    Costa, S
    Gaggioli, G
    Bottoni, N
    Bartoli, P
    Sutton, R
    [J]. EUROPACE, 2000, 2 (01): : 66 - 76
  • [8] Pacemaker therapy for prevention of syncope in patients with recurrent severe vasovagal syncope - Second Vasovagal Pacemaker Study (VPS II): A randomized trial
    Connolly, SJ
    Sheldon, R
    Thorpe, KE
    Roberts, RS
    Ellenbogen, KA
    Wilkoff, BL
    Morillo, C
    Gent, M
    [J]. JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 2003, 289 (17): : 2224 - 2229
  • [9] Lack of efficacy of atenolol for the prevention of neurally mediated syncope in a highly symptomatic population:: A prospective, double-blind, randomized and placebo-controlled study
    Madrid, AH
    Ortega, J
    Rebollo, JG
    Manzano, JG
    Segovia, JG
    Sánchez, A
    Peña, G
    Moro, C
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2001, 37 (02) : 554 - 559
  • [10] Effect of etilefrine in preventing syncopal recurrence in patients with vasovagal syncope - A double-blind, randomized, placebo-controlled trial
    Raviele, A
    Brignole, M
    Sutton, R
    Alboni, P
    Giani, P
    Menozzi, C
    Moya, A