A contrast-detail comparison of computed mammotomography and digital mammography

被引:0
作者
McKinley, Randolph L. [1 ,2 ]
Tornai, Martin P. [1 ,2 ,4 ]
Floyd, Carey E. [1 ,2 ]
Samei, Ehsan [1 ,2 ,3 ,4 ]
机构
[1] Duke Univ, Med Ctr, Dept Radiol, Durham, NC 27710 USA
[2] Duke Univ, Dept Biomed Engn, Durham, NC 27706 USA
[3] Duke Univ, Dept Phys, Durham, NC 27706 USA
[4] Duke Univ, Med Phys Grad Program, Durham, NC USA
来源
MEDICAL IMAGING 2007: PHYSICS OF MEDICAL IMAGING, PTS 1-3 | 2007年 / 6510卷
关键词
X-ray computed mammotomography; contrast-detail; breast imaging; cone-beam; CT; FFDM;
D O I
10.1117/12.713032
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
We use a contrast-detail observer study to compare performance of a novel 3D computed mammotomography (CmT) system with a commercially developed full-field digital mammography (FFDM) system. A contrast-detail phantom comprised of uniform acrylic spheres of various diameters was developed and placed in a variety of mediums including uniform water (simulating low contrast lesions within a uniform background), water and acrylic yarn (simulating low contrast lesions, with over/under-lying structure), oil only (simulating higher contrast lesions in a uniform background), and oil and acrylic yarn (simulating higher contrast lesions with over/under-lying structure). For CmT, the phantom was placed in a 14.6 cm diameter uncompressed breast phantom and projections acquired using a simple circular orbit, W-target tube, 60 kVp tube potential, 0.05 cm Ce filtration, 4 mAs per projection, and a CSI(T1) digital x-ray detector. Reconstructions used an iterative OSTR algorithm. For FFDM, the phantom was placed in a 5.3-cm-thick compressed breast phantom. Single CC-view mammograms were acquired using a clinical W-target tube with 50 um Rh filtration, 28 kVp, photo-timed mAs per our clinical mammography operation, and a Selenium-based flat-panel detector (Mammomat Novation, Siemens). Six observers evaluated the images in terms of the number of detectable spheres. FFDM performed significantly better for the low contrast lesions in uniform water background (p < 0.05). However, CmT performed significantly better for all other cases. (p < 0.05). Results indicate that CmT shows significant advantage in soft tissue detection over FFDM in otherwise low contrast dense breasts.
引用
收藏
页数:10
相关论文
共 14 条
  • [1] A comprehensive analysis of DgNCT coefficients for pendant-geometry cone-beam breast computed tomography
    Boone, JM
    Shah, N
    Nelson, TR
    [J]. MEDICAL PHYSICS, 2004, 31 (02) : 226 - 235
  • [2] BOONE JM, 2004, RSNA CATEGORICAL COU, P165
  • [3] Cone-beam volume CT breast imaging: Feasibility study
    Chen, B
    Ning, R
    [J]. MEDICAL PHYSICS, 2002, 29 (05) : 755 - 770
  • [4] CONTRAST DETAIL ASSESSMENT OF COMPUTED-TOMOGRAPHY SCANNERS
    FAULKNER, K
    MOORES, BM
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1986, 31 (09) : 993 - 1003
  • [5] A computer simulation study comparing lesion detection accuracy with digital mammography, breast tomosynthesis, and cone-beam CT breast imaging
    Gong, X
    Glick, SJ
    Liu, B
    Vedula, AA
    Thacker, S
    [J]. MEDICAL PHYSICS, 2006, 33 (04) : 1041 - 1052
  • [6] Progress in the development of a dedicated breast CT scanner
    Kwan, A
    Shah, N
    Burkett, G
    Seibert, JA
    Lindfors, KK
    Nelson, TR
    Boone, JM
    [J]. MEDICAL IMAGING 2004: PHYSICS OF MEDICAL IMAGING, PTS 1 AND 2, 2004, 5368 : 304 - 310
  • [7] SPECIFYING AND EVALUATING THE PERFORMANCE OF COMPUTED-TOMOGRAPHY (CT) SCANNERS
    MCCULLOUGH, EC
    [J]. MEDICAL PHYSICS, 1980, 7 (04) : 291 - 296
  • [8] Investigation of cone-beam acquisitions implemented using a novel dedicated mammotomography system with unique arbitrary orbit capability
    McKinley, RL
    Brzymialkiewicz, CN
    Madhav, P
    Tornai, MP
    [J]. Medical Imaging 2005: Physics of Medical Imaging, Pts 1 and 2, 2005, 5745 : 609 - 617
  • [9] Simulation study of a quasi-monochromatic beam for x-ray computed mammotomography
    McKinley, RL
    Tornai, MP
    Samei, E
    Bradshaw, ML
    [J]. MEDICAL PHYSICS, 2004, 31 (04) : 800 - 813
  • [10] A new test phantom with different breast tissue compositions for image quality assessment in conventional and digital mammography
    Pachoud, M
    Lepori, D
    Valley, JF
    Verdun, FR
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2004, 49 (23) : 5267 - 5281