Development of a temperature-controlled phantom for magnetic resonance quality assurance of diffusion, dynamic, and relaxometry measurements

被引:22
作者
Jerome, Neil P. [1 ,2 ]
Papoutsaki, Marianthi-Vasiliki [1 ,2 ,3 ]
Orton, Matthew R. [1 ]
Parkes, Harold G. [1 ]
Winfield, Jessica M. [1 ]
Boss, Michael A. [4 ]
Leach, Martin O. [1 ,2 ]
deSouza, Nandita M. [1 ,2 ,3 ]
Collins, David J. [1 ,2 ]
机构
[1] Canc Res UK Canc Imaging Ctr, Div Radiotherapy & Imaging, Inst Canc Res, 123 Old Brompton Rd, London SM2 5NG, England
[2] Royal Marsden Hosp, 123 Old Brompton Rd, London SM2 5NG, England
[3] Quantitat Imaging Canc Connecting Cellular Proc T, London, England
[4] NIST, 325 Broadway, Boulder, CO 80305 USA
基金
英国工程与自然科学研究理事会;
关键词
quality assurance; diffusion weighted imaging; quantitative imaging; DCE-MRI; IN-VIVO; RELAXATION; REPRODUCIBILITY; REPEATABILITY; BEVACIZUMAB; AGREEMENT; TISSUES; TESLA; T-1;
D O I
10.1118/1.4948997
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Diffusion-weighted (DW) and dynamic contrast-enhanced magnetic resonance imaging (MRI) are increasingly applied for the assessment of functional tissue biomarkers for diagnosis, lesion characterization, or for monitoring of treatment response. However, these techniques are vulnerable to the influence of various factors, so there is a necessity for a standardized MR quality assurance procedure utilizing a phantom to facilitate the reliable estimation of repeatability of these quantitative biomarkers arising from technical factors (e.g., B-1 variation) affecting acquisition on scanners of different vendors and field strengths. The purpose of this study is to present a novel phantom designed for use in quality assurance for multicenter trials, and the associated repeatability measurements of functional and quantitative imaging protocols across different MR vendors and field strengths. Methods: A cylindrical acrylic phantom was manufactured containing 7 vials of polyvinylpyrrolidone (PVP) solutions of different concentrations, ranging from 0% (distilled water) to 25% w/w, to create a range of different MR contrast parameters. Temperature control was achieved by equilibration with ice-water. Repeated MR imaging measurements of the phantom were performed on four clinical scanners (two at 1.5 T, two at 3.0 T; two vendors) using the same scanning protocol to assess the long-term and short-term repeatability. The scanning protocol consisted of DW measurements, inversion recovery (IR) T-1 measurements, multiecho T-2 measurement, and dynamic T-1-weighted sequence allowing multiple variable flip angle (VFA) estimation of T-1 values over time. For each measurement, the corresponding calculated parameter maps were produced. On each calculated map, regions of interest (ROIs) were drawn within each vial and the median value of these voxels was assessed. For the dynamic data, the autocorrelation function and their variance were calculated; for the assessment of the repeatability, the coefficients of variation (CoV) were calculated. Results: For both field strengths across the available vendors, the apparent diffusion coefficient (ADC) at 0 degrees C ranged from (1.12 +/- 0.01) x10(-3) mm(2)/s for pure water to (0.48 +/- 0.02)x10(-3) mm(2)/s for the 25% w/w PVP concentration, presenting a minor variability between the vendors and the field strengths. T-2 and IR-T-1 relaxation time results demonstrated variability between the field strengths and the vendors across the different acquisitions. Moreover, the T-1 values derived from the VFA method exhibited a large variation compared with the IR-T-1 values across all the scanners for all repeated measurements, although the calculation of the standard deviation of the VFA-T-1 estimate across each ROI and the autocorrelation showed a stability of the signal for three scanners, with autocorrelation of the signal over the dynamic series revealing a periodic variation in one scanner. Finally, the ADC, the T-2, and the IR-T-1 values exhibited an excellent repeatability across the scanners, whereas for the dynamic data, the CoVs were higher. Conclusions: The combination of a novel PVP phantom, with multiple compartments to give a physiologically relevant range of ADC and T-1 values, together with ice-water as a temperature-controlled medium, allows reliable quality assurance measurements that can be used to measure agreement between MRI scanners, critical in multicenter functional and quantitative imaging studies. (C) 2016 Author(s). All article content, except where otherwise noted, is licensed under a Creative Commons Attribution (CC BY) license
引用
收藏
页码:2998 / 3007
页数:10
相关论文
共 32 条
[1]   Diffusion-Weighted Magnetic Resonance Imaging as a Cancer Biomarker: Consensus and Recommendations [J].
不详 .
NEOPLASIA, 2009, 11 (02) :102-125
[2]  
Bland JM, 1999, STAT METHODS MED RES, V8, P135, DOI 10.1177/096228029900800204
[3]  
Boss M. A., 2014, P 22 INT SOC MAGN RE
[4]   Changes in Tumor Blood Flow as Measured by Dynamic Contrast-Enhanced Magnetic Resonance Imaging (DCE-MRI) May Predict Activity of Single Agent Bevacizumab in Recurrent Epithelial Ovarian (EOC) and Primary Peritoneal Cancer (PPC) Patients: An exploratory analysis of a Gynecologic Oncology Group Phase II study [J].
Chase, Dana M. ;
Sill, Michael W. ;
Monk, Bradley J. ;
Chambers, Mark D. ;
Darcy, Kathleen M. ;
Han, Ernest S. ;
Buening, Barbara J. ;
Sorosky, Joel I. ;
Fruehauf, John P. ;
Burger, Robert A. .
GYNECOLOGIC ONCOLOGY, 2012, 126 (03) :375-380
[5]   Rapid high-resolution T1 mapping by variable flip angles:: Accurate and precise measurements in the presence of radiofrequency field inhomogeneity [J].
Cheng, HLM ;
Wright, GA .
MAGNETIC RESONANCE IN MEDICINE, 2006, 55 (03) :566-574
[6]   Agreement and Reproducibility of Apparent Diffusion Coefficient Measurements of Dual-b-Value and Multi-b-Value Diffusion-Weighted Magnetic Resonance Imaging at 1.5 Tesla in Phantom and in Soft Tissues of the Abdomen [J].
Corona-Villalobos, Celia Pamela ;
Pan, Li ;
Halappa, Vivek Gowdra ;
Bonekamp, Susanne ;
Lorenz, Christine H. ;
Eng, John ;
Kamel, Ihab R. .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2013, 37 (01) :46-51
[7]   MR imaging relaxation times of abdominal and pelvic tissues measured in vivo at 3.0 T: Preliminary results [J].
de Bazelaire, CMJ ;
Duhamel, GD ;
Rofsky, NM ;
Alsop, DC .
RADIOLOGY, 2004, 230 (03) :652-659
[8]   Value of DCE-MRI and FDG-PET/CT in the prediction of response to preoperative chemotherapy with bevacizumab for colorectal liver metastases [J].
De Bruyne, S. ;
Van Damme, N. ;
Smeets, P. ;
Ferdinande, L. ;
Ceelen, W. ;
Mertens, J. ;
Van de Wiele, C. ;
Troisi, R. ;
Libbrecht, L. ;
Laurent, S. ;
Geboes, K. ;
Peeters, M. .
BRITISH JOURNAL OF CANCER, 2012, 106 (12) :1926-1933
[9]   Developing a quality control protocol for diffusion imaging on a clinical MRI system [J].
Delakis, I ;
Moore, EM ;
Leach, MO ;
De Wilde, JP .
PHYSICS IN MEDICINE AND BIOLOGY, 2004, 49 (08) :1409-1422
[10]   Rapid combined T1 and T2 mapping using gradient recalled acquisition in the steady state [J].
Deoni, SCL ;
Rutt, BK ;
Peters, TM .
MAGNETIC RESONANCE IN MEDICINE, 2003, 49 (03) :515-526