Axial Magnetic Bearing Development for the BiVACOR Rotary BiVAD/TAH

被引:31
作者
Greatrex, Nicholas A. [1 ]
Timms, Daniel L. [2 ]
Kurita, Nobuyuki [3 ]
Palmer, Edward W. [1 ]
Masuzawa, Toru [4 ]
机构
[1] Queensland Univ Technol, Inst Hlth & Biomed Innovat, Brisbane, Qld 4001, Australia
[2] Rhein Westfal TH Aachen, Helmholtz Inst, D-52074 Aachen, Germany
[3] Gunma Univ, Dept Elect Engn, Gunma 3718510, Japan
[4] Ibaraki Univ, Dept Mech Engn, Ibaraki 3108512, Japan
关键词
Artificial biological organs; blood pumps; electromagnetic forces; magnetic levitation; CENTRIFUGAL BLOOD PUMP;
D O I
10.1109/TBME.2009.2033389
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A suspension system for the BiVACOR biventricular assist device (BiVAD) has been developed and tested. The device features two semi-open centrifugal impellers mounted on a common rotating hub. Flow balancing is achieved through the movement of the rotor in the axial direction. The rotor is suspended in the pump casings by an active magnetic suspension system in the axial direction and a passive hydrodynamic bearing in the radial direction. This paper investigates the axial movement capacity of the magnetic bearing system and the power consumption at various operating points. The force capacity of the passive hydrodynamic bearing is investigated using a viscous glycerol solution. Axial rotor movement in the range of +/- 0.15 mm is confirmed and power consumption is under 15.5 W. The journal bearing is shown to stabilize the rotor in the radial direction at the required operating speed. Magnetic levitation is a viable suspension technique for the impeller of an artificial heart to improve device lifetime and reduce blood damage.
引用
收藏
页码:714 / 721
页数:8
相关论文
共 16 条
[1]   CENTRIFUGAL BLOOD PUMP WITH A MAGNETICALLY SUSPENDED IMPELLER [J].
AKAMATSU, T ;
NAKAZEKI, T ;
ITOH, H .
ARTIFICIAL ORGANS, 1992, 16 (03) :305-308
[2]   Which patient, which pump? [J].
DiGiorgi, PL ;
Rao, V ;
Naka, Y ;
Oz, MC .
JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2003, 22 (03) :221-235
[3]   Comparison of pulsatile with nonpulsatile mechanical support in a porcine model of profound cardiogenic shock [J].
Drakos, SG ;
Charitos, CE ;
Ntalianis, A ;
Terrovitis, JV ;
Siafakas, KX ;
Dolou, P ;
Pierrakos, C ;
Charitos, E ;
Karelas, J ;
Nanas, JN .
ASAIO JOURNAL, 2005, 51 (01) :26-29
[4]   Magnetically suspended centrifugal blood pump with a radial magnetic driver [J].
Hoshi, H ;
Katakoa, K ;
Ohuchi, K ;
Asama, J ;
Shinshi, T ;
Shimokohbe, A ;
Takatani, S .
ASAIO JOURNAL, 2005, 51 (01) :60-64
[5]  
Japikse D., 1997, Centrifugal Pump Design and Performance
[6]   SYSTEM CONSIDERATIONS FAVORING ROTARY ARTIFICIAL HEARTS WITH BLOOD-IMMERSED BEARINGS [J].
JARVIK, RK .
ARTIFICIAL ORGANS, 1995, 19 (07) :565-570
[7]  
Karassik Igor J., 2000, Pump Handbook, Vthird
[8]   Artificial hearts [J].
Maslen, EH ;
Bearnson, GB ;
Allaire, PE ;
Flack, RD ;
Baloh, M ;
Hilton, E ;
Noh, MD ;
Olson, DB ;
Khanwilkar, PS ;
Long, JD .
PROCEEDINGS OF THE 1997 IEEE INTERNATIONAL CONFERENCE ON CONTROL APPLICATIONS, 1997, :204-209
[9]   Magnetically suspended centrifugal blood pump with an axially levitated motor [J].
Masuzawa, T ;
Ezoe, S ;
Kato, T ;
Okada, Y .
ARTIFICIAL ORGANS, 2003, 27 (07) :631-638
[10]   Rotary blood pumps: A new horizon [J].
Olsen, DB .
ARTIFICIAL ORGANS, 1999, 23 (08) :695-696