A device for real-time live-cell microscopy during dynamic dual-modal mechanostimulation

被引:1
|
作者
Lorusso, D. [1 ,2 ,3 ]
Nikolov, H. N. [1 ]
Chmiel, T. [1 ]
Beach, R. J. [2 ,3 ]
Sims, S. M. [2 ,3 ]
Dixon, S. J. [2 ,3 ]
Holdsworth, D. W. [1 ,3 ]
机构
[1] Univ Western Ontario, Imaging Res Labs, Robarts Res Inst, London, ON, Canada
[2] Univ Western Ontario, Dept Physiol & Pharmacol, Schulich Sch Med & Dent, London, ON, Canada
[3] Univ Western Ontario, Bone & Joint Inst, London, ON, Canada
来源
MEDICAL IMAGING 2017: BIOMEDICAL APPLICATIONS IN MOLECULAR, STRUCTURAL, AND FUNCTIONAL IMAGING | 2017年 / 10137卷
基金
加拿大健康研究院;
关键词
mechanotransduction; vibration; fluid shear stress; mechanostimulation; microscopy; live-cell calcium imaging; fluorescence microscopy; HIGH-FREQUENCY; SHEAR-STRESS; MECHANICAL SIGNALS; HAIR-CELLS; BONE; MECHANOTRANSDUCTION; OSTEOPOROSIS; EPIDEMIOLOGY; VIBRATIONS; EXPOSURE;
D O I
10.1117/12.2253904
中图分类号
TP18 [人工智能理论];
学科分类号
081104 ; 0812 ; 0835 ; 1405 ;
摘要
Mechanotransduction - the process by which cells sense and respond to mechanical stimuli - is essential for several physiological processes including skeletal homeostasis. Mammalian cells are thought to be sensitive to different modes of mechanical stimuli, including vibration and fluid shear. To better understand the mechanisms underlying the early stages of mechanotransduction, we describe the development of devices for mechanostimulation (by vibration and fluid shear) of live cells that can be integrated with real-time optical microscopy. The integrated system can deliver up to 3 Pa of fluid shear simultaneous with high-frequency sinusoidal vibrations up to 1 g. Stimuli can be applied simultaneously or independently to cells during real-time microscopic imaging. A custom microfluidic chamber was prepared from polydimethylsiloxane on a glass-bottom cell culture dish. Fluid flow was applied with a syringe pump to induce shear stress. This device is compatible with a custom-designed motion control vibration system. A voice coil actuates the system that is suspended on linear air bushings. Accelerations produced by the system were monitored with an on-board accelerometer. Displacement was validated optically using particle tracking digital high-speed imaging (1200 frames per second). During operation at nominally 45 Hz and 0.3 g, displacements were observed to be within 3.56% of the expected value. MC3T3-E1 osteoblast like cells were seeded into the microfluidic device and loaded with the calcium sensitive fluorescent probe fura-2, then mounted onto the dual-modal mechanostimulation platform. Cells were then imaged and monitored for fluorescence emission. In summary, we have developed a system to deliver physiologically relevant vibrations and fluid shear to live cells during real-time imaging and photometry. Monitoring the behavior of live cells loaded with appropriate fluorescent probes will enable characterization of the signals activated during the initial stages of mechanotransduction.
引用
收藏
页数:7
相关论文
共 50 条
  • [1] Real-time dual-modal vein imaging system
    Mela, Christopher A.
    Lemmer, David P.
    Bao, Forrest Sheng
    Papay, Francis
    Hicks, Tyler
    Liu, Yang
    INTERNATIONAL JOURNAL OF COMPUTER ASSISTED RADIOLOGY AND SURGERY, 2019, 14 (02) : 203 - 213
  • [2] Real-time dual-modal vein imaging system
    Christopher A. Mela
    David P. Lemmer
    Forrest Sheng Bao
    Francis Papay
    Tyler Hicks
    Yang Liu
    International Journal of Computer Assisted Radiology and Surgery, 2019, 14 : 203 - 213
  • [3] Real-time dual-modal photoacoustic and fluorescence small animal imaging
    Sun, Yu
    Wang, Yibing
    Li, Wenzhao
    Li, Changhui
    PHOTOACOUSTICS, 2024, 36
  • [4] Real-Time Pedestrian Detection Based on Dual-Modal Relevant Image Fusion
    Bi Chengcheng
    Huang Miaohua
    Liu Ruoying
    Wang Liangzi
    LASER & OPTOELECTRONICS PROGRESS, 2024, 61 (08)
  • [5] Live-cell fluorescent microscopy platforms for real-time monitoring of polyplex-cell interaction: Basic guidelines
    Parhamifar, Ladan
    Wu, LinPing
    Andersen, Helene
    Moghimi, S. Moein
    METHODS, 2014, 68 (02) : 300 - 307
  • [6] A live-cell assay for the real-time assessment of extracellular ATP levels
    Niles, Andrew L.
    Cali, James J.
    Lazar, Dan F.
    ANALYTICAL BIOCHEMISTRY, 2021, 628
  • [7] Real-Time Runway Detection Using Dual-Modal Fusion of Visible and Infrared Data
    Yang, Lichun
    Wu, Jianghao
    Li, Hongguang
    Liu, Chunlei
    Wei, Shize
    REMOTE SENSING, 2025, 17 (04)
  • [8] Polymer live-cell array for real-time kinetic imaging of immune cells
    Zurgil, Naomi
    Afrimzon, Elena
    Deutsch, Assaf
    Namer, Yaniv
    Shafran, Yana
    Sobolev, Maria
    Tauber, Yishay
    Ravid-Hermesh, Orit
    Deutsch, Mordechai
    BIOMATERIALS, 2010, 31 (18) : 5022 - 5029
  • [9] Real-Time Segmentation of Artificial Targets Using a Dual-Modal Efficient Attention Fusion Network
    Shen, Ying
    Liu, Xiancai
    Zhang, Shuo
    Xu, Yixuan
    Zeng, Dawei
    Wang, Shu
    Huang, Feng
    REMOTE SENSING, 2023, 15 (18)
  • [10] Real-time visualization and quantification of neutrophil activation and function using live-cell analysis
    Bevan, Nicola
    Lovell, Gillan
    O'Clair, Belinda
    Kelsey, Lauren
    Szybut, Clare
    Campwala, Hinnah
    Jackson, Tim
    Holtz, Nevine
    Endsley, Eric
    Dale, Tim
    Trezise, Del
    JOURNAL OF IMMUNOLOGY, 2019, 202 (01):