Prion disease and recommended procedures for flexible endoscope reprocessing - a review of policies worldwide and proposal for a simplified approach

被引:13
作者
Kampf, G. [1 ]
Jung, M. [2 ]
Suchomel, M. [3 ]
Saliou, P. [4 ]
Griffiths, H. [5 ]
Vos, C. [6 ]
机构
[1] Univ Med Greifswald, Inst Hyg & Environm Med, Ferdinand Sauerbruch Str, D-17475 Greifswald, Germany
[2] Univ Hosp Frankfurt, Med Dept 1, Endoscopy, Frankfurt, Germany
[3] Med Univ Vienna, Inst Hyg & Appl Immunol, Vienna, Austria
[4] Brest Teaching Hosp, Infect Control Unit, Brest, France
[5] Brecon War Mem Hosp, Brecon, Powys, Wales
[6] Erasmus MC, Dept Med Microbiol & Infect Dis, Rotterdam, Netherlands
关键词
Prion disease; CJD; vCD; Flexible endoscope; Reprocessing; CREUTZFELDT-JAKOB-DISEASE; STAINLESS-STEEL; INACTIVATION; DECONTAMINATION; DISINFECTION; STERILIZATION; TRANSMISSION; INFECTIVITY; EPIDEMIOLOGY; PROTEIN;
D O I
10.1016/j.jhin.2019.08.005
中图分类号
R1 [预防医学、卫生学];
学科分类号
1004 ; 120402 ;
摘要
Several guidelines recommend specific treatments for endoscopes, procedures of quarantine for endoscopes, or additional treatments for the endoscope washer disinfector (EWD) in suspected or confirmed cases of Creutzfeldt-Jakob disease (CJD) or variant CJD (vCJD) but vary in many details. This study therefore reviewed guidelines on reprocessing flexible endoscopes after use in patients with suspected or confirmed prion disease. In addition, a literature search was performed in Medline on prion, CJD, vCJD, chemical inactivation, transmission healthcare, epidemiology healthcare, concentration tissue human and endoscope. Thus far, no case of CJD or vCJD transmitted by flexible endoscope has been reported. In animals it has been shown that oral uptake of 0.1-5 g of bovine spongiform encephalopathy (BSE)-infected brain homogenate is necessary for transmission. The maximum prion concentration in other tissues (e.g., terminal ileum) is at least 100-fold lower. Automated cleaning of endoscopes alone results in very low total residual protein <= 5.6 mg per duodenoscopes. Recommendations vary between countries, sometimes with additional cleaning, use of alkaline cleaners, no use of cleaners with fixative properties, use of disinfectants without fixative properties or single-use disinfectants. Sodium hydroxide (1 M) and sodium hypochlorite (10,000 and 25,000 mg/L) are very effective in preventing transmission via contaminated wires implanted into animal brains, but their relevance for endoscopes is questionable. Based on circumstantial evidence, it is proposed to consider validated reprocessing as appropriate in the case of delayed suspected prion disease when immediate bedside cleaning, routine use of alkaline cleaners, no fixative agents anywhere prior to disinfection and single use brushes and cleaning solutions can be assured. (C) 2019 The Healthcare Infection Society. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:92 / 110
页数:19
相关论文
共 51 条
  • [1] Advisory Committee on Dangerous Pathogens' Transmissible Spongiform Encephalopathy (ACDP TSE). Subgroup, 2015, MIN TRANSM SPONG ENC
  • [2] EVOTECH® endoscope cleaner and reprocessor (ECR) simulated-use and clinical-use evaluation of cleaning efficacy
    Alfa, Michelle J.
    DeGagne, Pat
    Olson, Nancy
    Fatima, Iram
    [J]. BMC INFECTIOUS DISEASES, 2010, 10
  • [3] [Anonymous], 2012, BUNDESGESUNDHEITSBLA, V55, P1244, DOI 10.1007/s00103-012-1548-6
  • [4] [Anonymous], 2011, INF PREV CONTR GUID
  • [5] Arbeitskreis CJK des BMG, 2016, RICHTL SCHUTZ UB CRE
  • [6] Variant Creutzfeldt-Jakob disease (vCJD) and gastrointestinal endoscopy
    Axon, ATR
    Beilenhoff, U
    Bramble, MG
    Ghosh, S
    Kruse, A
    McDonnell, GE
    Neumann, C
    Rey, JF
    Spencer, K
    [J]. ENDOSCOPY, 2001, 33 (12) : 1070 - 1078
  • [7] Activity of an alkaline 'cleaner' in the inactivation of the scrapie agent
    Baier, M
    Schwarz, A
    Mielke, M
    [J]. JOURNAL OF HOSPITAL INFECTION, 2004, 57 (01) : 80 - 84
  • [8] Prion diseases: infectious and lethal doses following oral challenge
    Baier, M
    Norley, S
    Schultz, J
    Burwinkel, M
    Schwarz, A
    Riemer, C
    [J]. JOURNAL OF GENERAL VIROLOGY, 2003, 84 : 1927 - 1929
  • [9] Infection control during GI endoscopy
    Banerjee, Subhas
    Shen, Bo
    Nelson, Douglas B.
    Lichtenstein, David R.
    Baron, Todd H.
    Anderson, Michelle A.
    Dominitz, Jason A.
    Gan, S. Ian
    Harrison, M. Edwyn
    Ikenberry, Steven O.
    Jagannath, Sanjay B.
    Fanelli, Robert D.
    Lee, Ken
    van Guilder, Trina
    Stewart, Leslie E.
    [J]. GASTROINTESTINAL ENDOSCOPY, 2008, 67 (06) : 781 - 790
  • [10] Fast, broad-range disinfection of bacteria, fungi, viruses and prions
    Beekes, Michael
    Lemmer, Karin
    Thomzig, Achim
    Joncic, Marion
    Tintelnot, Kathrin
    Mielke, Martin
    [J]. JOURNAL OF GENERAL VIROLOGY, 2010, 91 : 580 - 589