Influence of plate material and screw design on stiffness and ultimate load of locked plating in osteoporotic proximal humeral fractures

被引:51
作者
Katthagen, Jan Christoph [1 ]
Schwarze, Michael [2 ]
Warnhoff, Mara [1 ]
Voigt, Christine [1 ]
Hurschler, Christof [2 ]
Lill, Helmut [1 ]
机构
[1] Diakoniekrankenhaus Friederikenstift GGmbH, Dept Trauma & Reconstruct Surg, Humboldtstr 5, D-30169 Hannover, Germany
[2] MHH, Lab Biomech & Biomat, Anna von Borries Str 1-7, D-30625 Hannover, Germany
来源
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED | 2016年 / 47卷 / 03期
关键词
Carbon fibre; PEEK; Proximal humerus fracture; Locked plating; Screw; INTRAMEDULLARY FIBULAR ALLOGRAFT; BONE-MINERAL DENSITY; LOCKING PLATE; INTERNAL-FIXATION; JOINT LOADS; IMPLANTS; STABILITY; AUGMENTATION; PERFORMANCE; SUPPORT;
D O I
10.1016/j.injury.2016.01.004
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Introduction: The main purpose was to compare the biomechanical properties of a carbon-fibre reinforced polyetheretherketone (CF-PEEK) composite locking platewith pre-existing data of a titaniumalloy plate when used for fixation of an unstable 2-part fracture of the surgical neck of the humerus. The secondary purpose was to compare the mechanical behaviour of locking bolts and conventional locking cancellous screws. Methods: 7 pairs of fresh frozen human humeri were allocated to two equal groups. All specimens were fixed with the CF-PEEK plate. Cancellous screws (PEEK/screw) were compared to locking bolts (PEEK/bolt) for humeral head fixation. Stiffness, fracture gap deflection and ultimate load as well as load before screw perforation of the articular surface were assessed. Results were compared between groups and with pre-existing biomechanical data of a titanium-alloy plate. Results: The CF-PEEK plate featured significantly lower stiffness compared to the titanium-alloy plate (P < 0.001). In ultimate load testing, 6 out of 14 CF-PEEK plates failed due to irreversible deformation and cracking. No significant difference was observed between results of groups PEEK/screw and PEEK/bolt (P > 0.05). Discussion: The CF-PEEK plate has more elastic properties and significantly increases movement at the fracture site of an unstable proximal humeral fracture model compared to the commonly used titaniumalloy plate. The screw design however does neither affect the constructs primary mechanical behaviour in the constellation tested nor the load before screw perforation. (C) 2016 Elsevier Ltd. All rights reserved.
引用
收藏
页码:617 / 624
页数:8
相关论文
共 30 条
[1]  
[Anonymous], KNEE SURG SPORTS TRA
[2]   The biomechanical performance of locking plate fixation with intramedullary fibular strut graft augmentation in the treatment of unstable fractures of the proximal humerus [J].
Bae, J. -H. ;
Oh, J. -K. ;
Chon, C. -S. ;
Oh, C. -W. ;
Hwang, J. -H. ;
Yoon, Y. -C. .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2011, 93B (07) :937-941
[3]   In vivo glenohumeral contact forces -: Measurements in the first patient 7 months postoperatively [J].
Bergmann, G. ;
Graichen, F. ;
Bender, A. ;
Kaab, M. ;
Rohlmann, A. ;
Westerhoff, P. .
JOURNAL OF BIOMECHANICS, 2007, 40 (10) :2139-2149
[4]   In vivo gleno-humeral joint loads during forward flexion and abduction [J].
Bergmann, G. ;
Graichen, F. ;
Bender, A. ;
Rohlmann, A. ;
Halder, A. ;
Beier, A. ;
Westerhoff, P. .
JOURNAL OF BIOMECHANICS, 2011, 44 (08) :1543-1552
[5]   Proximal humeral fracture fixation: locking plate construct ± intramedullary fibular allograft [J].
Chow, Roxanne M. ;
Begum, Farhana ;
Beaupre, Lauren A. ;
Carey, Jason P. ;
Adeeb, Samer ;
Bouliane, Martin J. .
JOURNAL OF SHOULDER AND ELBOW SURGERY, 2012, 21 (07) :894-901
[6]   Influence of size and stability of the osteotomy gap on the success of fracture healing [J].
Claes, L ;
Augat, P ;
Suger, G ;
Wilke, HJ .
JOURNAL OF ORTHOPAEDIC RESEARCH, 1997, 15 (04) :577-584
[7]   Fractures and fracture-dislocations of the proximal humerus: A retrospective analysis of 82 cases treated with the Philos® locking plate [J].
Erasmo, Rocco ;
Guerra, Giovanni ;
Guerra, Luigi .
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2014, 45 :S43-S48
[8]   The Position and Number of Screws Influence Screw Perforation of the Humeral Head in Modern Locking Plates: A Cadaver Study [J].
Erhardt, Johannes B. ;
Stoffel, Karl ;
Kampshoff, Jorg ;
Badur, Nicole ;
Yates, Piers ;
Kuster, Markus S. .
JOURNAL OF ORTHOPAEDIC TRAUMA, 2012, 26 (10) :E188-E192
[9]   Investigation of metallic and carbon fibre PEEK fracture fixation devices for three-part proximal humeral fractures [J].
Feerick, Emer M. ;
Kennedy, Jim ;
Mullett, Hannan ;
FitzPatrick, David ;
McGarry, Patrick .
MEDICAL ENGINEERING & PHYSICS, 2013, 35 (06) :712-722
[10]   Dynamic-locking-screw (DLS)-leads to less secondary screw perforations in proximal humerus fractures [J].
Freude, Thomas ;
Schroeter, Steffen ;
Plecko, Michael ;
Bahrs, Christian ;
Martetschlaeger, Frank ;
Kraus, Tobias M. ;
Stoeckle, Ulrich ;
Doebele, Stefan .
BMC MUSCULOSKELETAL DISORDERS, 2014, 15