Ceramic-on-ceramic catastrophic liner failure in total hip arthroplasty: Morphological and compositional analysis of fractured ceramic components

被引:5
作者
Bistolfi, Alessandro [1 ]
Ferracini, Riccardo [2 ,3 ]
Lee, Gwo Chin [4 ]
Mellano, Danilo [1 ]
Guidotti, Claudio [1 ]
Baino, Francesco [5 ]
Verne, Enrica [5 ]
机构
[1] CTO Hosp, Dept Orthopaed Traumatol & Rehabil Med, AO Citta Salute & Sci, Turin, Italy
[2] Koelliker Orthopaed, Turin, Italy
[3] Univ Genoa, Dipartimento Sci Chirurg & Diagnost Integrate, Genoa, Italy
[4] Penn Hosp, Hosp Univ Penn, Penn Presbyterian Med Ctr, Philadelphia, PA 19107 USA
[5] Politecn Torino, Dept Appl Sci & Technol, Inst Mat Phys & Engn, Turin, Italy
关键词
Bioceramics; Total hip arthroplasty; Fracture; Scanning electron microscopy;
D O I
10.1016/j.ceramint.2020.12.225
中图分类号
TQ174 [陶瓷工业]; TB3 [工程材料学];
学科分类号
0805 ; 080502 ;
摘要
Surface bearing of total hip arthroplasty (THA) still is a strong subject of study due to the relatively high rate of failures caused by a multiplicity of factors including surgical technique, patient?s attitude, and type/characteristics of the materials used (metal, ceramics, polyethylene) with their specific risk factors. Fractures of the ceramic components are rare but catastrophic events, with many concerns among the orthopaedic surgeons. Such complication is usually evaluated from a clinical viewpoint; this study provides a materials scientist?s complementary perspective and comprehensively evaluates the surface and the mechanism of rupture of the ceramic liner in two cases with different ceramics (Biolox Forte and Biolox Delta) after ceramic-on-ceramic THA. The morphological and compositional analyses of the ceramic components were performed by field-emission scanning electron microscopy (FESEM) and energy dispersive spectroscopy (EDS), along with macroscopic pictures. The SEM analysis of the ceramic liner showed many wear signs in all directions, while the EDS revealed the presence of titanium near to the fracture border, which might be interpreted as a consequence of the catastrophic contact between ceramic liner and metal back. Biolox Delta and Biolox Forte ceramic liners showed different patterns of fracture and surface modifications that are illustrated and discussed.
引用
收藏
页码:11029 / 11036
页数:8
相关论文
共 36 条
[21]   Outcomes of Ceramic Bearings After Primary Total Hip Arthroplasty in the Medicare Population [J].
Kurtz, Steven M. ;
Lau, Edmund ;
Baykal, Doruk ;
Springer, Bryan D. .
JOURNAL OF ARTHROPLASTY, 2017, 32 (03) :743-749
[22]   Bearing Surfaces for Total Hip Arthroplasty [J].
Lachiewicz, Paul F. ;
Kleeman, Lindsay T. ;
Seyler, Thorsten .
JOURNAL OF THE AMERICAN ACADEMY OF ORTHOPAEDIC SURGEONS, 2018, 26 (02) :45-57
[23]   The operation of the century: total hip replacement [J].
Learmonth, Ian D. ;
Young, Claire ;
Rorabeck, Cecil .
LANCET, 2007, 370 (9597) :1508-1519
[24]  
Lee G.-C., 2015, Seminars in Arthroplasty, Current Concepts in Joint Alternative Bearings, V26, P11
[25]   Does Biolox® Delta ceramic reduce the rate of component fractures in total hip replacement? [J].
Massin, P. ;
Lopes, R. ;
Masson, B. ;
Mainard, D. .
ORTHOPAEDICS & TRAUMATOLOGY-SURGERY & RESEARCH, 2014, 100 (06) :S317-S321
[26]   Long-Term Performance of Ceramic and Metal Femoral Heads on Conventional Polyethylene in Young and Active Patients A Matched-Pair Analysis [J].
Meftah, Morteza ;
Klingenstein, Gregory G. ;
Yun, Richard J. ;
Ranawat, Amar S. ;
Ranawat, Chitranjan S. .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2013, 95A (13) :1193-1197
[27]   Ceramics for prosthetic hip and knee joint replacement [J].
Rahaman, Mohamed N. ;
Yao, Aihua ;
Bal, B. Sonny ;
Garino, Jonathan P. ;
Ries, Michael D. .
JOURNAL OF THE AMERICAN CERAMIC SOCIETY, 2007, 90 (07) :1965-1988
[28]   Mechanisms of fatigue-crack propagation in ductile and brittle solids [J].
Ritchie, RO .
INTERNATIONAL JOURNAL OF FRACTURE, 1999, 100 (01) :55-83
[29]  
SCHMALZRIED TP, 1992, CLIN ORTHOP RELAT R, P60
[30]   Squeaking in Third- and Fourth-Generation Ceramic-on-Ceramic Total Hip Arthroplasty Meta-Analysis and Systematic Review [J].
Stanat, Scott J. C. ;
Capozzi, James D. .
JOURNAL OF ARTHROPLASTY, 2012, 27 (03) :445-453