The long-term stability of amorphous silicon flat panel imaging devices for dosimetry purposes

被引:72
作者
Louwe, RJW [1 ]
McDermott, LN [1 ]
Sonke, JJ [1 ]
Tielenburg, R [1 ]
Wendling, M [1 ]
van Herk, MB [1 ]
Mijnheer, BJ [1 ]
机构
[1] Antoni Van Leeuwenhoek Hosp, Netherlands Canc Inst, Dept Radiotherapy, Amsterdam, Netherlands
关键词
radiotherapy; dose verification; flat panel imager; portal dosimetry; quality assurance;
D O I
10.1118/1.1803751
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
This study was carried out to determine the stability of the response of amorphous silicon (a-Si)-flat panel imagers for dosimetry applications. Measurements of the imager's response under reference conditions were performed on a regular basis for four detectors of the same manufacturer. We found that the ambient temperature influenced the dark-field, while the gain of the imager signal was unaffected. Therefore, temperature fluctuations were corrected for by applying a "dynamic" dark-field correction. This correction method also removed the influence of a small, irreversible increase of the dark-field current, which was equal to 0.5% of the dynamic range of the imager per year and was probably caused by mild radiation damage to the a-Si array. By applying a dynamic dark-field correction, excellent stability of the response over the entire panel of all imagers of 0.5% (1 SD) was obtained over an observation period up to 23 months. However, two imagers had to be replaced after several months. For one imager, an image segment stopped functioning, while the image quality of the other imager degraded significantly. We conclude that the tested a-Si EPIDs have a very stable response and are therefore well suited for dosimetry. We recommend, however, applying quality assurance tests dedicated to both imaging and dosimetry. (C) 2004 American Association of Physicists in Medicine.
引用
收藏
页码:2989 / 2995
页数:7
相关论文
共 16 条
[1]  
ANTONUK LE, 1998, AAPM MED PHYSICS MON, V24, P371
[2]  
Bernstein R, 1997, INTERNATIONAL CONFERENCE ON IMAGE PROCESSING - PROCEEDINGS, VOL I, P287, DOI 10.1109/ICIP.1997.647761
[3]   The dose response relationship of a liquid-filled electronic portal imaging device [J].
Boellaard, R ;
vanHerk, M ;
Mijnheer, BJ .
MEDICAL PHYSICS, 1996, 23 (09) :1601-1611
[4]   DAILY DOSIMETRIC QUALITY-CONTROL OF THE MM50 RACETRACK MICROTRON USING AN ELECTRONIC PORTAL IMAGING DEVICE [J].
DIRKX, MLP ;
KROONWIJK, M ;
DEBOER, JCJ ;
HEIJMEN, BJM .
RADIOTHERAPY AND ONCOLOGY, 1995, 37 (01) :55-60
[5]   Relative dosimetry using active matrix flat-panel imager (AMFPI) technology [J].
El-Mohri, Y ;
Antonuk, LE ;
Yorkston, J ;
Jee, KW ;
Maolinbay, M ;
Lam, KL ;
Siewerdsen, JH .
MEDICAL PHYSICS, 1999, 26 (08) :1530-1541
[6]   Dosimetric properties of the Theraview™ fluoroscopic electronic portal imaging device [J].
Glendinning, AG ;
Bonnett, DE .
BRITISH JOURNAL OF RADIOLOGY, 2000, 73 (869) :517-530
[7]   Dosimetric properties of an amorphous silicon electronic portal imaging device for verification of dynamic intensity modulated radiation therapy [J].
Greer, PB ;
Popescu, CC .
MEDICAL PHYSICS, 2003, 30 (07) :1618-1627
[8]   Clinical use of electronic portal imaging: Report of AAPM radiation therapy committee Task Group 58 [J].
Herman, MG ;
Balter, JM ;
Jaffray, DA ;
McGee, KP ;
Munro, P ;
Shalev, S ;
Van Herk, M ;
Wong, JW .
MEDICAL PHYSICS, 2001, 28 (05) :712-737
[9]   The stability of liquid-filled matrix ionization chamber electronic portal imaging devices for dosimetry purposes [J].
Louwe, RJW ;
Tielenburg, R ;
van Ingen, KM ;
Mijnheer, BJ ;
van Herk, MB .
MEDICAL PHYSICS, 2004, 31 (04) :819-827
[10]   Three-dimensional dose reconstruction of breast cancer treatment using portal imaging [J].
Louwe, RJW ;
Damen, EMF ;
van Herk, M ;
Minken, AWH ;
Törzsök, O ;
Mijnheer, BJ .
MEDICAL PHYSICS, 2003, 30 (09) :2376-2389