Particle size spectrometer using inertial classification and electrical measurement techniques for real-time monitoring of particle size distribution

被引:4
作者
Kwon, Hong-Beom [1 ]
Kim, Hong-Lae [1 ]
Hong, Ui-Seon [1 ]
Yoo, Seong-Jae [1 ]
Kim, Kyongtae [1 ]
Han, Jangseop [1 ]
Kim, Min-Ki [1 ]
Hwang, Jungho [1 ]
Kim, Yong-Jun [1 ]
机构
[1] Yonsei Univ, Sch Mech Engn, 50 Yonsei Ro, Seoul 03722, South Korea
基金
新加坡国家研究基金会;
关键词
PARTICULATE MATTER; CARDIOVASCULAR-DISEASE; FINE PARTICLES; AIR-POLLUTION; DEPOSITION; AEROSOL; MORTALITY; TERM;
D O I
10.1039/c8lc00429c
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
To achieve real-time monitoring of aerodynamic submicron particle size distributions at a point-of-interest, we developed a high-performance particle size spectrometer that is compact, low-cost, and portable. The present system consists of four key components: a unipolar mini-discharger for electrically charging particles, an inertial size-separator for classifying charged particles into five size fractions in terms of their aerodynamic sizes, a portable multi-channel electrometer for detecting femto-ampere currents carried by charged particles at each stage, and a retrieval algorithm for converting the current data into a smooth particle size distribution. The unipolar mini-discharger and inertial size separator were quantitatively characterised by using standard polystyrene latex (PSL) particles. The experimentally determined cut-off diameters at each stage in the inertial size separator were 1.17, 0.94, 0.71, 0.54, and 0.23 m, respectively. Then, the system was compared with a commercial reference aerodynamic particle sizer (APS) in the environment where the number concentration and the average size of TiO2 particles were changing. The present system resolved peak size and geometric standard deviation of particles to within 11.2%, and 6.3%, respectively, indicating that the system can be used to accurately monitor submicron particle size distributions in real time.
引用
收藏
页码:2642 / 2652
页数:11
相关论文
共 50 条
  • [1] Particulate Matter Air Pollution and Cardiovascular Disease An Update to the Scientific Statement From the American Heart Association
    Brook, Robert D.
    Rajagopalan, Sanjay
    Pope, C. Arden, III
    Brook, Jeffrey R.
    Bhatnagar, Aruni
    Diez-Roux, Ana V.
    Holguin, Fernando
    Hong, Yuling
    Luepker, Russell V.
    Mittleman, Murray A.
    Peters, Annette
    Siscovick, David
    Smith, Sidney C., Jr.
    Whitsel, Laurie
    Kaufman, Joel D.
    [J]. CIRCULATION, 2010, 121 (21) : 2331 - 2378
  • [2] Particle Deposition in a Child Respiratory Tract Model: In Vivo Regional Deposition of Fine and Ultrafine Aerosols in Baboons
    Albuquerque-Silva, Iolanda
    Vecellio, Laurent
    Durand, Marc
    Avet, John
    Le Pennec, Deborah
    de Monte, Michele
    Montharu, Jerome
    Diot, Patrice
    Cottier, Michele
    Dubois, Francis
    Pourchez, Jeremie
    [J]. PLOS ONE, 2014, 9 (04):
  • [3] [Anonymous], 1994, Ann ICRP, V24, P1
  • [4] [Anonymous], ATMOS CHEM PHYS
  • [5] Fine particle components and health-a systematic review and meta-analysis of epidemiological time series studies of daily mortality and hospital admissions
    Atkinson, Richard W.
    Mills, Inga C.
    Walton, Heather A.
    Anderson, H. Ross
    [J]. JOURNAL OF EXPOSURE SCIENCE AND ENVIRONMENTAL EPIDEMIOLOGY, 2015, 25 (02) : 208 - 214
  • [6] Local particle deposition patterns may play a key role in the development of lung cancer
    Balásházy, I
    Hofmann, W
    Heistracher, T
    [J]. JOURNAL OF APPLIED PHYSIOLOGY, 2003, 94 (05) : 1719 - 1725
  • [7] Baron P.A., 2001, AEROSOL MEASUREMENT, V2nd
  • [8] BOUCHIKHI A, 1988, EUR RESPIR J, V1, P547
  • [9] Thoracic and respirable particle definitions for human health risk assessment
    Brown, James S.
    Gordon, Terry
    Price, Owen
    Asgharian, Bahman
    [J]. PARTICLE AND FIBRE TOXICOLOGY, 2013, 10
  • [10] PERFORMANCE OF A TSI AERODYNAMIC PARTICLE SIZER
    CHEN, BT
    CHENG, YS
    YEH, HC
    [J]. AEROSOL SCIENCE AND TECHNOLOGY, 1985, 4 (01) : 89 - 97