Development of computer-assisted minimal-dose system with beam blanker for TEM

被引:12
作者
Hayashida, Misa [1 ]
Nomaguchi, Tsunenori [1 ]
Kimura, Yoshihide [1 ]
Takai, Yoshizo [1 ]
机构
[1] Osaka Univ, Dept Mat & Life Sci, Grad Sch Engn, Suita, Osaka 5650871, Japan
关键词
TEM; radiation damage; computer control;
D O I
10.1016/j.micron.2006.07.024
中图分类号
TH742 [显微镜];
学科分类号
摘要
A computer-assisied minimal-dose system has been developed for the high-resolution observation of radiation-sensitive samples using a transmission electron microscope (TEM). This system consists of a CCD camera, a beam blanker and a control computer (PC) that also controls the TEM. A sample is illuminated by an electron beam only when the CCD camera takes images; otherwise, the beam blanker cuts off the electron beam. Emulated images, which are calculated from the images taken and the variable parameters of the TEM, such as magnification and sample stage position, are displayed on the control PC display. After a few times of repetition of exposures and emulations, the sample is positioned to final observation area. Subsequently to select the observation area, the system automatically adjusts the focus position from two different illuminating angle images and the TEM conditions are appropriate for taking a final image. The total electron dose before the final image is taken can be markedly decreased because the sample is irradiated by an electron beam only when images are taken. For the final image, a three-dimensional Fourier filtering method (3DFFM), which corrects the spherical aberration for the through-focus image series and enables the selection of the optimum focus condition later, is also included in our system. (c) 2006 Elsevier Ltd. All rights reserved.
引用
收藏
页码:505 / 512
页数:8
相关论文
共 11 条
[1]   Electron tomography of molecules and cells [J].
Baumeister, W ;
Grimm, R ;
Walz, J .
TRENDS IN CELL BIOLOGY, 1999, 9 (02) :81-85
[2]  
FUJIYOSHI Y, 1980, ULTRAMICROSCOPY, V5, P459, DOI 10.1016/0304-3991(80)90046-7
[3]   Wave field restoration using three-dimensional Fourier filtering method [J].
Kawasaki, T ;
Takai, Y ;
Ikuta, T ;
Shimizu, R .
ULTRAMICROSCOPY, 2001, 90 (01) :47-59
[4]   Development of a real-time defocus image modulation processing electron microscope. I. Construction [J].
Kimura, Y ;
Takai, Y ;
Kawasaki, T ;
Shimizu, R ;
Ikuta, T ;
Isakozawa, S ;
Satou, Y ;
Ichihashi, M .
JOURNAL OF ELECTRON MICROSCOPY, 1999, 48 (06) :873-878
[5]   PRACTICAL AUTOALIGNMENT OF TRANSMISSION ELECTRON-MICROSCOPES [J].
KOSTER, AJ ;
DERUIJTER, WJ .
ULTRAMICROSCOPY, 1992, 40 (02) :89-107
[6]   Automated acquisition of cryo-electron micrographs for single particle reconstruction on an FEI Tecnai electron microscope [J].
Lei, JL ;
Frank, J .
JOURNAL OF STRUCTURAL BIOLOGY, 2005, 150 (01) :69-80
[7]   MINIMAL BEAM EXPOSURE WITH A FIELD-EMISSION SOURCE [J].
OHTSUKI, M ;
ZEITLER, E .
ULTRAMICROSCOPY, 1975, 1 (02) :163-165
[8]  
TANIGUCHI Y, 1992, J ELECTRON MICROSC, V41, P21
[9]   PHASE-CONTRAST CHARACTERISTICS IN BRIGHT FIELD ELECTRON-MICROSCOPY [J].
WADE, RH .
ULTRAMICROSCOPY, 1978, 3 (03) :329-334
[10]   ELECTRON MICROSCOPY OF TOBACCO MOSAIC VIRUS UNDER CONDITIONS OF MINIMAL BEAM EXPOSURE [J].
WILLIAMS, RC ;
FISHER, HW .
JOURNAL OF MOLECULAR BIOLOGY, 1970, 52 (01) :121-&