Double inversion black-blood fast spin-echo imaging of the human heart: A comparison between 1.5T and 3.0T

被引:103
作者
Greenman, RL
Shirosky, JE
Mulkern, RV
Rofsky, NM
机构
[1] Beth Israel Deaconess Med Ctr, Dept Radiol, Boston, MA 02115 USA
[2] Harvard Univ, Sch Med, Boston, MA USA
[3] Boston Childrens Hosp, Dept Radiol, Boston, MA USA
关键词
black-blood imaging; fast spin-echo; high-field imaging; B1; mapping; cardiovascular MRI; 3T MRI;
D O I
10.1002/jmri.10316
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To evaluate the effectiveness of blood suppression and the quality of black-blood cardiac images acquired at 3.0 Tesla using a double-inversion recovery fast spin-echo sequence by comparing data acquired at 3.0T to data acquired at 1.5T. Materials and Methods: Black-blood T2-weighted fast spin-echo images of the heart were acquired from five normal volunteers at 1.5T and five normal volunteers at 3.0T. Region-of-interest signal intensity measurements were performed at several locations in the suppressed blood regions of the left and right ventricles and around the left ventricle walls to assess the effectiveness and uniformity of the blood suppression, the myocardial signal-to-noise ratio (SNR), and the signal uniformity at both field strengths. B I field maps were produced in phantoms and in subjects at both field strengths. Results: Blood suppression performance is equivalent at 1.5T and 3.0T. The improvement in SNR at 3.0T compared with 1.5T is less than has been predicted in previous studies. The signal uniformity is significantly poorer at 3.0T than at 1.5T due to dielectric effects and shorter radio frequency wavelengths (P < 0.005). Conclusion: Spin-echo and spin-echo echo-train sequences that perform well at 1.5T will produce large signal variations in the chest cavity at 3.0T without modifications. B1 insensitive methods must be explored and implemented for spin-echo sequences to fully realize the advantages of using these sequences for high-field MRI.
引用
收藏
页码:648 / 655
页数:8
相关论文
共 26 条
[1]   SUPPRESSION OF ARTIFACTS IN 2-DIMENSIONAL J SPECTROSCOPY [J].
BODENHAUSEN, G ;
FREEMAN, R ;
TURNER, DL .
JOURNAL OF MAGNETIC RESONANCE, 1977, 27 (03) :511-514
[2]   RF MAGNETIC-FIELD PENETRATION, PHASE-SHIFT AND POWER DISSIPATION IN BIOLOGICAL TISSUE - IMPLICATIONS FOR NMR IMAGING [J].
BOTTOMLEY, PA ;
ANDREW, ER .
PHYSICS IN MEDICINE AND BIOLOGY, 1978, 23 (04) :630-643
[3]   ESTIMATING RADIOFREQUENCY POWER DEPOSITION IN BODY NMR IMAGING [J].
BOTTOMLEY, PA ;
REDINGTON, RW ;
EDELSTEIN, WA ;
SCHENCK, JF .
MAGNETIC RESONANCE IN MEDICINE, 1985, 2 (04) :336-349
[4]   MR IMAGING - CLINICAL USE OF THE INVERSION RECOVERY SEQUENCE [J].
BYDDER, GM ;
YOUNG, IR .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1985, 9 (04) :659-675
[5]  
Dougherty L, 2001, MAGNET RESON MED, V45, P176, DOI 10.1002/1522-2594(200101)45:1<176::AID-MRM1025>3.0.CO
[6]  
2-E
[7]   In vivo magnetic resonance evaluation of atherosclerotic plaques in the human thoracic aorta - A comparison with transesophageal echocardiography [J].
Fayad, ZA ;
Nahar, T ;
Fallon, JT ;
Goldman, M ;
Aguinaldo, JG ;
Badimon, JJ ;
Shinnar, M ;
Chesebro, JH ;
Fuster, V .
CIRCULATION, 2000, 101 (21) :2503-2509
[8]  
Fayad ZA, 2000, CIRCULATION, V102, P506
[9]   PROPERTIES OF THE PHASE-ALTERNATING PHASE-SHIFT (PHAPS) MULTIPLE SPIN-ECHO PROTOCOL IN MRI - A STUDY OF THE EFFECTS OF IMPERFECT RF PULSES [J].
FRANSSON, A ;
ERICSSON, A ;
JUNG, B ;
SPERBER, GO .
MAGNETIC RESONANCE IMAGING, 1993, 11 (06) :771-784
[10]  
GARWOOD M, 1993, PHYSICS MRI, P110