The use of extended dose range film for dosimetric calibration of a scanning liquid-filled ionization chamber electronic portal imaging device

被引:8
作者
Mohammadi, Mohammad
Bezak, Eva
Reich, Paul
机构
[1] Royal Adelaide Hosp, Dept Med Phys, Adelaide, SA 5000, Australia
[2] Univ Adelaide, Sch Chem & Phys, Adelaide, SA 5005, Australia
[3] Hamadan Univ Med Sci, Fac Med, Dept Med Phys, Hamadan, Iran
来源
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS | 2007年 / 8卷 / 01期
关键词
EDR2; film; electronic portal imaging; dose verification; transit dosimetry; treatment planning system;
D O I
10.1120/jacmp.v8i1.2305
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A scanning liquid-filled ionization chamber electronic portal imaging device (SLIC-EPID) and extended dose range (EDR2) films were used to evaluate transmitted dose profiles for homogeneous and inhomogeneous phantoms. Calibrated ionization chamber measurements were used to convert the pixel values acquired from the electronic portal images to dose. Because SLIC-EPID was developed to have a uniform response for all liquid ionization chambers, the off-axis dose values were reconstructed using a correction factor matrix, defined as the ratio of the relative EDR2 film and the corresponding EPID dose values measured in air. The transmitted dose distributions in the EPID detector layer were also modeled using a Pinnacle(3) treatment planning system (TPS: Philips Radiation Oncology Systems, Milpitas, CA). The gamma function algorithm was then used to assess agreement between transmitted dose distributions measured using a SLIC-EPID and EDR2 film, and those calculated using the TPS. For homogenous and inhomogeneous phantoms, more than 90% agreement was achieved using gamma criteria of 2% and 3 mm and 3% and 2.5 mm respectively. Our results indicate that the calibration procedure proposed in the present study should be performed if SLIC-EPID is to be used as a reliable two-dimensional transmitted dosimeter for clinical purposes.
引用
收藏
页码:69 / 84
页数:16
相关论文
共 50 条
  • [1] COLLAPSED CONE CONVOLUTION OF RADIANT ENERGY FOR PHOTON DOSE CALCULATION IN HETEROGENEOUS MEDIA
    AHNESJO, A
    [J]. MEDICAL PHYSICS, 1989, 16 (04) : 577 - 592
  • [2] Two-dimensional exit dosimetry using a liquid-filled electronic portal imaging device and a convolution model
    Boellaard, R
    vanHerk, M
    Uiterwaal, H
    Mijnheer, B
    [J]. RADIOTHERAPY AND ONCOLOGY, 1997, 44 (02) : 149 - 157
  • [3] First clinical tests using a liquid-filled electronic portal imaging device and a convolution model for the verification of the midplane dose
    Boellaard, R
    van Herk, M
    Uiterwaal, H
    Mijnheer, B
    [J]. RADIOTHERAPY AND ONCOLOGY, 1998, 47 (03) : 303 - 312
  • [4] New method to obtain the midplane dose using portal in vivo dosimetry
    Boellaard, R
    Essers, M
    van Herk, M
    Mijnheer, BJ
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1998, 41 (02): : 465 - 474
  • [5] A convolution model to convert transmission dose images to exit dose distributions
    Boellaard, R
    vanHerk, M
    Mijnheer, BJ
    [J]. MEDICAL PHYSICS, 1997, 24 (02) : 189 - 199
  • [6] Interinstitutional variations of sensitometric curves of radiographic dosimetric films
    Bos, LJ
    Danciu, C
    Cheng, CW
    Brugmans, MJP
    van der Horst, A
    Minken, A
    Mijnheer, BJ
    [J]. MEDICAL PHYSICS, 2002, 29 (08) : 1772 - 1780
  • [7] A REVIEW OF ELECTRONIC PORTAL IMAGING DEVICES (EPIDS)
    BOYER, AL
    ANTONUK, L
    FENSTER, A
    VANHERK, M
    MEERTENS, H
    MUNRO, P
    REINSTEIN, LE
    WONG, J
    [J]. MEDICAL PHYSICS, 1992, 19 (01) : 1 - 16
  • [8] A new approach to film dosimetry for high energy photon beams: Lateral scatter filtering
    Burch, SE
    Kearfott, KJ
    Trueblood, JH
    Sheils, WC
    Yeo, JI
    Wang, CKC
    [J]. MEDICAL PHYSICS, 1997, 24 (05) : 775 - 783
  • [9] An iterative EPID calibration procedure for dosimetric verification that considers the EPID scattering factor
    Chang, JH
    Mageras, GS
    Ling, CC
    Lutz, W
    [J]. MEDICAL PHYSICS, 2001, 28 (11) : 2247 - 2257
  • [10] Evaluation of rapid dose map acquisition of a scanning liquid-filled ionization chamber electronic portal imaging device
    Chang, JW
    Mageras, GS
    Ling, CC
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2003, 55 (05): : 1432 - 1445