Catheter ablation of atrial flutter guided by electroanatomic mapping (CARTO):: A randomized comparison to the conventional approach

被引:69
作者
Willems, S [1 ]
Weiss, C [1 ]
Ventura, R [1 ]
Rüppel, R [1 ]
Risius, T [1 ]
Hoffmann, M [1 ]
Meinertz, T [1 ]
机构
[1] Univ Hamburg, Hosp Eppendorf, Dept Cardiol, D-20246 Hamburg, Germany
关键词
atrial flutter; catheter ablation; electroanatomic mapping;
D O I
10.1046/j.1540-8167.2000.01223.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Introduction: Three-dimensional electroanatomic (CARTO) activation mapping of the cavotricuspid isthmus can be helpful to guide atrial flutter ablation, but to date has not been investigated in comparison to conventional strategies. The aim of the present study was to assess the efficacy of the CARTO navigation system, especially with respect to the fluoroscopy time required for successful atrial flutter ablation. Methods and Results: Eighty patients with recurrent common-type atrial flutter were randomly assigned to temperature-controlled radiofrequency (RF) catheter ablation, either guided by conventional criteria (group 1) or additionally oriented on electroanatomic mapping (group 2). In all patients, similar multipolar catheters were inserted into the coronary sinus and placed at the tricuspid annulus, respectively. In group 2, positioning of the mapping electrode and delivery of RF pulses within the cavotricuspid isthmus was mainly oriented on the CARTO map to achieve the most linear and continuous RF lesions. Abolition of intra-atrial conduction verified by conventional criteria (group 1) and electroanatomic mapping (group 2) could be verified in all patients. The overall number of RF pulses (group 1: 16.7 +/- 6.5; group 2: 13.2 +/- 5.3) and mean procedure duration (group 1: 172.5 +/- 47.4 min; group 2: 169.3 +/- 47.3 min) were not different between the two groups, but mean fluoroscopy time was significantly shorter when the CARTO technology was used (group 1: 29.2 +/- 9.4 min; group 2: 7.7 +/- 2.8 min; P = 0.0001). Recurrence of atrial flutter was observed in 3 (9%) patients in each group after a mean follow-up of 8.5 +/- 2.8 months. Conclusion: Atrial flutter can be abolished effectively using the conventional technique as well as oriented on electroanatomic mapping. However, overall X-ray exposure can be significantly reduced by the CARTO-guided approach without prolongation of procedure duration.
引用
收藏
页码:1223 / 1230
页数:8
相关论文
共 18 条
[1]  
CHAUCHEMEZ B, 1996, CIRCULATION, V93, P284
[2]  
Gepstein L, 1997, CIRCULATION, V95, P1611
[3]   Successful irrigated-tip catheter ablation of atrial flutter resistant to conventional radiofrequency ablation [J].
Jaïs, P ;
Haïssaguerre, M ;
Shah, DC ;
Takahashi, A ;
Hocini, M ;
Lavergne, T ;
Lafitte, S ;
Le Mouroux, A ;
Fischer, B ;
Clémenty, J .
CIRCULATION, 1998, 98 (09) :835-838
[4]   Nonfluoroscopic three-dimensional mapping for arrhythmia ablation: Tool or toy? [J].
Khongphatthanayothin, A ;
Kosar, E ;
Nademanee, K .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2000, 11 (03) :239-243
[5]   Three-dimensional electromagnetic catheter technology: Electroanatomical mapping of the right atrium and ablation of ectopic atrial tachycardia [J].
Kottkamp, H ;
Hindricks, G ;
Breithardt, G ;
Borggrefe, M .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 1997, 8 (12) :1332-1337
[6]   Risk to patients from radiation associated with radiofrequency ablation for supraventricular tachycardia [J].
Kovoor, P ;
Ricciardello, M ;
Collins, L ;
Uther, JB ;
Ross, DL .
CIRCULATION, 1998, 98 (15) :1534-1540
[7]  
LESH M, 1999, CARDIAC ELECTROPHYSI, P1009
[8]   Use of a three-dimensional, nonfluoroscopic mapping system for catheter ablation of typical atrial flutter [J].
Nakagawa, H ;
Jackman, WM .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1998, 21 (06) :1279-1286
[9]   Radio frequency catheter ablation of atrial flutter - Further insights into the various types of isthmus block: Application to ablation during sinus rhythm [J].
Poty, H ;
Saoudi, N ;
Nair, M ;
Anselme, F ;
Letac, B .
CIRCULATION, 1996, 94 (12) :3204-3213
[10]  
POTY H, 1995, CIRCULATION, V92, P1389