Results From the Imaging and Radiation Oncology Core Houston's Anthropomorphic Phantoms Used for Proton Therapy Clinical Trial Credentialing

被引:41
作者
Taylor, Paige A. [1 ]
Kry, Stephen F. [1 ]
Alvarez, Paola [1 ]
Keith, Tyler [1 ]
Lujano, Carrie [1 ]
Hernandez, Nadia [1 ]
Followill, David S. [1 ]
机构
[1] Univ Texas MD Anderson Canc Ctr, Dept Radiat Phys, Unit 607,1515 Holcombe Blvd, Houston, TX 77030 USA
来源
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS | 2016年 / 95卷 / 01期
关键词
DOSE CALCULATIONS;
D O I
10.1016/j.ijrobp.2016.01.061
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: The purpose of this study was to summarize the findings of anthropomorphic proton phantom irradiations analyzed by the Imaging and Radiation Oncology Core Houston QA Center (IROC Houston). Methods and Materials: A total of 103 phantoms were irradiated by proton therapy centers participating in clinical trials. The anthropomorphic phantoms simulated heterogeneous anatomy of a head, liver, lung, prostate, and spine. Treatment plans included those for scattered, uniform scanning, and pencil beam scanning beam delivery modalities using 5 different treatment planning systems. For every phantom irradiation, point doses and planar doses were measured using thermoluminescent dosimeters (TLD) and film, respectively. Differences between measured and planned doses were studied as a function of phantom, beam delivery modality, motion, repeat attempt, treatment planning system, and date of irradiation. Results: The phantom pass rate (overall, 79%) was high for simple phantoms and lower for phantoms that introduced higher levels of difficulty, such as motion, multiple targets, or increased heterogeneity. All treatment planning systems overestimated dose to the target, compared to TLD measurements. Errors in range calculation resulted in several failed phantoms. There was no correlation between treatment planning system and pass rate. The pass rates for each individual phantom are not improving over time, but when individual institutions received feedback about failed phantom irradiations, pass rates did improve. Conclusions: The proton phantom pass rates are not as high as desired and emphasize potential deficiencies in proton therapy planning and/or delivery. There are many areas for improvement with the proton phantom irradiations, such as treatment planning system dose agreement, range calculations, accounting for motion, and irradiation of multiple targets. Published by Elsevier Inc.
引用
收藏
页码:242 / 248
页数:7
相关论文
共 13 条
[1]  
Blatnica A, 2011, THESIS U TEXAS HOUST
[2]   Partial Breast Irradiation Delivered With Proton Beam: Results of a Phase II Trial [J].
Bush, David A. ;
Slater, Jerry D. ;
Garberoglio, Carlos ;
Do, Sharon ;
Lum, Sharon ;
Slater, James M. .
CLINICAL BREAST CANCER, 2011, 11 (04) :241-245
[3]   Phase 2 Study of High-Dose Proton Therapy With Concurrent Chemotherapy for Unresectable Stage III Nonsmall Cell Lung Cancer [J].
Chang, Joe Y. ;
Komaki, Ritsuko ;
Lu, Charles ;
Wen, Hong Y. ;
Allen, Pamela K. ;
Tsao, Anne ;
Gillin, Michael ;
Mohan, Radhe ;
Cox, James D. .
CANCER, 2011, 117 (20) :4707-4713
[4]   Relative stopping power measurements to aid in the design of anthropomorphic phantoms for proton radiotherapy [J].
Grant, Ryan L. ;
Summers, Paige A. ;
Neihart, James L. ;
Blatnica, Anthony P. ;
Sahoo, Narayan ;
Gillin, Michael T. ;
Followill, David S. ;
Ibbott, Geoffrey S. .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2014, 15 (02) :121-126
[5]   Independent evaluations of IMRT through the use of an anthropomorphic phantom [J].
Ibbott, Geoffrey S. ;
Molineu, Andrea ;
Followill, David S. .
TECHNOLOGY IN CANCER RESEARCH & TREATMENT, 2006, 5 (05) :481-487
[6]  
IROC Houston QA Center, GUID US PROT RAD THE
[7]   UNCERTAINTY ANALYSIS OF ABSORBED DOSE CALCULATIONS FROM THERMOLUMINESCENCE DOSIMETERS [J].
KIRBY, TH ;
HANSON, WF ;
JOHNSTON, DA .
MEDICAL PHYSICS, 1992, 19 (06) :1427-1433
[8]   Algorithms Used in Heterogeneous Dose Calculations Show Systematic Differences as Measured With the Radiological Physics Center's Anthropomorphic Thorax Phantom Used for RTOG Credentialing [J].
Kry, Stephen F. ;
Alvarez, Paola ;
Molineu, Andrea ;
Amador, Carrie ;
Galvin, James ;
Followill, David S. .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2013, 85 (01) :E95-E100
[9]   EARLY OUTCOMES FROM THREE PROSPECTIVE TRIALS OF IMAGE-GUIDED PROTON THERAPY FOR PROSTATE CANCER [J].
Mendenhall, Nancy P. ;
Li, Zuofeng ;
Hoppe, Bradford S. ;
Marcus, Robert B., Jr. ;
Mendenhall, William M. ;
Nichols, R. Charles ;
Morris, Christopher G. ;
Williams, Christopher R. ;
Costa, Joseph ;
Henderson, Randal .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2012, 82 (01) :213-221
[10]   Design and implementation of an anthropomorphic quality assurance phantom for intensity-modulated radiation therapy for the radiation therapy oncology group [J].
Molineu, A ;
Followill, DS ;
Balter, PA ;
Hanson, WF ;
Gillin, MT ;
Huq, MS ;
Eisbruch, A ;
Ibbott, GS .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2005, 63 (02) :577-583