Adverse Events in Robotic Surgery: A Retrospective Study of 14 Years of FDA Data

被引:223
作者
Alemzadeh, Homa [1 ]
Raman, Jaishankar [2 ]
Leveson, Nancy [3 ]
Kalbarczyk, Zbigniew [1 ]
Iyer, Ravishankar K. [1 ]
机构
[1] Univ Illinois, Coordinated Sci Lab, Urbana, IL USA
[2] Rush Univ, Med Ctr, Dept Surg, Chicago, IL 60612 USA
[3] MIT, Dept Aeronaut & Astronaut, Cambridge, MA 02139 USA
基金
美国国家科学基金会;
关键词
LAPAROSCOPIC PROSTATECTOMY; SURGICAL SYSTEM; MALFUNCTION; FAILURE;
D O I
10.1371/journal.pone.0151470
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
Background Use of robotic systems for minimally invasive surgery has rapidly increased during the last decade. Understanding the causes of adverse events and their impact on patients in robot-assisted surgery will help improve systems and operational practices to avoid incidents in the future. Methods By developing an automated natural language processing tool, we performed a comprehensive analysis of the adverse events reported to the publicly available MAUDE database (maintained by the U.S. Food and Drug Administration) from 2000 to 2013. We determined the number of events reported per procedure and per surgical specialty, the most common types of device malfunctions and their impact on patients, and the potential causes for catastrophic events such as patient injuries and deaths. Results During the study period, 144 deaths (1.4% of the 10,624 reports), 1,391 patient injuries (13.1%), and 8,061 device malfunctions (75.9%) were reported. The numbers of injury and death events per procedure have stayed relatively constant (mean = 83.4, 95% confidence interval (CI), 74.2-92.7 per 100,000 procedures) over the years. Surgical specialties for which robots are extensively used, such as gynecology and urology, had lower numbers of injuries, deaths, and conversions per procedure than more complex surgeries, such as cardiothoracic and head and neck (106.3 vs. 232.9 per 100,000 procedures, Risk Ratio = 2.2, 95% CI, 1.9-2.6). Device and instrument malfunctions, such as falling of burnt/broken pieces of instruments into the patient (14.7%), electrical arcing of instruments (10.5%), unintended operation of instruments (8.6%), system errors (5%), and video/imaging problems (2.6%), constituted a major part of the reports. Device malfunctions impacted patients in terms of injuries or procedure interruptions. In 1,104 (10.4%) of all the events, the procedure was interrupted to restart the system (3.1%), to convert the procedure to non-robotic techniques (7.3%), or to reschedule it (2.5%). Conclusions Despite widespread adoption of robotic systems for minimally invasive surgery in the U.S., a non-negligible number of technical difficulties and complications are still being experienced during procedures. Adoption of advanced techniques in design and operation of robotic surgical systems and enhanced mechanisms for adverse event reporting may reduce these preventable incidents in the future.
引用
收藏
页数:20
相关论文
共 49 条
[1]   Malfunction and failure of robotic systems during general surgical procedures [J].
Agcaoglu, Orhan ;
Aliyev, Shamil ;
Taskin, Halit Eren ;
Chalikonda, Sricharan ;
Walsh, Matthew ;
Costedio, Meagan M. ;
Kroh, Matthew ;
Rogula, Tomasz ;
Chand, Bipan ;
Gorgun, Emre ;
Siperstein, Allan ;
Berber, Eren .
SURGICAL ENDOSCOPY AND OTHER INTERVENTIONAL TECHNIQUES, 2012, 26 (12) :3580-3583
[2]  
Andonian S, 2008, CAN J UROL, V15, P3912
[3]  
[Anonymous], 2013, INT SURG COMM MED DE
[4]   New Technology and Health Care Costs - The Case of Robot-Assisted Surgery [J].
Barbash, Gabriel I. ;
Glied, Sherry A. .
NEW ENGLAND JOURNAL OF MEDICINE, 2010, 363 (08) :701-704
[5]  
Borden Lester S Jr, 2007, Can J Urol, V14, P3499
[6]   Robotic-assisted versus laparoscopic cholecystectomy - Outcome and cost analyses of a case-matched control study [J].
Breitenstein, Stefan ;
Nocito, Antonio ;
Pithan, Milo ;
Held, Ulrike ;
Weber, Markus ;
Clavien, Pierre-Alain .
ANNALS OF SURGERY, 2008, 247 (06) :987-993
[7]   Reliability of robotic system during general surgical procedures in a university hospital [J].
Buchs, Nicolas C. ;
Pugin, Francois ;
Volonte, Francesco ;
Morel, Philippe .
AMERICAN JOURNAL OF SURGERY, 2014, 207 (01) :84-88
[8]   Malfunction of the da Vinci robotic system in urology [J].
Chen, Cheng-Che ;
Ou, Yen-Chuan ;
Yang, Cheng-Kuang ;
Chiu, Kun-Yuan ;
Wang, Shian-Shiang ;
Su, Chung-Kuang ;
Ho, Hao-Chung ;
Cheng, Chen-Li ;
Chen, Chuan-Shu ;
Lee, Jian-Ri ;
Chen, Wen-Min .
INTERNATIONAL JOURNAL OF UROLOGY, 2012, 19 (08) :736-740
[9]  
Christopher D.Manning., 1999, FDN STAT NATURAL LAN
[10]  
Citron Research, 2013, INT SURG ANG BROK WI