Improved image quality in abdominal computed tomography reconstructed with a novel Deep Learning Image Reconstruction technique - initial clinical experience

被引:15
作者
Njolstad, Tormund [1 ,2 ,3 ]
Schulz, Anselm [1 ,2 ]
Godt, Johannes C. [1 ]
Brogger, Helga M. [1 ]
Johansen, Cathrine K. [1 ]
Andersen, Hilde K. [2 ]
Martinsen, Anne Catrine T. [2 ,4 ]
机构
[1] Oslo Univ Hosp Ulleval, Dept Radiol & Nucl Med, Pb 4950, N-0424 Oslo, Norway
[2] Oslo Univ Hosp, Dept Diagnost Phys, Oslo, Norway
[3] Haukeland Hosp, Dept Radiol, Bergen, Norway
[4] Oslo Metropolitan Univ, Fac Hlth Sci, Oslo, Norway
关键词
Abdominal computed tomography; deep learning image reconstruction; image quality;
D O I
10.1177/20584601211008391
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background: A novel Deep Learning Image Reconstruction (DLIR) technique for computed tomography has recently received clinical approval. Purpose: To assess image quality in abdominal computed tomography reconstructed with DLIR, and compare with standardly applied iterative reconstruction. Material and methods: Ten abdominal computed tomography scans were reconstructed with iterative reconstruction and DLIR of medium and high strength, with 0.625 mm and 2.5 mm slice thickness. Image quality was assessed using eight visual grading criteria in a side-by-side comparative setting. All series were presented twice to evaluate intraobserver agreement. Reader scores were compared using univariate logistic regression. Image noise and contrast-to-noise ratio were calculated for quantitative analyses. Results: For 2.5 mm slice thickness, DLIR images were more frequently perceived as equal or better than iterative reconstruction across all visual grading criteria (for both DLIR of medium and high strength, p < 0.001). Correspondingly, DLIR images were more frequently perceived as better (as opposed to equal or in favor of iterative reconstruction) for visual reproduction of liver parenchyma, intrahepatic vascular structures as well as overall impression of image noise and texture (p < 0.001). This improved image quality was also observed for 0.625 mm slice images reconstructed with DLIR of high strength when directly comparing to traditional iterative reconstruction in 2.5 mm slices. Image noise was significantly lower and contrast-to-noise ratio measurements significantly higher for images reconstructed with DLIR compared to iterative reconstruction (p < 0.01). Conclusions: Abdominal computed tomography images reconstructed using a DLIR technique shows improved image quality when compared to standardly applied iterative reconstruction across a variety of clinical image quality criteria.
引用
收藏
页数:9
相关论文
共 50 条
[31]   Image quality improvement in low-dose chest CT with deep learning image reconstruction [J].
Tian, Qian ;
Li, Xinyu ;
Li, Jianying ;
Cheng, Yannan ;
Niu, Xinyi ;
Zhu, Shumeng ;
Xu, Wenting ;
Guo, Jianxin .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2022, 23 (12)
[32]   Deep learning versus iterative reconstruction on image quality and dose reduction in abdominal CT: a live animal study [J].
Zhang, Jason Z. ;
Ganesh, Halemane ;
Raslau, Flavius D. ;
Nair, Rashmi ;
Escott, Edward ;
Wang, Chi ;
Wang, Ge ;
Zhang, Jie .
PHYSICS IN MEDICINE AND BIOLOGY, 2022, 67 (14)
[33]   Metal artifact reduction combined with deep learning image reconstruction algorithm for CT image quality optimization: a phantom study [J].
Zou, Huachun ;
Wang, Zonghuo ;
Guo, Mengya ;
Peng, Kun ;
Zhou, Jian ;
Zhou, Lili ;
Fan, Bing .
PEERJ, 2025, 13
[34]   Investigation of effect of reconstruction filters on cone-beam computed tomography image quality [J].
Srinivasan, Kavitha ;
Mohammadi, Mohammad ;
Shepherd, Justin .
AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2014, 37 (03) :607-614
[35]   Investigation of effect of reconstruction filters on cone-beam computed tomography image quality [J].
Kavitha Srinivasan ;
Mohammad Mohammadi ;
Justin Shepherd .
Australasian Physical & Engineering Sciences in Medicine, 2014, 37 :607-614
[36]   Virtual Monochromatic Imaging of Half-Iodine-Load, Contrast-Enhanced Computed Tomography with Deep Learning Image Reconstruction in Patients with Renal Insufficiency: A Clinical Pilot Study [J].
Harashima, Shingo ;
Fukui, Rika ;
Samejima, Wakana ;
Hirose, Yuta ;
Kariyasu, Toshiya ;
Nishikawa, Makiko ;
Yamaguchi, Hidenori ;
Machida, Haruhiko .
JOURNAL OF NIPPON MEDICAL SCHOOL, 2025, 92 (01) :69-79
[37]   Improved Image Quality in Dual-Energy Abdominal CT: Comparison of Iterative Reconstruction in Image Space and Filtered Back Projection Reconstruction [J].
Wang, Rui ;
Yu, Wei ;
Wu, Runze ;
Yang, Hua ;
Lu, Dongxu ;
Liu, Jiayi ;
Zhang, Zhaoqi ;
Zhang, Chuanchen .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2012, 199 (02) :402-406
[38]   Comparison of image quality from filtered back projection, statistical iterative reconstruction, and model-based iterative reconstruction algorithms in abdominal computed tomography [J].
Kuo, Yu ;
Lin, Yi-Yang ;
Lee, Rheun-Chuan ;
Lin, Chung-Jung ;
Chiou, Yi-You ;
Guo, Wan-Yuo .
MEDICINE, 2016, 95 (31)
[39]   Assessing the influence of X-ray spectrum on image quality in abdominal computed tomography scans [J].
El Mansouri, M'hamed ;
Talbi, Mohammed ;
Khatar, Zakaria ;
Nhila, Oussama ;
Sekkat, Hamza ;
Tahiri, Zakaria .
RADIATION EFFECTS AND DEFECTS IN SOLIDS, 2024, 179 (9-10) :1178-1187
[40]   Initial Phantom Study Comparing Image Quality in Computed Tomography Using Adaptive Statistical Iterative Reconstruction and New Adaptive Statistical Iterative Reconstruction V [J].
Lim, Kyungjae ;
Kwon, Heejin ;
Cho, Jinhan ;
Oh, Jongyoung ;
Yoon, Seongkuk ;
Kang, Myungjin ;
Ha, Dongho ;
Lee, Jinhwa ;
Kang, Eunju .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2015, 39 (03) :443-448