Treating patients with real-time tumor tracking using the Vero gimbaled linac system: Implementation and first review

被引:98
作者
Depuydt, Tom [1 ,2 ,3 ]
Poels, Kenneth [2 ]
Verellen, Dirk [1 ,2 ]
Engels, Benedikt [1 ]
Collen, Christine [1 ]
Buleteanu, Manuela [1 ]
Van den Begin, Robbe [1 ]
Boussaer, Marlies [1 ]
Duchateau, Michael [1 ]
Gevaert, Thierry [1 ]
Storme, Guy [1 ,2 ]
De Ridder, Mark [1 ,2 ]
机构
[1] Vrije Univ Brussel, UZ Brussel, Dept Radiotherapy, B-1090 Brussels, Belgium
[2] Vrije Univ Brussel, Med Imaging & Phys Sci Grp, Fac Med & Pharm, B-1090 Brussels, Belgium
[3] Univ Hosp Leuven, Dept Radiat Oncol, Leuven, Belgium
关键词
Tumor tracking; Stereotactic body radiation therapy; Organ motion; 4-DIMENSIONAL COMPUTED-TOMOGRAPHY; STEREOTACTIC BODY RADIOTHERAPY; IMAGE-GUIDED RADIOTHERAPY; CELL LUNG-CANCER; TREATMENT MARGINS; MOTION; CYBERKNIFE; ACCURACY; VOLUME;
D O I
10.1016/j.radonc.2014.05.017
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: To report on the first clinical application of a real-time tumor tracking (RTTT) solution based on the Vero SBRT gimbaled linac system for treatment of moving tumors. Methods and materials: A first group of 10 SBRT patients diagnosed with NSCLC or oligometastatic disease in lung or liver was treated with the RTTT technique. The PTV volumes and OAR exposure were benchmarked against the widely used ITV approach. Based on data acquired during execution of RTTT treatments, a first review was performed of the process. Results: The 35% PTV volume reduction with RTTT of the studied single lesions SBRT irradiations of small target volumes is expected to result in a small (<1%) reduction of lung or liver NTCP. A GTV-PTV margin of 5.0 mm was applied for treatment planning of RTTT. From patient data on residual geometric uncertainties, a CTV-PTV margin of 3.2 mm was calculated. Reduction of the GTV-PTV margin below 5.0 mm without better understanding of biological definition of tumor boundaries was discouraged. Total treatment times were reduced to 34.4 min on average. Conclusion: A considerable PTV volume reduction was achieved applying RTTT and time efficiency for respiratory correlated SBRT was reestablished with Vero RTTT. (C) 2014 Elsevier Ireland Ltd. All rights reserved.
引用
收藏
页码:343 / 351
页数:9
相关论文
共 43 条
[1]   Predictive uncertainty in infrared marker-based dynamic tumor tracking with Vero4DRT [J].
Akimoto, Mami ;
Nakamura, Mitsuhiro ;
Mukumoto, Nobutaka ;
Tanabe, Hiroaki ;
Yamada, Masahiro ;
Matsuo, Yukinori ;
Monzen, Hajime ;
Mizowaki, Takashi ;
Kokubo, Masaki ;
Hiraoka, Masahiro .
MEDICAL PHYSICS, 2013, 40 (09)
[2]  
[Anonymous], CLIN STUD PROT RTOG
[3]   DOSE COVERAGE BEYOND THE GROSS TUMOR VOLUME FOR VARIOUS STEREOTACTIC BODY RADIOTHERAPY PLANNING TECHNIQUES REPORTING SIMILAR CONTROL RATES FOR STAGE I NON-SMALL-CELL LUNG CANCER [J].
Arvidson, Noah B. ;
Mehta, Mtnesh P. ;
Tome, Wolfgang A. .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2008, 72 (05) :1597-1603
[4]  
Collins B, 2009, RADIAT ONCOL, V2, P39
[5]   Individualized image guided iso-NTCP based liver cancer SBRT [J].
Dawson, Laura A. ;
Eccles, Cynthia ;
Craig, Tim .
ACTA ONCOLOGICA, 2006, 45 (07) :856-864
[6]   Initial assessment of tumor tracking with a gimbaled linac system in clinical circumstances: A patient simulation study [J].
Depuydt, Tom ;
Poels, Kenneth ;
Verellen, Dirk ;
Engels, Benedikt ;
Collen, Christine ;
Haverbeke, Chloe ;
Gevaert, Thierry ;
Buls, Nico ;
Van Gompel, Gert ;
Reynders, Truus ;
Duchateau, Michael ;
Tournel, Koen ;
Boussaer, Marlies ;
Steenbeke, Femke ;
Vandenbroucke, Frederik ;
De Ridder, Mark .
RADIOTHERAPY AND ONCOLOGY, 2013, 106 (02) :236-240
[7]   Geometric accuracy of a novel gimbals based radiation therapy tumor tracking system [J].
Depuydt, Tom ;
Verellen, Dirk ;
Haas, Olivier ;
Gevaert, Thierry ;
Linthout, Nadine ;
Duchateau, Michael ;
Tournel, Koen ;
Reynders, Truus ;
Leysen, Katrien ;
Hoogeman, Mischa ;
Storme, Guy ;
De Ridder, Mark .
RADIOTHERAPY AND ONCOLOGY, 2011, 98 (03) :365-372
[8]   CONFORMAL ARC RADIOTHERAPY FOR PROSTATE CANCER: INCREASED BIOCHEMICAL FAILURE IN PATIENTS WITH DISTENDED RECTUM ON THE PLANNING COMPUTED TOMOGRAM DESPITE IMAGE GUIDANCE BY IMPLANTED MARKERS [J].
Engels, Benedikt ;
Soete, Guy ;
Verellen, D. ;
Storme, Guy .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2009, 74 (02) :388-391
[9]   Fast Monte Carlo dose calculation for photon beams based on the VMC electron algorithm [J].
Fippel, M .
MEDICAL PHYSICS, 1999, 26 (08) :1466-1475
[10]   Loss of biological effect in prolonged fraction delivery [J].
Fowler, JF ;
Welsh, JS ;
Howard, SP .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2004, 59 (01) :242-249