Bromine and iodine in atmospheric particles or coal can cause environmental problems such as destruction of ozone in the atmosphere: therefore, the presence of these compounds has recently received increased attention. Here, a rapid and reliable method for the simultaneous determination of total bromine and iodine using ICP-MS analysis is described. Samples were dissolved in mixtures of 5 mL of HNO(3) and 2 mL of H(2)O(2) in a high pressure microwave digester. The solution was then oxidized by persulfate (Na(2)S(2)O(8)) in addition to a small amount of silver nitrate, after which the total bromine and iodine were measured simultaneously by ICP-MS. The signal memory effects of bromine and iodine during analysis were effectively decreased by washing with a new mixture agent (2% alcohol acidic solution, pH 1-2 adjusted with HCl). The detection limits for bromine and iodine using this method were about 3.2 mu g L(-1) and 1.1 mu g L(-1), respectively. Additionally, the spike recoveries were between 78.7% and 121% for bromine and iodine analysis, while the relative standard deviations ranged from 4.3% to 9.7%, and from 1.5% to 3.4% for bromine and iodine, respectively. The results of this study indicate that the method described here is suitable for the analysis of micro-amounts of bromine and iodine in atmospheric particles and coal samples. (C) 2009 Elsevier B.V. All rights reserved.