Effects of filter pore size on efficacy of continuous arteriovenous hemofiltration therapy for Staphylococcus aureus-induced septicemia in immature swine

被引:68
作者
Lee, PA
Wegner, GW
Pryor, RW
Matson, JR
机构
[1] Dallas Hosp, Med City, Dept Clin Res, Dallas, TX USA
[2] Dallas Hosp, Med City, Div Pediat Crit Care, Dallas, TX USA
关键词
swine; extracorporeal; mediators; Gram-positive infection; multiple organ failure; molecular weight exclusion;
D O I
10.1097/00003246-199804000-00024
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Objective: To evaluate the effect of hemofilter pore size on the efficacy of continuous arteriovenous hemofiltration (CAVH) in improving morbidity and mortality in an immature swine model of Staphylococcus aureus-induced septicemia. Design: Prospective, randomized study with age-matched controls. Setting: Biomedical research facility. Subjects: Fourteen 4 to 8-wk-old, weaned Poland-China swine, weighing 5 to 10 kg. Interventions: spontaneously breathing, ketamine-sedated swine (4 to 8 wks of age) were given an intravenous lethal dose of live S. aureus, Animals were then filtered with either a 50-kilodalton (kD) pore size filter (control) or a 100-kD pore size filter (experimental). No animals received antibiotics. Measurements and Main Results: Physiologic, biochemical, and hematologic parameters were measured in all animals every 1 to 3 hrs. Animals were monitored continuously and survival time (hr) was recorded (permanent survival = 168 hrs/7 days). Animals filtered with the 100-kD filter survived significantly longer than control animals (103 +/- 18 [SEM] vs. 56 +/- 9 hrs), The 10O kD-filtered group had one permanent survivor (168 hrs), Protein concentration of the ultrafiltrate obtained from the 100-kD-filtered animals was eight-fold higher than control ultrafiltrate. The protein removed did not contain a high percentage of albumin (as determined by autoanalyzer methods). No significant differences were seen in any of the other measured parameters. Conclusions: CAVH significantly improved survival in swine with S. aureus-induced sepsis. The superior performance of the 100-kD filter vs. the 50-kD filter suggests that higher molecular weight mediators that are not removed efficiently by the 50-kD tilter may be responsible for the morbidity and mortality seen in this model of sepsis, These mediators may be removed In greater proportion by our customized (100-kD pore size) filter.
引用
收藏
页码:730 / 737
页数:8
相关论文
共 22 条
[1]   USE OF EXTRACORPOREAL SUPPORTIVE TECHNIQUES AS ADDITIONAL TREATMENT FOR SEPTIC-INDUCED MULTIPLE ORGAN FAILURE PATIENTS [J].
BARZILAY, E ;
KESSLER, D ;
BERLOT, G ;
GULLO, A ;
GEBER, D ;
BENZEEV, I .
CRITICAL CARE MEDICINE, 1989, 17 (07) :634-637
[2]  
Barzilay E, 1987, Rev Esp Anestesiol Reanim, V34, P459
[3]  
BRADFORD MM, 1976, ANAL BIOCHEM, V72, P248, DOI 10.1016/0003-2697(76)90527-3
[4]  
CORAIM F, 1985, CONTINUOUS ARTERIOVE, P103
[5]  
Coraim F., 1985, CONTINUOUS ARTERIOVE, P116
[6]  
GOLDFARB S, 1994, J AM SOC NEPHROL, V5, P228
[7]   SEQUENTIAL HEMOFILTRATION IN NONOLIGURIC HIGH CAPILLARY-PERMEABILITY PULMONARY-EDEMA OF SEVERE SEPSIS - PRELIMINARY-REPORT [J].
GOTLOIB, L ;
BARZILAY, E ;
SHUSTAK, A ;
LEV, A .
CRITICAL CARE MEDICINE, 1984, 12 (11) :997-1000
[8]   HEMOFILTRATION IN SEPTIC ARDS - THE ARTIFICIAL-KIDNEY AS AN ARTIFICIAL ENDOCRINE LUNG [J].
GOTLOIB, L ;
BARZILAY, E ;
SHUSTAK, A ;
WAIS, Z ;
JAICHENKO, J ;
LEV, A .
RESUSCITATION, 1986, 13 (02) :123-132
[9]  
HINSHAW LB, 1988, CIRC SHOCK, V26, P257
[10]  