Use of a plastic scintillator detector for patient-specific quality assurance of VMAT SRS

被引:9
作者
Snyder, Jesse D. [1 ]
Sullivan, Rodney J. [1 ]
Wu, Xingen [1 ]
Covington, Elizabeth L. [1 ]
Popple, Richard A. [1 ]
机构
[1] Univ Alabama Birmingham, Dept Radiat Oncol, Birmingham, AL 35294 USA
来源
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS | 2019年 / 20卷 / 09期
关键词
dosimetry; quality assurance; radiosurgery; scintillator; volumetric-modulated arc therapy; ASTRO PRACTICE GUIDELINE; ENERGY BEAM DOSIMETRY; AMERICAN SOCIETY; RADIOLOGY ACR; PLAN QUALITY; RADIATION; RADIOSURGERY; DELIVERY; COLLEGE; IMRT;
D O I
10.1002/acm2.12705
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose To evaluate a scintillator detector for patient-specific quality assurance of VMAT radiosurgery plans. Methods The detector was comprised of a 1 mm diameter, 1 mm high scintillator coupled to an acrylic optical fiber. Sixty VMAT SRS plans for treatment of single targets having sizes ranging from 3 mm to 30.2 mm equivalent diameter (median 16.3 mm) were selected. The plans were delivered to a 20 cm x 20 cm x 15 cm water equivalent plastic phantom having either the scintillator detector or radiochromic film at the center. Calibration films were obtained for each measurement session. The films were scanned and converted to dose using a 3-channel technique. Results The mean difference between scintillator and film was -0.45% (95% confidence interval -0.1% to 0.8%). For target equivalent diameter smaller than the median, the mean difference was 1.1% (95% confidence interval 0.5% to 1.7%). For targets larger than the median, the mean difference was -0.2% (95% confidence interval -0.7% to 0.1%). Conclusions The scintillator detector response is independent of target size for targets as small as 3 mm and is well-suited for patient-specific quality assurance of VMAT SRS plans. Further work is needed to evaluate the accuracy for VMAT plans that treat multiple targets using a single isocenter.
引用
收藏
页码:143 / 148
页数:6
相关论文
共 16 条
  • [1] WATER-EQUIVALENT PLASTIC SCINTILLATION DETECTORS FOR HIGH-ENERGY BEAM DOSIMETRY .2. PROPERTIES AND MEASUREMENTS
    BEDDAR, AS
    MACKIE, TR
    ATTIX, FH
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1992, 37 (10) : 1901 - 1913
  • [2] WATER-EQUIVALENT PLASTIC SCINTILLATION DETECTORS FOR HIGH-ENERGY BEAM DOSIMETRY .1. PHYSICAL CHARACTERISTICS AND THEORETICAL CONSIDERATIONS
    BEDDAR, AS
    MACKIE, TR
    ATTIX, FH
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1992, 37 (10) : 1883 - 1900
  • [3] The practical application of scintillation dosimetry in small-field photon-beam radiotherapy
    Burke, Elisa
    Poppinga, Daniela
    Schoenfeld, Andreas A.
    Harder, Dietrich
    Poppe, Bjoern
    Looe, Hui Khee
    [J]. ZEITSCHRIFT FUR MEDIZINISCHE PHYSIK, 2017, 27 (04): : 324 - 333
  • [4] Plan quality and treatment planning technique for single isocenter cranial radiosurgery with volumetric modulated arc therapy
    Clark, Grant M.
    Popple, Richard A.
    Prendergast, Brendan M.
    Spencer, Sharon A.
    Thomas, Evan M.
    Stewart, John G.
    Guthrie, Barton L.
    Markert, James M.
    Fiveash, John B.
    [J]. PRACTICAL RADIATION ONCOLOGY, 2012, 2 (04) : 306 - 313
  • [5] Guidance document on delivery, treatment planning, and clinical implementation of IMRT: Report of the IMRT subcommittee of the AAPM radiation therapy committee
    Ezzell, GA
    Galvin, JM
    Low, D
    Palta, JR
    Rosen, I
    Sharpe, MB
    Xia, P
    Xiao, Y
    Xing, L
    Yu, CX
    [J]. MEDICAL PHYSICS, 2003, 30 (08) : 2089 - 2115
  • [6] Spectral discrimination of Cerenkov radiation in scintillating dosimeters
    Frelin, AM
    Fontbonne, JM
    Ban, G
    Colin, J
    Labalme, M
    Batalla, A
    Isambert, A
    Vela, A
    Leroux, T
    [J]. MEDICAL PHYSICS, 2005, 32 (09) : 3000 - 3006
  • [7] Spectral method for the correction of the Cerenkov light effect in plastic scintillation detectors: A comparison study of calibration procedures and validation in Cerenkov light-dominated situations
    Guillot, Mathieu
    Gingras, Luc
    Archambault, Louis
    Beddar, Sam
    Beaulieu, Luc
    [J]. MEDICAL PHYSICS, 2011, 38 (04) : 2140 - 2150
  • [8] American College of Radiology (ACR) and American Society for Radiation Oncology (ASTRO) Practice Guideline for Intensity-modulated Radiation Therapy (IMRT)
    Hartford, Alan C.
    Galvin, James M.
    Beyer, David C.
    Eichler, Thomas J.
    Ibbott, Geoffrey S.
    Kavanagh, Brian
    Schultz, Christopher J.
    Rosenthal, Seth A.
    [J]. AMERICAN JOURNAL OF CLINICAL ONCOLOGY-CANCER CLINICAL TRIALS, 2012, 35 (06): : 612 - 617
  • [9] Multichannel film dosimetry with nonuniformity correction
    Micke, Andre
    Lewis, David F.
    Yu, Xiang
    [J]. MEDICAL PHYSICS, 2011, 38 (05) : 2523 - 2534
  • [10] Safety considerations for IMRT: Executive summary
    Moran, Jean M.
    Dempsey, Melanie
    Eisbruch, Avraham
    Fraass, Benedick A.
    Galvin, James M.
    Ibbott, Geoffrey S.
    Marks, Lawrence B.
    [J]. MEDICAL PHYSICS, 2011, 38 (09) : 5067 - 5072