Evanescent wave fiber optic fluorosensor: effect of tapering configuration on the signal acquisition

被引:43
作者
Nath, N [1 ]
Anand, S [1 ]
机构
[1] Indian Inst Technol, Ctr Biomed Engn, New Delhi 110016, India
关键词
evanescent wave; fiber-optic immunosensor; tapered fibers;
D O I
10.1117/1.601709
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
A comparative study of the signal acquisition from straight fiber and tapered optical fibers has been analyzed theoretically for the design of an evanescent wave fiber optic immunosensor. It has been shown that tapered optical fibers have higher sensitivity compared to straight fibers. Among the tapered fibers combination tapered fibers were also found to have higher sensitivity than continuous tapered fibers. Theoretical analysis indicates that an increase in signal is a result of three factors, namely, V number matching, an increase in evanescent absorption by the fluorescent medium, and an increase in fluorescence emission coupling into the fiber. Maximum sensitivity is obtained when the radius of the fiber is reduced to the V number matching radius within 1 cm from the start of the taper with rest of the fiber remaining untapered. Also, a higher signal level is obtained when the launching light fills the numerical aperture determined by the tapering ratio of the fiber. (C) 1998 Society of Photo-Optical Instrumentation Engineers. [S0091-3286(98)04301-3].
引用
收藏
页码:220 / 228
页数:9
相关论文
共 15 条
[1]   DEVELOPMENT OF AN EVANESCENT-WAVE FIBER OPTIC BIOSENSOR [J].
ANDERSON, GP ;
GOLDEN, JP ;
CAO, LK ;
WIJESURIYA, D ;
SHRIVERLAKE, LC ;
LIGLER, FS .
IEEE ENGINEERING IN MEDICINE AND BIOLOGY MAGAZINE, 1994, 13 (03) :358-368
[2]   A FIBER OPTIC BIOSENSOR - COMBINATION TAPERED FIBERS DESIGNED FOR IMPROVED SIGNAL ACQUISITION [J].
ANDERSON, GP ;
GOLDEN, JP ;
LIGLER, FS .
BIOSENSORS & BIOELECTRONICS, 1993, 8 (05) :249-256
[3]   SLOWLY VARYING OPTICAL FIBERS [J].
ANKIEWICZ, A ;
PASK, C ;
SNYDER, AW .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA, 1982, 72 (02) :198-203
[4]  
Boisde G, 1996, CHEM BIOCH SENSING O
[5]   Tapered fiber tips for fiber optic biosensors [J].
Gao, HH ;
Chen, ZP ;
Kumar, J ;
Tripathy, SK ;
Kaplan, DL .
OPTICAL ENGINEERING, 1995, 34 (12) :3465-3470
[6]   EFFECT OF NUMERICAL APERTURE ON SIGNAL LEVEL IN CYLINDRICAL WAVE-GUIDE EVANESCENT FLUOROSENSORS [J].
GLASS, TR ;
LACKIE, S ;
HIRSCHFELD, T .
APPLIED OPTICS, 1987, 26 (11) :2181-2187
[7]   WEAKLY GUIDING FIBERS [J].
GLOGE, D .
APPLIED OPTICS, 1971, 10 (10) :2252-&
[8]   FLUOROMETER AND TAPERED FIBER OPTIC PROBES FOR SENSING IN THE EVANESCENT WAVE [J].
GOLDEN, JP ;
SHRIVERLAKE, LC ;
ANDERSON, GP ;
THOMPSON, RB ;
LIGLER, FS .
OPTICAL ENGINEERING, 1992, 31 (07) :1458-1462
[9]  
LOVE WF, 1992, BIOSENSORS FIBRE OPT, P139
[10]   ENHANCED EVANESCENT WAVE SENSORS BASED ON SOL GEL-DERIVED POROUS-GLASS COATINGS [J].
MACCRAITH, BD .
SENSORS AND ACTUATORS B-CHEMICAL, 1993, 11 (1-3) :29-34