Negative Impedance Capacitive Electrode for ECG Sensing Through Fabric Layer

被引:19
作者
Wang, Ting-Wei [1 ]
Lin, Shien-Fong [1 ]
机构
[1] Natl Chiao Tung Univ, Inst Biomed Engn, Hsinchu 30010, Taiwan
关键词
Capacitive coupling; negative impedance capacitive electrode; noncontact electrocardiogram (ECG) measurement;
D O I
10.1109/TIM.2020.3045187
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Many applications utilize flexible capacitive electrode implementations and analog front-end optimizations for improving noncontact electrocardiogram (ECG) measurements. However, the influence of the negative impedance in the circuit front end on noncontact measurements has not been investigated. The objective of this study was to develop negative impedance capacitive electrodes for improving the degraded noncontact ECG signals induced by the high-impedance interface between the skin and electrode. A capacitive electrode with a negative impedance of -200 M Omega was designed to reduce the impedance mismatch between the body and measurement electrode. The proposed electrodes were applied on a subject's chest to continuously measure the cardiac signals through a fabric layer and further evaluate the signal-to-noise (SNR) performance for noncontact ECG measurements. The negative impedance in the circuit front end reduced the coupling effect between the skin and the measurement electrodes, thus improved the SNR in theory and in practice. Importantly, our system produced a distinguishable heartbeat signal even with a maximum cloth thickness of 1.8 mm and skin-electrode capacitance exceeding 57 pF, which is suitable for wearable applications. This study is a pilot effort toward validating the efficacy of negative impedance capacitive electrodes for noncontact physiological measurements. This novel electrode design is expected to potentially address the limitations imposed by a fabric layer and its thickness in noncontact hioelectric sensing in the future.
引用
收藏
页数:8
相关论文
共 22 条
  • [1] Conductive Polymer Foam Surface Improves the Performance of a Capacitive EEG Electrode
    Baek, Hyun Jae
    Lee, Hong Ji
    Lim, Yong Gyu
    Park, Kwang Suk
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2012, 59 (12) : 3422 - 3431
  • [2] The power output and efficiency of a negative capacitance shunt for vibration control of a flexural system
    Beck, Benjamin S.
    Cunefare, Kenneth A.
    Collet, Manuel
    [J]. SMART MATERIALS AND STRUCTURES, 2013, 22 (06)
  • [3] Charlton P.H., 2016, Secondary Anal. Electron. Health Rec., P377, DOI [DOI 10.1007/978-3-319-43742-2_26, 10.1007/978-3-319-43742-2]
  • [4] An assessment of algorithms to estimate respiratory rate from the electrocardiogram and photoplethysmogram
    Charlton, Peter H.
    Bonnici, Timothy
    Tarassenko, Lionel
    Clifton, David A.
    Beale, Richard
    Watkinson, Peter J.
    [J]. PHYSIOLOGICAL MEASUREMENT, 2016, 37 (04) : 610 - 626
  • [5] A 400 GΩ Input-Impedance Active Electrode for Non-Contact Capacitively Coupled ECG Acquisition With Large Linear-Input-Range and High CM-Interference-Tolerance
    Chen, Mingyi
    Chun, Ho Sung
    Castro, Ivan Dario
    Torfs, Tom
    Lin, Qiuyang
    van Hoof, Chris
    Wang, Guoxing
    Lian, Yong
    van Helleputte, Nick
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL CIRCUITS AND SYSTEMS, 2019, 13 (02) : 376 - 386
  • [6] Chi Yu Mike, 2010, IEEE Rev Biomed Eng, V3, P106, DOI 10.1109/RBME.2010.2084078
  • [7] Heart Monitor Using Flexible Capacitive ECG Electrodes
    Gao, Yang
    Soman, Varun V.
    Lombardi, Jack P.
    Rajbhandari, Pravakar Prasad
    Dhakal, Tara P.
    Wilson, Dale G.
    Poliks, Mark D.
    Ghose, Kanad
    Turner, James N.
    Jin, Zhanpeng
    [J]. IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2020, 69 (07) : 4314 - 4323
  • [8] Novel Dry Polymer Foam Electrodes for Long-Term EEG Measurement
    Lin, Chin-Teng
    Liao, Lun-De
    Liu, Yu-Hang
    Wang, I-Jan
    Lin, Bor-Shyh
    Chang, Jyh-Yeong
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2011, 58 (05) : 1200 - 1207
  • [9] Master equation for operational amplifiers: stability of negative differential converters, crossover frequency and pass-bandwidth
    Mishonov, T. M.
    Danchev, V., I
    Petkov, E. G.
    Gourev, V. N.
    Dimitrova, I. M.
    Serafimov, N. S.
    Stefanov, A. A.
    Varonov, A. M.
    [J]. JOURNAL OF PHYSICS COMMUNICATIONS, 2019, 3 (03):
  • [10] Characterization of Textile-Insulated Capacitive Biosensors
    Ng, Charn Loong