The effect of clearance in total hip endoprostheses

被引:1
作者
Munteanu, Florin [1 ]
Sirbu, Paul [2 ]
Barbarosie, Constantin [3 ]
Munteanu, Anca [2 ]
Botez, Paul [2 ]
机构
[1] Univ Med & Pharm Iasi, Fac Med Bioengn, Iasi, Romania
[2] Univ Med & Pharm Iasi, Fac Med, Iasi, Romania
[3] Univ Med & Pharm Iasi, Clin Rehabil Hosp Iasi, Iasi, Romania
来源
AT-EQUAL 2009: 2009 ECSIS SYMPOSIUM ON ADVANCED TECHNOLOGIES FOR ENHANCED QUALITY OF LIFE: LAB-RS AND ARTIPED 2009 | 2009年
关键词
D O I
10.1109/AT-EQUAL.2009.48
中图分类号
TP18 [人工智能理论];
学科分类号
081104 ; 0812 ; 0835 ; 1405 ;
摘要
The aim of this paper is to establish a theoretical experimental method which would emphasize the importance of radial clearance between the cotyloid cup and the femoral head in total hip endoprostheses. Three UHMWPE acetabular cups of different diameters, with the same diameter of the femoral head (CoCrMo), were tested on equipment that determines the tribologic parameters of the biomaterials used in total hip endprostheses. The results showed minimum wear for a 0.1-0.15mm radial clearance and maximum wear at clearance values close to 0, or higher than 0.5mm.