Daily online localization using implanted fiducial markers and its impact on planning target volume for carcinoma prostate

被引:9
作者
Khosa, Robin [1 ]
Nangia, Sapna [2 ]
Chufal, Kundan S. [1 ]
Ghosh, D. [1 ]
Kaul, Rakesh [3 ]
Sharma, Lalit [1 ]
机构
[1] Batra Hosp & Med Res Ctr, Batra Canc Ctr, New Delhi 110062, India
[2] Fortis Hosp, Noida, India
[3] Max Hosp, New Delhi, India
关键词
Daily online localization; prostate internal organ motion; prostate planning target volume; 3-DIMENSIONAL CONFORMAL RADIOTHERAPY; TREATMENT MARGINS; ORGAN MOTION; ERRORS; UNCERTAINTY; CANCER;
D O I
10.4103/0973-1482.65244
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background: Aim of the study was to assess prostate motion on daily basis with respect to setup and to compare the shifts based on bony anatomy and gold fiducial markers. Materials and Methods: Gold fiducial markers were inserted in prostate under U/S guidance and daily portal images were taken and compared with digitally reconstructed images, both using bony landmarks and fiducial markers as reference. A dose of 2 MU was given for two orthogonal images daily. The mean and standard deviation of displacement using gold seeds and bone were calculated. Systematic and random errors were generated. The planning target volume (PTV) was calculated using the Van Herk formula. Results: A total of 180 portal images from 10 patients were studied. The mean displacement along x, y and z axes was 1.67 mm, 3.58 mm, and 1.76 mm using fiducial markers and 2.12 mm, 3.47 mm, and 2.09 mm using bony landmarks, respectively. The mean internal organ motion was 1.23 mm (+1.45), 3.11 mm (+2.69 mm); and 1.87 mm (+1.67 mm) along x, y and z axes, respectively. The PTV to account for prostate motion if daily matching was not done was 4.64 mm, 10.41 mm and 4.40 mm along lateral, superoinferior, and anteroposterior directions, respectively. If bony landmarks were used for daily matching, margins of 3.61 mm, 7.31 mm, and 4.72 mm in lateral, superoinferior, and anteroposterior directions should be added to the clinical target volume. Conclusion: Daily alignment using gold fiducial markers is an effective method of localizing prostate displacement. It provides the option of reducing margins, thus limiting normal tissue toxicity and allowing the possibility of dose escalation for better long-term control.
引用
收藏
页码:172 / 178
页数:7
相关论文
共 17 条
[1]   Portal imaging for evaluation of daily on-line setup errors and off-line organ motion during conformal irradiation of carcinoma of the prostate [J].
Alasti, H ;
Petric, MP ;
Catton, CN ;
Warde, PR .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2001, 49 (03) :869-884
[2]   Variation in prostate position relative to adjacent bony anatomy [J].
Althof, VGM ;
Hoekstra, CJM ;
Loo, HJT .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1996, 34 (03) :709-715
[3]   Prostate target volume variations during a course of radiotherapy [J].
Antolak, JA ;
Rosen, II ;
Childress, CH ;
Zagars, GK ;
Pollack, A .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1998, 42 (03) :661-672
[4]   PROSTATE MOTION DURING STANDARD RADIOTHERAPY AS ASSESSED BY FIDUCIAL MARKERS [J].
CROOK, JM ;
RAYMOND, Y ;
SALHANI, D ;
YANG, H ;
ESCHE, B .
RADIOTHERAPY AND ONCOLOGY, 1995, 37 (01) :35-42
[5]   Measurement of patient positioning errors in three-dimensional conformal radiotherapy of the prostate [J].
Hanley, J ;
Lumley, MA ;
Mageras, GS ;
Sun, J ;
Zelefsky, MJ ;
Leibel, SA ;
Fuks, Z ;
Kutcher, GJ .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1997, 37 (02) :435-444
[6]   Technical aspects of daily online positioning of the prostate for three-dimensional conformal radiotherapy using an electronic portal imaging device [J].
Herman, MG ;
Pisansky, TM ;
Kruse, JJ ;
Prisciandaro, JI ;
Davis, BJ ;
King, BF .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2003, 57 (04) :1131-1140
[7]  
International Commission on Radiation Units and Measurements, 2000, 62 ICRU
[8]   Use of portal images and bat ultrasonography to measure setup error and organ motion for prostate IMRT: Implications for treatment margins [J].
Little, DJ ;
Dong, L ;
Levy, LB ;
Chandra, A ;
Kuban, DA .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2003, 56 (05) :1218-1224
[9]   Comparison of localization performance with implanted fiducial markers and cone-beam computed tomography for on-line image-guided radiotherapy of the prostate [J].
Moseley, Douglas J. ;
White, Elizabeth A. ;
Wiltshire, Kirsty L. ;
Rosewall, Tara ;
Sharpe, Michael B. ;
Siewerdsen, Jeffrey H. ;
Bissonnette, Jean-Pierre ;
Gospodarowicz, Mary ;
Warde, Padraig ;
Catton, Charles N. ;
Jaffray, David A. .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2007, 67 (03) :942-953
[10]   Measurements and clinical consequences of prostate motion during a radiotherapy fraction [J].
Nederveen, AJ ;
van der Heide, UA ;
Dehnad, H ;
van Moorselaar, RJA ;
Hofman, P ;
Lagendijk, JJW .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2002, 53 (01) :206-214