Comparative performance study of ICP mass spectrometers by means of U "isotopic measurements'"

被引:30
作者
Quétel, CR [1 ]
Vogl, J [1 ]
Prohaska, T [1 ]
Nelms, S [1 ]
Taylor, PDP [1 ]
De Bièvre, P [1 ]
机构
[1] Commiss European Communities, Joint Res Ctr, Inst Reference Mat & Measurements, Isotope Measurements Unit, B-2440 Geel, Belgium
来源
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY | 2000年 / 368卷 / 2-3期
关键词
D O I
10.1007/s002160000499
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The performance of four commercially available ICPMS instruments of three different types was compared by means of uranium "isotopic measurements". Examined were two quadrupole sector (different generation, different manufacturer), one single detector double focusing magnetic sector and one multiple collector double focusing magnetic sector instruments. The same samples of the IRMM-072 series were used under routine conditions to measure the U-233/U-235 and the U-233/U-238 ratios which, in these samples, vary over almost three orders of magnitude from similar to 1 to similar to 2 (.) 10(-3). Within expanded (k = 2) uncertainties, good agreement was observed between the certified values and the data internally corrected for mass-discrimination effects. The magnitude of the evaluated uncertainties was different for each type of instrument. With the multiple collector instrument, expanded uncertainties varied from +/- 0.04% to+/- 0.24% for the U-233/U-235 ratio, and from +/- 0.08% to +/- 0.27% for the U-233/U-238 ratio. They were similar to 1 to 5 times larger with the single detector magnetic sector instrument, and similar to 10 to 25 times larger with both quadrupole sector instruments. With the multiple collector instrument, repeatability of the measurements seemed to be limited by the difficulty of correcting properly for instrumental background, whereas with the single detector magnetic sector instrument the counting statistics was the only limitation (on smallest ratios). Apparent mass-discrimination was clearly found to be larger but more reproducible (and hence easier to correct for) in the case of magnetic sector instruments than for both quadrupole: sector instruments. If space charge effects were the main source of mass-discrimination for all instruments, these results are in contradiction with the hypothesis of the size of mass-discrimination decreasing with the acceleration voltage. With the single detector magnetic sector instrument in particular (when operated by changing the ion energy only), our results pointed at more than only one major source of mass-discrimination with variable size depending on the ratios measured.
引用
收藏
页码:148 / 155
页数:8
相关论文
共 24 条
[1]  
[Anonymous], 1995, GUID EXPR UNC MEAS
[2]   Characterisation and correction of instrumental bias in inductively coupled plasma quadrupole mass spectrometry for accurate measurement of lead isotope ratios [J].
Begley, IS ;
Sharp, BL .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1997, 12 (04) :395-402
[3]   OCCURRENCE AND REDUCTION OF NOISE IN INDUCTIVELY-COUPLED PLASMA-MASS SPECTROMETRY FOR ENHANCED PRECISION IN ISOTOPE RATIO MEASUREMENT [J].
BEGLEY, IS ;
SHARP, BL .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1994, 9 (03) :171-176
[4]   A new variable dispersion double-focusing plasma mass spectrometer with performance illustrated for Pb isotopes [J].
Belshaw, NS ;
Freedman, PA ;
O'Nions, RK ;
Frank, M ;
Guo, Y .
INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 1998, 181 :51-58
[5]  
BRUNNEE C, 1987, INT J MASS SPECTROM, V76, P125, DOI 10.1016/0168-1176(87)80030-7
[6]   Space charge evaluation in a plasma-source mass spectrograph [J].
Burgoyne, TW ;
Hieftje, GM ;
Hites, RA .
ANALYTICAL CHEMISTRY, 1997, 69 (03) :485-489
[7]   HIGH-PRECISION URANIUM, THORIUM AND RADIUM ISOTOPE RATIO MEASUREMENTS BY HIGH DYNAMIC-RANGE THERMAL IONIZATION MASS-SPECTROMETRY [J].
COHEN, AS ;
BELSHAW, NS ;
ONIONS, RK .
INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 1992, 116 (01) :71-81
[8]  
DAWSON PH, 1995, QUADRUPOLE MASS SPEC, P143
[9]  
*EURACHEM, 1995, QUANT UNC AN MEAS
[10]   NONSPECTROSCOPIC INTERELEMENT INTERFERENCES IN INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY [J].
GILLSON, GR ;
DOUGLAS, DJ ;
FULFORD, JE ;
HALLIGAN, KW ;
TANNER, SD .
ANALYTICAL CHEMISTRY, 1988, 60 (14) :1472-1474