COMPARISON OF PADC NEUTRON DETECTORS FROM DIFFERENT SUPPLIERS
被引:6
作者:
Caresana, M.
论文数: 0引用数: 0
h-index: 0
机构:
Politecn Milan, Dipartimento Energia, Via La Masa 34, I-20156 Milan, ItalyPolitecn Milan, Dipartimento Energia, Via La Masa 34, I-20156 Milan, Italy
Caresana, M.
[1
]
Parravicini, A.
论文数: 0引用数: 0
h-index: 0
机构:
MI AM Srl, Via De Amicis 5, Fabbiano, ItalyPolitecn Milan, Dipartimento Energia, Via La Masa 34, I-20156 Milan, Italy
Parravicini, A.
[2
]
Coria, S.
论文数: 0引用数: 0
h-index: 0
机构:
MI AM Srl, Via De Amicis 5, Fabbiano, ItalyPolitecn Milan, Dipartimento Energia, Via La Masa 34, I-20156 Milan, Italy
Coria, S.
[2
]
Stabilini, A.
论文数: 0引用数: 0
h-index: 0
机构:
Politecn Milan, Dipartimento Energia, Via La Masa 34, I-20156 Milan, Italy
Paul Scherrer Inst, Dept Radiat Safety & Secur, CH-5232 Villigen, SwitzerlandPolitecn Milan, Dipartimento Energia, Via La Masa 34, I-20156 Milan, Italy
Stabilini, A.
[1
,3
]
Assenmacher, F.
论文数: 0引用数: 0
h-index: 0
机构:
Paul Scherrer Inst, Dept Radiat Safety & Secur, CH-5232 Villigen, SwitzerlandPolitecn Milan, Dipartimento Energia, Via La Masa 34, I-20156 Milan, Italy
Assenmacher, F.
[3
]
Yukihara, E. G.
论文数: 0引用数: 0
h-index: 0
机构:
Paul Scherrer Inst, Dept Radiat Safety & Secur, CH-5232 Villigen, Switzerland
Oklahoma State Univ, Dept Phys, Stillwater, OK 74078 USAPolitecn Milan, Dipartimento Energia, Via La Masa 34, I-20156 Milan, Italy
Yukihara, E. G.
[3
,4
]
机构:
[1] Politecn Milan, Dipartimento Energia, Via La Masa 34, I-20156 Milan, Italy
The objective of this study was to compare poly-allyl diglycol carbonate (PADC) track detectors from different suppliers for linear energy transfer (LET) spectrometry and neutron dosimetry. PADCs are commonly used for passive personal neutron dosimetry, where a common approach is to couple the PADC with a plastic radiator to generate secondary charged particles. The neutron dose can be estimated using the track density or the average LET of the secondary particles. The characterisation of PADC in terms of LET spectrometric capability and neutron sensitivity was previously performed using PADC from Intercast S.r.l., Parma, Italy, but this company stopped the production. Since it is common experience that material from different suppliers can have different properties, it became necessary to perform a comparison of PADCs from different suppliers with the Intercast material. The study permits to compare the reading procedures used at Politecnico di Milano and at the Paul Scherrer Institute.