100 OHM transfer standard

被引:1
|
作者
Satrapinski, A [1 ]
Seppä, H [1 ]
机构
[1] VTT Automat, Measurement Technol, FIN-02044 Espoo, Finland
关键词
D O I
10.1109/CPEM.2000.850897
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
A new 100 Omega transfer standard has been built and placed in service at VTT Automation. The temperature of the resistor is maintained at 30.72 degrees C where the temperature coefficient is zero. The daily variation in temperature is less than +/- 0.005 degrees C. The temperature during transportation is about 0.06 degrees C less than the nominal value and it can be hold for about 40 hours. The behavior of the standard was verified in the resistance comparison between VTT (Finland), PTB (Germany), BIPM (International) and BNM-LCIE (France) in November - December 1998. The result indicates that the variation of the standard during the transportation is less than 6.10(-9).
引用
收藏
页码:102 / 103
页数:2
相关论文
共 50 条
  • [1] Quick comparison of QHR systems using a 100 OHM transfer resistor
    Satrapinski, A
    Seppä, H
    Schumacher, B
    Warnecke, P
    Delahaye, F
    Poirier, W
    Piquemal, F
    2000 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS DIGEST, 2000, : 560 - 561
  • [2] Development of a precision one ohm resistance standard
    Grossenbacher, AC
    1998 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS DIGEST, 1998, : 520 - 521
  • [3] A comparison of the prototype standard measurements of the international ohm
    Jouaust, R
    COMPTES RENDUS HEBDOMADAIRES DES SEANCES DE L ACADEMIE DES SCIENCES, 1922, 175 : 94 - 96
  • [4] STATE PRIMARY STANDARD OF UNIT OF ELECTRIC RESISTANCE - OHM
    KROTKOV, IN
    SHIGORIN, VP
    MEASUREMENT TECHNIQUES-USSR, 1971, 14 (12): : 1791 - &
  • [5] Construction and evaluation of 100 MΩ resistance transfer standard
    Lindic, M
    2000 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS DIGEST, 2000, : 291 - 292
  • [6] ESTABLISHMENT OF A CHECKING STANDARD AND NEW WORKING STANDARDS OF AN OHM
    SHIGORIN, VP
    LOPATNIK.AN
    MEASUREMENT TECHNIQUES-USSR, 1967, (06): : 697 - &
  • [7] A-D CONVERTER LINEARIZES 100-OHM TEMPERATURE DETECTOR
    PARISE, A
    ELECTRONICS, 1982, 55 (18): : 143 - 143
  • [8] Calibration of 1 Ohm Standard Resistor in National Metrology Institute of Malaysia (NMIM)
    Ishak, Nirul Trwan
    Ibarahim, Zahari
    2017 UKM FST POSTGRADUATE COLLOQUIUM, 2018, 1940
  • [9] Improved system for the automatic calibration of standard resistors in the meg-ohm range
    Tadros, N. N.
    Ali, Rasha S. M.
    MEASUREMENT, 2013, 46 (07) : 2077 - 2081
  • [10] Calibration of 1 Ohm based on QHR standard using a commercial DCC bridge
    Yu, KM
    Kim, KT
    Song, YS
    Ihm, G
    2002 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS, CONFERENCE DIGEST, 2002, : 60 - 61