Effect of different saline chaser volumes and flow rates on intravascular contrast enhancement in CT using a circulation phantom

被引:33
作者
Behrendt, Florian F. [1 ]
Bruners, Philipp [1 ]
Keil, Sebastian [1 ]
Plumhans, Cedric [1 ]
Mahnken, Andreas H. [1 ]
Das, Marco [1 ]
Ackermann, Diana [2 ]
Guenther, Rolf W. [1 ]
Muehlenbruch, Georg [1 ]
机构
[1] Univ Hosp RWTH Aachen, Dept Diagnost & Intervent Radiol, D-52057 Aachen, Germany
[2] Univ Hosp RWTH Aachen, Inst Med Stat, D-52057 Aachen, Germany
关键词
Saline chaser; Contrast media; Phantom; CT; COMPUTED-TOMOGRAPHY; PULMONARY-ARTERY; PEAK ENHANCEMENT; POWER INJECTOR; ABDOMINAL CT; FLUSH; ANGIOGRAPHY; MODEL; REDUCTION; AORTA;
D O I
10.1016/j.ejrad.2009.01.008
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To evaluate the influence of different saline chaser volumes and different saline chaser flow rates on the intravascular contrast enhancement in MDCT. Materials and methods: In a physiological flow phantom contrast medium (120 ml, 300 mgl/ml, Ultravist 300) was administered at a flow rate of 6 ml/s followed by different saline chaser volumes (0, 30, 60 and 90 ml) at the same injection rate or followed by a 30-ml saline chaser at different injection rates (2, 4, 6 and 8 ml/s). Serial CT-scans at a level covering the pulmonary artery, the ascending and the descending aorta replica were obtained. Time-enhancement curves were computed and both pulmonary and aortic peak enhancement and peak time were determined. Results: Compared to contrast medium injection without a saline chaser the pushing with a saline chaser (30, 60, and 90 ml) resulted in a statistically significant increased pulmonary peak enhancement (all p = 0.008) and prolonged peak time (p = 0.032, p = 0.024 and p = 0.008, respectively). Highest aortic peak enhancement values were detected for a saline chaser volume of 30 ml. A saline chaser flow rate of 8 ml/s resulted in the highest pulmonary peak enhancement values compared to flow rates of 2, 4 and Omits (all p = 0.008). Aortic peak enhancement showed the highest values for a flow rate of 6 ml/s. Conclusion: A saline chaser volume of 30 ml and an injection rate of 6 ml/s are sufficient to best improve vascular contrast enhancement in the pulmonary artery and the aorta in MDCT. (C) 2009 Published by Elsevier Ireland Ltd.
引用
收藏
页码:688 / 693
页数:6
相关论文
共 17 条
[1]   Simulation of aortic peak enhancement on MDCT using a contrast material flow phantom: Feasibility study [J].
Awai, K ;
Hatcho, A ;
Nakayama, Y ;
Kusunoki, S ;
Liu, D ;
Hatemura, M ;
Funama, Y ;
Denbo, M ;
Sato, N ;
Yamashita, Y .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2006, 186 (02) :379-385
[2]   Aortic and hepatic contrast medium enhancement at CT part I. Prediction with a computer model [J].
Bae, KT ;
Heiken, JP ;
Brink, JA .
RADIOLOGY, 1998, 207 (03) :647-655
[3]   Aortic and hepatic peak enhancement at CT: Effect of contrast medium injection rate - Pharmacokinetic analysis and experimental porcine model [J].
Bae, KT ;
Heiken, JP ;
Brink, JA .
RADIOLOGY, 1998, 206 (02) :455-464
[4]   Introduction of a dedicated circulation phantom for comprehensive in vitro analysis of intravascular contrast material application [J].
Behrendt, Florian F. ;
Bruners, Philipp ;
Kalafut, John ;
Mahnken, Andreas H. ;
Keil, Sebastian ;
Plumhans, Cedric ;
Das, Marco ;
Stanzel, Sven ;
Wildberger, Joachim E. ;
Pfeffer, Jochen ;
Guenther, Rolf W. ;
Muehlenbruch, Georg .
INVESTIGATIVE RADIOLOGY, 2008, 43 (10) :729-736
[5]   Using a saline chaser to decrease contrast media in abdominal CT [J].
Dorio, PJ ;
Lee, FT ;
Henseler, KP ;
Pilot, M ;
Pozniak, MA ;
Winter, TC ;
Shock, SA .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2003, 180 (04) :929-934
[6]   Mathematical analysis of arterial enhancement and optimization of bolus geometry for CT angiography using the discrete Fourier transform [J].
Fleischmann, D ;
Hittmair, K .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1999, 23 (03) :474-484
[7]   Reduction of contrast material dose and artifacts by a saline flush using a double power injector in helical CT of the thorax [J].
Haage, P ;
Schmitz-Rode, T ;
Hübner, D ;
Piroth, W ;
Günther, RW .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2000, 174 (04) :1049-1053
[8]   Thoracic spiral CT: Delivery of contrast material pushed with injectable saline solution in a power injector [J].
Hopper, KD ;
Mosher, TJ ;
Kasales, CJ ;
TenHave, TR ;
Tully, DA ;
Weaver, JS .
RADIOLOGY, 1997, 205 (01) :269-271
[9]   Contrast-enhanced CT with saline flush technique using two automated injectors: How much contrast medium does it save? [J].
Irie, T ;
Kajitani, M ;
Yamaguchi, M ;
Itai, Y .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2002, 26 (02) :287-291
[10]   Saline flush effect for enhancement of aorta and coronary arteries at multidetector CT coronary angiography [J].
Kim, Dae Jung ;
Kim, Tae Hoon ;
Kim, Sang Jin ;
Kim, Dong Phil ;
Oh, Chi Suk ;
Ryu, Young Hoon ;
Kim, Young Jin ;
Choi, Byoung Wook .
RADIOLOGY, 2008, 246 (01) :110-115