Portable Accelerometer-Based Navigation System for Cup Placement of Total Hip Arthroplasty: A Prospective, Randomized, Controlled Study

被引:42
作者
Tanino, Hiromasa [1 ]
Nishida, Yasuhiro [1 ]
Mitsutake, Ryo [1 ]
Ito, Hiroshi [1 ]
机构
[1] Asahikawa Med Univ, Dept Orthopaed Surg, Midorigaoka Higashi 2-1-1-1, Asahikawa, Hokkaido 0788510, Japan
关键词
navigation; randomized trial; portable; total hip arthroplasty; cup position; TOTAL KNEE ARTHROPLASTY; COMPUTER-ASSISTED NAVIGATION; ACETABULAR COMPONENT; SAFE ZONE; ACCURACY; ALIGNMENT;
D O I
10.1016/j.arth.2019.08.044
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: Malposition of the acetabular component during total hip arthroplasty (THA) is associated with increased risk of dislocation, reduced range of motion, and accelerated wear. The purpose of this study is to compare cup positioning with a portable, accelerometer-based hip navigation system and conventional surgical technique. Methods: In a prospective, randomized, clinical study, cups were implanted with a portable, accelerometer-based hip navigation system (navigation group; n = 55) or conventional technique (conventional group; n = 55). THA was conducted in the lateral position and through posterior approach. The cup position was determined postoperatively on pelvic radiograph and computed tomography scans. Results: An average cup abduction of 39.2 degrees +/- 4.6 degrees (range, 27 degrees to 50 degrees) and an average cup anteversion of 14.6 degrees +/- 6.1 degrees (range, 1 degrees to 27.5 degrees) were found in the navigation group, and an average cup abduction of 42.9 degrees +/- 8.0 degrees (range, 23 degrees to 73 degrees) and an average cup anteversion of 11.6v +/- 7.7 degrees (range, -12.1 degrees to 25 degrees) in the conventional group. A smaller variation in the navigation group was indicated for cup abduction (P =.001). The deviations from the target cup position were significantly lower in the navigation group (P =.001,.016). While only 37 of 55 cups in the conventional group were inside the Lewinnek safe zone, 51 of 55 cups in the navigation group were placed inside this safe zone (P =.006). The navigation procedure took a mean of 10 minutes longer than the conventional technique. Conclusion: Use of the portable, accelerometer-based hip navigation system can improve cup positioning in THA. (C) 2019 Elsevier Inc. All rights reserved.
引用
收藏
页码:172 / 177
页数:6
相关论文
共 50 条
[31]   Accuracy of tibial component placement in unicompartmental knee arthroplasty performed using an accelerometer-based portable navigation system [J].
Yoshio Matsui ;
Shinichi Fukuoka ;
Sho Masuda ;
Masanori Matsuura ;
Toshiaki Masada ;
Kenji Fukunaga .
Knee Surgery, Sports Traumatology, Arthroscopy, 2020, 28 :3733-3739
[32]   Accuracy of tibial component placement in unicompartmental knee arthroplasty performed using an accelerometer-based portable navigation system [J].
Matsui, Yoshio ;
Fukuoka, Shinichi ;
Masuda, Sho ;
Matsuura, Masanori ;
Masada, Toshiaki ;
Fukunaga, Kenji .
KNEE SURGERY SPORTS TRAUMATOLOGY ARTHROSCOPY, 2020, 28 (12) :3733-3739
[33]   Evaluation of the accuracy of acetabular cup orientation using the accelerometer-based portable navigation system [J].
Hayashi, Shinya ;
Hashimoto, Shingo ;
Takayama, Koji ;
Matsumoto, Tomoyuki ;
Kamenaga, Tomoyuki ;
Fujishiro, Takaaki ;
Hiranaka, Takafumi ;
Niikura, Takahiro ;
Kuroda, Ryosuke .
JOURNAL OF ORTHOPAEDIC SCIENCE, 2020, 25 (04) :612-617
[34]   Accuracy of acetabular cup insertion in an anterolateral supine approach using an accelerometer-based portable navigation system [J].
Masahiro Hasegawa ;
Yohei Naito ;
Shine Tone ;
Hiroki Wakabayashi ;
Akihiro Sudo .
Journal of Artificial Organs, 2021, 24 :82-89
[35]   Accuracy of acetabular cup insertion in an anterolateral supine approach using an accelerometer-based portable navigation system [J].
Hasegawa, Masahiro ;
Naito, Yohei ;
Tone, Shine ;
Wakabayashi, Hiroki ;
Sudo, Akihiro .
JOURNAL OF ARTIFICIAL ORGANS, 2021, 24 (01) :82-89
[36]   An accelerometer-based navigation system provides acetabular cup orientation accuracy comparable to that of computed tomography-based navigation during total hip arthroplasty in the supine position [J].
Tetsunaga, Tomonori ;
Yamada, Kazuki ;
Tetsunaga, Tomoko ;
Sanki, Tomoaki ;
Kawamura, Yoshi ;
Ozaki, Toshifumi .
JOURNAL OF ORTHOPAEDIC SURGERY AND RESEARCH, 2020, 15 (01)
[37]   Computer-Assisted Total Hip Arthroplasty Improves Acetabular Prosthesis Placement Accuracy: A Multicenter, Randomized Controlled Clinical Study [J].
Zhou, Ge ;
Geng, Xiao ;
Zhang, Ming ;
Sun, Zhiwen ;
Li, Feng ;
Zhao, Minwei ;
Tian, Hua .
ORTHOPAEDIC SURGERY, 2024, 16 (12) :3078-3087
[38]   Precision of Cup Positioning Using a Novel Computed Tomography Based Navigation System in Total Hip Arthroplasty [J].
Nemati, Hassan M. ;
Christensson, Albin ;
Pettersson, Andreas ;
Nemeth, Gunnar ;
Flivik, Gunnar .
MEDICINA-LITHUANIA, 2024, 60 (10)
[39]   Extramedullary Guides Versus Portable, Accelerometer-Based Navigation for Tibial Alignment in Total Knee Arthroplasty: A Randomized, Controlled Trial: Winner of the 2013 HAP PAUL Award [J].
Nam, Denis ;
Cody, Elizabeth A. ;
Nguyen, Joseph T. ;
Figgie, Mark P. ;
Mayman, David J. .
JOURNAL OF ARTHROPLASTY, 2014, 29 (02) :288-294
[40]   Validation of navigation assisted cup placement in total hip arthroplasty [J].
Kumar, M. Ajith ;
Shetty, M. Shantharam ;
Kiran, Kanthi G. ;
Kini, Abhishek R. .
INTERNATIONAL ORTHOPAEDICS, 2012, 36 (01) :17-22