Noninvasive intracranial compliance monitoring - Technical note and clinical results

被引:27
作者
Lang, EW
Paulat, K
Witte, C
Zolondz, J
Mehdorn, HM
机构
[1] Univ Kiel, Dept Neurosurg, Kiel, Germany
[2] Univ Appl Sci, Dept Med Cybernet, Fachhochsch Ulm, Germany
[3] DWL Inc, Sipplingen, Germany
关键词
intracranial compliance; intracranial pressure monitoring; waveform analysis;
D O I
10.3171/jns.2003.98.1.0214
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Although invasive measurement of intracranial pressure (ICP) involving high-resolution waveform analysis allows assessment of intracranial compliance (ICC), it is only feasible in a few selected neurosurgical conditions. Intracranial compliance can be assessed using the high-frequency centroid (HFC), which is the power-weighted mean frequency within the 4 to 15-Hz band of the ICP waveform. The authors have systematically tested the utility, performance, and reliability of a noninvasive monitor of ICC. The underlying principle of this device is that the ICP transmission and its infrasonic waves are transmitted through the inner ear toward the tympanic membrane. If the outer ear is sealed in an airtight fashion, motions of the tympanic membrane cause air pressure fluctuations that can be recorded using a special sensor. The authors compared the HFC calculated from an intraparenchymal ICP sensor with that obtained simultaneously from an ipsilaterally placed noninvasive device during half of a respiratory cycle (peak to baseline) as well as for three random samples of three heart cycles. They analyzed 32 sessions in 13 patients in whom mechanical ventilation had been established. In four (11%) of 36 sessions they could not demonstrate an adequate signal. For the peak-to-baseline cycle, the mean invasively recorded HFC was 8.05 +/- 0.55 Hz (range 6.7-9 Hz) whereas the mean noninvasively recorded HFC was 8.04 +/- 0.49 Hz (range 7-9.3 Hz). The ICP was 8.5 +/- 5 mm Hg (range 2-24 mm Hg). For the three heart cycles randomly sampled, the values were 7.73 +/- 0.51 Hz (range 6.7-8.6 Hz) and 7.76 +/- 0.56 mm Hg (range 6.5-8.8 mm Hg), respectively. This device allows noninvasive assessment of ICC based on the HFC waveform analysis that is equivalent to that obtained by invasive intraparenchymal recording. The monitoring device may become a valuable tool for monitoring parameters in patients in whom placement of an intracranial sensor is not feasible but assessment of ICC as an alternative to ICP measurement is desired.
引用
收藏
页码:214 / 218
页数:5
相关论文
共 34 条
[1]   MR-intracranial pressure (ICP): A method to measure intracranial elastance and pressure noninvasively by means of MR imaging: Baboon and human study [J].
Alperin, NJ ;
Lee, SH ;
Loth, F ;
Raksin, PB ;
Lichtor, T .
RADIOLOGY, 2000, 217 (03) :877-885
[2]  
Andrews JC, 1996, INTRACRANIAL AND INTRALABYRINTHINE FLUIDS, P85
[3]   CEREBROSPINAL-FLUID PULSE PRESSURE AND INTRACRANIAL VOLUME-PRESSURE RELATIONSHIPS [J].
AVEZAAT, CJJ ;
VANEIJNDHOVEN, JHM ;
WYPER, DJ .
JOURNAL OF NEUROLOGY NEUROSURGERY AND PSYCHIATRY, 1979, 42 (08) :687-700
[4]  
Casselbrant M, 1979, Acta Otolaryngol Suppl, V362, P3
[5]  
Dubin MJ, 1998, NEUROL RES, V20, P533
[6]   A STUDY OF CEREBROSPINAL FLUID PULSE WAVE [J].
DUNBAR, HS ;
GUTHRIE, TC ;
KARPELL, B .
ARCHIVES OF NEUROLOGY, 1966, 14 (06) :624-&
[7]  
FOLTZ EL, 1979, ADV NEUROSURG, V8, P295
[8]  
Germon K, 1988, J Neurosci Nurs, V20, P344
[9]   Anatomy of the normal human cochlear aqueduct with functional implications [J].
Gopen, Q ;
Rosowski, JJ ;
Merchant, SN .
HEARING RESEARCH, 1997, 107 (1-2) :9-22
[10]   EPIDURAL PULSE WAVEFORM AS AN INDICATOR OF INTRACRANIAL-PRESSURE DYNAMICS [J].
HIRAI, O ;
HANDA, H ;
ISHIKAWA, M ;
KIM, SH .
SURGICAL NEUROLOGY, 1984, 21 (01) :67-74