Are We Being Fooled by Fluoroscopy? Distortion May Affect Limb-Length Measurements in Direct Anterior Total Hip Arthroplasty

被引:13
作者
Carlson, Victor R. [1 ]
Elliott, Iain S. [2 ]
DeKeyser, Graham J. [1 ]
Pelt, Christopher E. [1 ]
Anderson, Lucas A. [1 ]
Gililland, Jeremy M. [1 ]
机构
[1] Univ Utah, Dept Orthopaed, 590 Wakara Way, Salt Lake City, UT 84108 USA
[2] Univ Washington, Harborview Med Ctr, Dept Orthopaed, Seattle, WA 98104 USA
关键词
direct anterior total hip arthroplasty; image-intensified fluoroscopy; distortion; limb-length measurements; component positioning; AAPM/RSNA PHYSICS TUTORIAL; GEOMETRIC IMAGE DISTORTION; LEG LENGTH; GLOBAL METHOD; DISCREPANCY; DISLOCATION; RESIDENTS; AGREEMENT; POSITION;
D O I
10.1016/j.arth.2020.11.021
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: Distortion is an intrinsic phenomenon associated with image-intensified fluoroscopy that is both poorly understood and infrequently appreciated by orthopedic surgeons. Little information exists regarding its potential influence on intraoperative parameters during orthopedic surgery, let alone during direct anterior (DA) total hip arthroplasty (THA). The purpose of this study was to quantify the amount of potential error caused by fluoroscopic distortion during DA THA. Methods: Intra-operative fluoroscopic pelvic images from 74 DA THAs were reviewed by two independent readers. All images were obtained using the same fluoroscopic C-arm unit with a radiopaque grid attached to the image intensifier. The vertical distortion from a straight central horizontal line at the peripheries of images were measured and summed to yield the combined vertical distortion similar to how a surgeon calculates a side to side comparison of limb lengths. Simple linear regression was used to evaluate associations between total distortion and patient demographics, operating theaters, and various operative parameters. Results: The average combined distortion was 10.0mm (range 2.0-20.0mm). There was a significant difference in the average distortion observed in different theaters (P <.001). There was no association between distortion and patient demographics or fluoroscopic time (all, P >.05). Conclusion: Fluoroscopic distortion is unpredictable and can cause a substantial amount of error when comparing limb lengths during DA THA. This is a critical finding as this amount of inaccuracy could lead to unintended implant positioning and limb-length discrepancies if unaccounted for. (C) 2020 Elsevier Inc. All rights reserved.
引用
收藏
页码:1799 / 1803
页数:5
相关论文
共 32 条
[1]   A VIDEOFLUOROSCOPY METHOD FOR OPTICAL DISTORTION CORRECTION AND MEASUREMENT OF KNEE-JOINT KINEMATICS [J].
BALTZOPOULOS, V .
CLINICAL BIOMECHANICS, 1995, 10 (02) :85-92
[2]   Does Intraoperative Fluoroscopy Improve Limb-Length Discrepancy and Acetabular Component Positioning During Direct Anterior Total Hip Arthroplasty? [J].
Bingham, Joshua S. ;
Spangehl, Mark J. ;
Hines, Jeremy T. ;
Taunton, Michael J. ;
Schwartz, Adam J. .
JOURNAL OF ARTHROPLASTY, 2018, 33 (09) :2927-2931
[3]   Applying the right statistics: analyses of measurement studies [J].
Bland, JM ;
Altman, DG .
ULTRASOUND IN OBSTETRICS & GYNECOLOGY, 2003, 22 (01) :85-93
[4]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[5]  
Bolink S, 2019, CLIN BIOMECH BRISTOL, V68
[6]   The effect of X-ray beam distortion on the Edwards Sapien XT (TM) trans-catheter aortic valve replacement prosthesis [J].
Crowhurst, James A. ;
Poon, Karl K. ;
Murdoch, Dale ;
Incani, Alexander ;
Raffel, Owen C. ;
Liddicoat, Annelise ;
Walters, Darren .
JOURNAL OF MEDICAL RADIATION SCIENCES, 2015, 62 (04) :239-245
[7]  
Dudda Marcel, 2010, J Orthop Surg (Hong Kong), V18, P179
[8]   Three-dimensional computed tomographic reconstruction using a C-arm mounted XRII: Correction of image intensifier distortion [J].
Fahrig, R ;
Moreau, M ;
Holdsworth, DW .
MEDICAL PHYSICS, 1997, 24 (07) :1097-1106
[9]   A global method based on thin-plate splines for correction of geometric distortion: An application to fluoroscopic images [J].
Fantozzi, S ;
Cappello, A ;
Leardini, A .
MEDICAL PHYSICS, 2003, 30 (02) :124-131
[10]   A Fluoroscopic Grid in Supine Total Hip Arthroplasty Improving Cup Position, Limb Length, and Hip Offset [J].
Gililland, Jeremy M. ;
Anderson, Lucas A. ;
Boffeli, Shannon L. ;
Pelt, Christopher E. ;
Peters, Christopher L. ;
Kubiak, Erik N. .
JOURNAL OF ARTHROPLASTY, 2012, 27 (08) :111-116