Improving the trade-off between simulation time and accuracy in efficiency calibrations with the code DETEFF

被引:12
作者
Cornejo Diaz, N. [2 ]
Jurado Vargas, M. [1 ]
机构
[1] Univ Extremadura, Dept Phys, E-06071 Badajoz, Spain
[2] Ctr Radiol Protect & Hyg, Havana, Cuba
关键词
Gamma-ray spectrometry; Efficiency; HPGe detectors; Monte Carlo simulation; GAMMA-RAY SPECTROMETRY; HPGE DETECTOR;
D O I
10.1016/j.apradiso.2009.11.021
中图分类号
O61 [无机化学];
学科分类号
070301 ; 081704 ;
摘要
Quick and relatively simple procedures were incorporated into the Monte Carlo code DETEFF in order to consider the escape of Bremsstrahlung radiation and secondary electrons. The relative bias in efficiency values was thus reduced for photon energies between 1500 and 2000 keV, without any noticeable increment of the simulation time A relatively simple method was also included to consider the rounding of detector edges The validation studies showed relative deviations of about 1% in the energy range 10-2000 key. (C) 2009 Elsevier Ltd All rights reserved
引用
收藏
页码:1413 / 1417
页数:5
相关论文
共 16 条
  • [1] [Anonymous], LAUR003581 LOS AL NA
  • [2] BRIESMESITER JF, 2000, CCC700 LOS AL NAT LA
  • [3] *CANB IND, 2003, CAT CANBERRA PROD CA
  • [4] DETEFF:: An improved Monte Carlo computer program for evaluating the efficiency in coaxial gamma-ray detectors
    Diaz, N. Cornejo
    Vargas, M. Jurado
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2008, 586 (02) : 204 - 210
  • [5] Experimental and MC determination of HPGe detector efficiency in the 40-2754 keV energy range for measuring point source geometry with the source-to-detector distance of 25 cm
    Dryak, Pavel
    Kovar, Petr
    [J]. APPLIED RADIATION AND ISOTOPES, 2006, 64 (10-11) : 1346 - 1349
  • [6] Monte Carlo modelling of germanium crystals that are tilted and have rounded front edges
    Gasparro, Joel
    Hult, Mikael
    Johnston, Peter N.
    Tagziria, Hamid
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2008, 594 (02) : 196 - 201
  • [7] The use of Monte Carlo calculations in the determination of a Ge detector efficiency curve
    Helmer, RG
    Hardy, JC
    Iacob, VE
    Sanchez-Vega, M
    Neilson, RG
    Nelson, J
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2003, 511 (03) : 360 - 381
  • [8] GEANT4 code for simulation of a germanium gamma-ray detector and its application to efficiency calibration
    Hurtado, S
    García-León, M
    García-Tenorio, R
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2004, 518 (03) : 764 - 774
  • [9] Intercomparison of efficiency transfer software for gamma-ray spectrometry
    Lépy, MC
    Altzitzoglou, T
    Arnold, D
    Bronson, F
    Noy, RC
    Décombaz, M
    De Corte, F
    Edelmaier, R
    Peraza, EH
    Klemola, S
    Korun, M
    Kralik, M
    Neder, H
    Plagnard, J
    Pommé, S
    de Sanoit, J
    Sima, O
    Ugletveit, F
    Van Velzen, L
    Vidmar, T
    [J]. APPLIED RADIATION AND ISOTOPES, 2001, 55 (04) : 493 - 503
  • [10] LEPY MC, 2000, R5894 CEA