In silico evaluation of the thermal stress induced by MRI switched gradient fields in patients with metallic hip implant

被引:16
作者
Arduino, Alessandro [1 ]
Bottauscio, Oriano [1 ]
Bruehl, Ruediger [2 ,3 ]
Chiampi, Mario [1 ]
Zilberti, Luca [1 ]
机构
[1] Ist Nazl Ric Metrol, Turin, Italy
[2] PTB, Braunschweig, Germany
[3] PTB, Berlin, Germany
基金
欧盟地平线“2020”;
关键词
dosimetry; magnetic resonance imaging (MRI); MR safety; gradient coils; medical implants; prostheses; numerical simulation; TEMPERATURE ELEVATION; BODY; PROSTHESES; COILS;
D O I
10.1088/1361-6560/ab5428
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This work focuses on the in silico evaluation of the energy deposed by MRI switched gradient fields in bulk metallic implants and the consequent temperature increase in the surrounding tissues. An original computational strategy, based on the subdivision of the gradient coil switching sequences into sub-signals and on the time-harmonic electromagnetic field solution, allows to realistically simulate the evolution of the phenomena produced by the gradient coils fed according to any MRI sequence. Then, Pennes' bioheat equation is solved through a Douglas-Gunn time split scheme to compute the time-dependent temperature increase. The procedure is validated by comparison with laboratory results, using a component of a realistic hip implant embedded within a phantom, obtaining an agreement on the temperature increase better than 5%, lower than the overall measurement uncertainty. The heating generated inside the body of a patient with a unilateral hip implant when undergoing an Echo-Planar Imaging (EPI) MRI sequence is evaluated and the role of the parameters affecting the thermal results (body position, coil performing the frequency encoding, effects of thermoregulation) is discussed. The results show that the gradient coils can generate local increases of temperature up to some kelvin when acting without radiofrequency excitation. Hence, their contribution in general should not be disregarded when evaluating patients' safety.
引用
收藏
页数:16
相关论文
共 29 条
[22]   Numerical simulation of SAR induced around Co-Cr-Mo hip prostheses in situ exposed to RF fields associated with 1.5 and 3 T MRI body coils [J].
Powell, John ;
Papadaki, Annie ;
Hand, Jeff ;
Hart, Alister ;
McRobbie, Donald .
MAGNETIC RESONANCE IN MEDICINE, 2012, 68 (03) :960-968
[23]   On quasi-static models hidden in Maxwell's equations [J].
Rapetti, Francesca ;
Rousseaux, Germain .
APPLIED NUMERICAL MATHEMATICS, 2014, 79 :92-106
[24]   Retrospective analysis of RF heating measurements of passive medical implants [J].
Song, Ting ;
Xu, Zhiheng ;
Iacono, Maria Ida ;
Angelone, Leonardo M. ;
Rajan, Sunder .
MAGNETIC RESONANCE IN MEDICINE, 2018, 80 (06) :2726-2730
[25]   Review of the mathematical functions used to model the temperature dependence of electrical and thermal conductivities of biological tissue in radiofrequency ablation [J].
Trujillo, Macarena ;
Berjano, Enrique .
INTERNATIONAL JOURNAL OF HYPERTHERMIA, 2013, 29 (06) :590-597
[26]   Thermal Effects Associated with RF Exposures in Diagnostic MRI: Overview of Existing and Emerging Concepts of Protection [J].
van den Brink, Johan S. .
CONCEPTS IN MAGNETIC RESONANCE PART B-MAGNETIC RESONANCE ENGINEERING, 2019, 2019
[27]   The underestimated role of gradient coils in MRI safety [J].
Zilberti, Luca ;
Arduino, Alessandro ;
Bottauscio, Oriano ;
Chiampi, Mario .
MAGNETIC RESONANCE IN MEDICINE, 2017, 77 (01) :13-15
[28]   Numerical prediction of temperature elevation induced around metallic hip prostheses by traditional, split, and uniplanar gradient coils [J].
Zilberti, Luca ;
Bottauscio, Oriano ;
Chiampi, Mario ;
Hand, Jeffrey ;
Lopez, Hector Sanchez ;
Bruehl, Ruediger ;
Crozier, Stuart .
MAGNETIC RESONANCE IN MEDICINE, 2015, 74 (01) :272-279
[29]   Collateral Thermal Effect of MRI-LINAC Gradient Coils on Metallic Hip Prostheses [J].
Zilberti, Luca ;
Bottauscio, Oriano ;
Chiampi, Mario ;
Hand, Jeff ;
Lopez, Hector Sanchez ;
Crozier, Stuart .
IEEE TRANSACTIONS ON MAGNETICS, 2014, 50 (11)