Effects of Raw Material and Rotational Speed on the Cyclic Fatigue of ProFile Vortex Rotary Instruments

被引:112
作者
Gao, Yong [1 ]
Shotton, Vincent [1 ]
Wilkinson, Kevin [1 ]
Phillips, Greg [1 ]
Ben Johnson, William [1 ]
机构
[1] DENTSPLY Tulsa Dent Specialties, Tulsa, OK 74135 USA
关键词
Cyclic fatigue; endodontic rotary instrument; nickel-titanium; ProFile Vortex; rotational speed; TITANIUM ENDODONTIC INSTRUMENTS; SHAPE-MEMORY ALLOYS; FRACTURE TOUGHENING MECHANISM; CLINICAL-USE; MARTENSITE-TRANSFORMATION; NITINOL; DRIVEN; FILES; BREAKAGE; DEFECTS;
D O I
10.1016/j.joen.2010.02.015
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Introduction: This in vitro study compared the cyclic fatigue resistance of Profile Vortex rotary instruments (DENTSPLY Tulsa Dental, Tulsa, OK) made of two different raw materials M-Wire and regular superelastic wire (SE-wire) at two different rotational speeds Two rotary-based Profile Vortex instruments with taper 0 04 and 0 06 (25 mm in length and #30 in size) were compared Methods. Cyclic fatigue testing was performed by rotating instruments in an artificially constructed stainless steel canal with a 5-mm radius and a 90 degrees angle of curvature at two different test frequencies 300 and 500 rpm. The time to failure was recorded, and the total number of cycles to failure was calculated and compared for a total of 160 samples Statistical analysis was performed with one-way analysis of variance and the Tukey test for individual comparisons (confidence level of p < 0.05) Fracture surfaces of broken instruments were also observed under scanning electron microscopy Over 50% of broken files made of SE-wire exhibited multiple crack initiation sites compared with the single crack initiation on files made of M-Wire. Results: Overall, there were significant differences in rotary instruments made of different materials (p < 0.001) However, for rotary instruments made from the same material (either M-Wire or SE-wire), there was no significant difference (p > 0 05) of cyclic fatigue life under different rotational speeds (300 and 500 rpm) Conclusions: Pro File Vortex files made of M-Wire exhibited superior cyclic fatigue resistance (similar to 150% longer in fatigue life) compared with those made of regular SE-wire at two test frequencies (300 and 500 rpm). (J Endod 2010,36.1205-1209)
引用
收藏
页码:1205 / 1209
页数:5
相关论文
共 26 条
  • [1] Metallurgical Characterization of a New Nickel-Titanium Wire for Rotary Endodontic Instruments
    Alapati, Satish B.
    Brantley, William A.
    Iijima, Masahiro
    Clark, William A. T.
    Kovarik, Libor
    Buie, Caesar
    Liu, Jie
    Ben Johnson, William
    [J]. JOURNAL OF ENDODONTICS, 2009, 35 (11) : 1589 - 1593
  • [2] Berendt C., 2007, US Patent, Patent No. [20070072147A1, 20070072147]
  • [3] Defects in ProTaper S1 instruments after clinical use: fractographic examination
    Cheung, GSP
    Peng, B
    Bian, Z
    Shen, Y
    Darvell, BW
    [J]. INTERNATIONAL ENDODONTIC JOURNAL, 2005, 38 (11) : 802 - 809
  • [4] Comparison of fracture rate, deformation rate, and efficiency between rotary endodontic instruments driven at 150 rpm and 350 rpm
    Daugherty, DW
    Gound, TG
    Comer, TL
    [J]. JOURNAL OF ENDODONTICS, 2001, 27 (02) : 93 - 95
  • [5] Dauskardt RH, 1989, SHAPE MEMORY MAT, V9, P243
  • [6] Effect of rotational speed on the breakage of nickel-titanium rotary files
    Dietz, DB
    Di Fiore, PM
    Bahcall, JK
    Lautenschlager, EP
    [J]. JOURNAL OF ENDODONTICS, 2000, 26 (02) : 68 - 71
  • [7] Effect of rotational speed on nickel-titanium file distortion
    Gabel, WP
    Hoen, M
    Steiman, HR
    Pink, FE
    Dietz, R
    [J]. JOURNAL OF ENDODONTICS, 1999, 25 (11) : 752 - 754
  • [8] Fatigue resistance of engine-driven rotary nickel-titanium instruments produced by new manufacturing methods
    Gambarini, Gianluca
    Grande, Nicola Maria
    Plotino, Gianluca
    Somma, Francesco
    Garala, Manish
    De Luca, Massimo
    Testarelli, Luca
    [J]. JOURNAL OF ENDODONTICS, 2008, 34 (08) : 1003 - 1005
  • [9] Dynamic and cyclic fatigue of engine-driven rotary nickel-titanium endodontic instruments
    Haïkel, Y
    Serfaty, R
    Bateman, G
    Senger, B
    Allemann, C
    [J]. JOURNAL OF ENDODONTICS, 1999, 25 (06) : 434 - 440
  • [10] Comparison between a Novel Nickel-Titanium Alloy and 508 Nitinol on the Cyclic Fatigue Life of ProFile 25/.04 Rotary Instruments
    Johnson, Eugenia
    Lloyd, Adain
    Kuttler, Sergio
    Namerow, Kenneth
    [J]. JOURNAL OF ENDODONTICS, 2008, 34 (11) : 1406 - 1409