Evaluation of monoenergetic imaging to reduce metallic instrumentation artifacts in computed tomography of the cervical spine

被引:34
作者
Komlosi, Peter [1 ]
Grady, Deborah [1 ]
Smith, Justin S. [2 ]
Shaffrey, Christopher I. [2 ]
Goode, Allen R. [1 ]
Judy, Patricia G. [1 ]
Shaffrey, Mark [2 ]
Wintermark, Max [1 ,3 ]
机构
[1] Univ Virginia Hlth Syst, Dept Radiol & Med Imaging, Div Neuroradiol, Charlottesville, VA USA
[2] Univ Virginia Hlth Syst, Dept Neurosurg, Charlottesville, VA USA
[3] CHU Vaudois, Dept Radiol, CH-1011 Lausanne, Switzerland
关键词
dual-energy CT; cervical spine; spine hardware; image artifacts; technique; DUAL-ENERGY CT; IMPLANTS;
D O I
10.3171/2014.10.SPINE14463
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
OBJECT Monoenergetic imaging with dual-energy CT has been proposed to reduce metallic artifacts in comparison with conventional polychromatic CT. The purpose of this study is to systematically evaluate and define the optimal dual-energy CT imaging parameters for specific cervical spinal implant alloy compositions. METHODS Spinal fixation rods of cobalt-chromium or titanium alloy inserted into the cervical spine section of an Alderson Rando anthropomorphic phantom were imaged ex vivo with fast-kilovoltage switching CT at 80 and 140 peak kV. The collimation width and field of view were varied between 20 and 40 mm and medium to large, respectively. Extrapolated monoenergetic images were generated at 70, 90, 110, and 130 kiloelectron volts (keV). The standard deviation of voxel intensities along a circular line profile around the spine was used as an index of the magnitude of metallic artifact. RESULTS The metallic artifact was more conspicuous around the fixation rods made of cobalt-chromium than those of titanium alloy. The magnitude of metallic artifact seen with titanium fixation rods was minimized at monoenergies of 90 keV and higher, using a collimation width of 20 mm and large field of view. The magnitude of metallic artifact with cobalt-chromium fixation rods was minimized at monoenergies of 110 keV and higher; collimation width or field of view had no effect. CONCLUSIONS Optimization of acquisition settings used with monoenergetic CT studies might yield reduced metallic artifacts.
引用
收藏
页码:34 / 38
页数:5
相关论文
共 18 条
  • [2] BONE-MARROW DOSE IN CHEST RADIOGRAPHY - POSTEROANTERIOR VS ANTEROPOSTERIOR PROJECTION
    ARCHER, BR
    WHITMORE, RC
    NORTH, LB
    BUSHONG, SC
    [J]. RADIOLOGY, 1979, 133 (01) : 211 - 216
  • [3] Metal artifact reduction by dual energy computed tomography using monoenergetic extrapolation
    Bamberg, Fabian
    Dierks, Alexander
    Nikolaou, Konstantin
    Reiser, Maximilian F.
    Becker, Christoph R.
    Johnson, Thorsten R. C.
    [J]. EUROPEAN RADIOLOGY, 2011, 21 (07) : 1424 - 1429
  • [4] Dose and image quality at CT pulmonary angiography-comparison of first and second generation dual-energy CT and 64-slice CT
    Bauer, Ralf W.
    Kramer, Sebastian
    Renker, Matthias
    Schell, Boris
    Larson, Maya Christina
    Beeres, Martin
    Lehnert, Thomas
    Jacobi, Volkmar
    Vogl, Thomas J.
    Kerl, Josef Matthias
    [J]. EUROPEAN RADIOLOGY, 2011, 21 (10) : 2139 - 2147
  • [5] Dual Energy CT Myelography after Lumbar Osteosynthesis
    Grams, A. E.
    Sender, J.
    Moritz, R.
    Obert, M.
    Stein, M.
    Oertel, M.
    Krombach, G. A.
    Gizewski, E. R.
    Schmidt, T.
    [J]. ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2014, 186 (07): : 670 - 674
  • [6] Metallic artefact reduction with monoenergetic dual-energy CT: systematic ex vivo evaluation of posterior spinal fusion implants from various vendors and different spine levels
    Guggenberger, R.
    Winklhofer, S.
    Osterhoff, G.
    Wanner, G. A.
    Fortunati, M.
    Andreisek, G.
    Alkadhi, H.
    Stolzmann, P.
    [J]. EUROPEAN RADIOLOGY, 2012, 22 (11) : 2357 - 2364
  • [7] HEMMINGSSON A, 1986, J COMPUT ASSIST TOMO, V10, P490
  • [8] An operation for progressive spinal deformities - A preliminary report of three cases from the service of the orthopaedic hospital
    Hibbs, Russell A.
    [J]. CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2007, (460) : 17 - 20
  • [9] THE EFFECTS OF SCATTER IN X-RAY COMPUTED-TOMOGRAPHY
    JOSEPH, PM
    SPITAL, RD
    [J]. MEDICAL PHYSICS, 1982, 9 (04) : 464 - 472
  • [10] A comparison of magnetic and radiographic imaging artifact after using three types of metal rods: stainless steel, titanium, and vitallium
    Knott, Patrick T.
    Mardjetko, Steven M.
    Kim, Richard H.
    Cotter, Timothy M.
    Dunn, Megan M.
    Patel, Shivani T.
    Spencer, Matthew J.
    Wilson, Alan S.
    Tager, David S.
    [J]. SPINE JOURNAL, 2010, 10 (09) : 789 - 794