Acetabular Cartilage Thickness: Accuracy of Three-Dimensional Reconstructions from Multidetector CT Arthrograms in a Cadaver Study

被引:39
作者
Allen, Bryce C. [1 ]
Peters, Christopher L. [1 ]
Brown, Nicholas A. T. [1 ]
Anderson, Andrew E. [1 ,2 ,3 ]
机构
[1] Univ Utah, Harold K Dunn Orthopaed Res Lab, Dept Orthopaed, Salt Lake City, UT 84108 USA
[2] Univ Utah, Dept Bioengn, Salt Lake City, UT 84108 USA
[3] Univ Utah, Sci Comp & Imaging Inst, Salt Lake City, UT 84108 USA
基金
美国国家卫生研究院;
关键词
HUMAN ARTICULAR-CARTILAGE; HUMAN HIP-JOINT; MR-ARTHROGRAPHY; FEMOROACETABULAR IMPINGEMENT; PRESSURE DISTRIBUTION; CONTRAST AGENT; CONTACT AREAS; LOWER-LIMB; REPRODUCIBILITY; OSTEOARTHRITIS;
D O I
10.1148/radiol.10081876
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To prospectively quantify the accuracy of hip cartilage thickness estimated from three-dimensional (3D) surfaces, generated by segmenting multidetector computed tomographic (CT) arthrograms by using direct physical measurements of cartilage thickness as the reference standard. Materials and Methods: Four fresh-frozen cadaver hip joints from two male donors, ages 43 and 46 years, were obtained; institutional review board approval for cadaver research was also obtained. Sixteen holes were drilled perpendicular to the cartilage of four cadaveric acetabula (two specimens). Hip capsules were surgically closed, injected with contrast material, and scanned by using multidetector CT. After scanning, 5.3-mm cores were harvested concentrically at each drill hole and cartilage thickness was measured with a microscope. Cartilage was reconstructed in 3D by using commercial software. Segmentations were repeated by two authors. Reconstructed cartilage thickness was determined by using a published algorithm. Bland-Altman plots and linear regression were used to assess accuracy. Repeatability was quantified by using the coefficient of variation, intraclass correlation coefficient (ICC), repeatability coefficient, and percentage variability. Results: Cartilage was reconstructed to a bias of -0.13 mm and a repeatability coefficient of +/- 0.46 mm. Regression of the scatterplots indicated a tendency for multidetector CT to overestimate thickness. Intra-and interobserver repeatability were very good. For intraobserver correlation, the coefficient of variation was 14.80%, the ICC was 0.88, the repeatability coefficient was 0.55 mm, and the percentage variability was 11.77%. For interobserver correlation, the coefficient of variation was 13.47%, the ICC was 0.90, the repeatability coefficient was 0.52 mm, and the percentage variability was 11.63%. Conclusion: Assuming that an accuracy of approximately +/- 0.5 mm is sufficient, reconstructions of cartilage geometry from multidetector CT arthrographic data could be used as a preoperative surgical planning tool. (C) RSNA, 2010
引用
收藏
页码:544 / 552
页数:9
相关论文
共 55 条
  • [1] The distribution of cartilage thickness within the joints of the lower limb of elderly individuals
    Adam, C
    Eckstein, F
    Milz, S
    Putz, R
    [J]. JOURNAL OF ANATOMY, 1998, 193 : 203 - 214
  • [2] MEASUREMENT IN MEDICINE - THE ANALYSIS OF METHOD COMPARISON STUDIES
    ALTMAN, DG
    BLAND, JM
    [J]. JOURNAL OF THE ROYAL STATISTICAL SOCIETY SERIES D-THE STATISTICIAN, 1983, 32 (03) : 307 - 317
  • [3] Subject-specific finite element model of the pelvis: Development, validation and sensitivity studies
    Anderson, AE
    Peters, CL
    Tuttle, BD
    Weiss, JA
    [J]. JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 2005, 127 (03): : 364 - 373
  • [4] Cartilage thickness: Factors influencing multidetector CT measurements in a phantom study
    Anderson, Andrew E.
    Ellis, Benjamin J.
    Peters, Christopher L.
    Weiss, Jeffrey A.
    [J]. RADIOLOGY, 2008, 246 (01) : 133 - 141
  • [5] Andriacchi TP, 2000, J REHABIL RES DEV, V37, P163
  • [6] QUANTITATION OF ARTICULAR SURFACE-TOPOGRAPHY AND CARTILAGE THICKNESS IN KNEE JOINTS USING STEREOPHOTOGRAMMETRY
    ATESHIAN, GA
    SOSLOWSKY, LJ
    MOW, VC
    [J]. JOURNAL OF BIOMECHANICS, 1991, 24 (08) : 761 - 776
  • [7] COMPARATIVE-STUDY OF THE INTRINSIC MECHANICAL-PROPERTIES OF THE HUMAN ACETABULAR AND FEMORAL-HEAD CARTILAGE
    ATHANASIOU, KA
    AGARWAL, A
    DZIDA, FJ
    [J]. JOURNAL OF ORTHOPAEDIC RESEARCH, 1994, 12 (03) : 340 - 349
  • [8] Magnetic resonance imaging with gadolinium arthrography to assess acetabular cartilage delamination -: A report of four cases
    Beaulé, PE
    Zaragoza, E
    Copelan, N
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2004, 86A (10) : 2294 - 2298
  • [9] Bland JM, 1999, STAT METHODS MED RES, V8, P135, DOI 10.1177/096228029900800204
  • [10] Comparison of double-contrast CT arthrography image quality with nonionic contrast agents - Isotonic dimeric iodixanol 270 mg I/mL and monomeric iohexol 300 mg I/mL
    Blum, AG
    Simon, JM
    Cotten, A
    Quirin-Cosmidis, I
    Boyer, B
    Boutry, N
    Antonini, JP
    [J]. INVESTIGATIVE RADIOLOGY, 2000, 35 (05) : 304 - 310