VIBRO-ACOUSTIC RESPONSE OF TYMPANIC-MEMBRANE-LIKE MODELS

被引:0
作者
Rona, Aldo [1 ,2 ]
Cockrill, Martin [1 ,2 ]
Picinali, Lorenzo [3 ]
Panday, Prakash [3 ]
Tripathi, Ramesh [3 ]
Ayub, Mohammed [3 ]
机构
[1] Univ Leicester, ASDEC, Leicester LE1 7RH, Leics, England
[2] Univ Leicester, Dept Engn, Leicester LE1 7RH, Leics, England
[3] De Montfort Univ, Fac Technol, Leicester LE1 9BH, Leics, England
来源
PROCEEDINGS OF THE 22ND INTERNATIONAL CONGRESS ON SOUND AND VIBRATION: MAJOR CHALLENGES IN ACOUSTICS, NOISE AND VIBRATION RESEARCH, 2015 | 2015年
关键词
D O I
暂无
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Equivalent measurements to the Real Ear Measurement (REM) procedure are sought using faster and contactless Laser Doppler Vibrometry. A feasibility study has been performed on tympanic membrane-like rubber and latex models, in which the vibro-acoustic response was measured using both standard REM equipment and a PSV-500-HM Laser Doppler Vibrometer (LDV). The membrane surface velocity distribution was measured by the LDV and the transfer function between the forcing acoustic field and the vibration response of the membrane was calculated. The variability in the transfer function due to membrane shape, membrane type, and measurement position was determined. This study indicates that the transfer functions from measurements in the neighbourhood of the modelled umbo spot repeated several days apart were found to be consistent within typical clinical measurement tolerances. These results seem to indicate that Laser Doppler Vibrometry has the potential for replacing the standard REM procedure and equipment in patient-specific tests.
引用
收藏
页数:8
相关论文
共 8 条
[1]   Children's amplification needs - same or different from adults? [J].
Ching, TYC ;
Dillon, H ;
Byrne, D .
SCANDINAVIAN AUDIOLOGY, 2001, 30 :54-60
[2]   SHAPE AND DERIVED GEOMETRICAL PARAMETERS OF THE ADULT, HUMAN TYMPANIC MEMBRANE MEASURED WITH A PHASE-SHIFT MOIRE INTERFEROMETER [J].
DECRAEMER, WF ;
DIRCKX, JJJ ;
FUNNELL, WRJ .
HEARING RESEARCH, 1991, 51 (01) :107-122
[3]  
Goode RL, 1996, AM J OTOL, V17, P813
[4]  
Hiipakka M., 2009, IEEE WORKSH APPL SIG
[5]  
Katz J., 1994, HDB CLIN AUDIOLOGY, P124
[6]  
Naser N. E., 2012, NEW IRAQI J MED, V8, P62
[7]  
Picinali L., 2012, NONINVASIVE CALIBRAT
[8]  
Strutt J. W., 1897, THEORY SOUND, VI, P306